Masimo Rad-8 Operator's Manual

Masimo Rad-8 Operator's Manual

Signal extraction pulse oximeter
Hide thumbs Also See for Rad-8:
Table of Contents

Advertisement

Rad-8
®
Pulse Oximeter
Signal Extraction
Pulse Oximeter
O P E R A T O R ' S
M A N U A L

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Rad-8 and is the answer not in the manual?

Questions and answers

Trish
April 4, 2025

how do to save alarm limits

1 comments:
Mr. Anderson
April 4, 2025

To save alarm limits on the Masimo Rad-8:

1. Change the alarm parameter (e.g., High or Low Saturation Alarm) using the Up/Down buttons.
2. Press the Mode/Enter button to accept and save the new setting.
3. Verify the updated value is shown on the Saturation Alarm Limit Display.

This process saves the alarm limit setting.

This answer is automatically generated

Summary of Contents for Masimo Rad-8

  • Page 1 Rad-8 ® Pulse Oximeter Signal Extraction Pulse Oximeter O P E R A T O R ’ S M A N U A L...
  • Page 2 Rad-8 Pulse Oximeter Signal Extraction Pulse Oximeter OPERATOR’S MANUAL...
  • Page 4 The Rad-8 Operating Instructions provide the necessary information for proper operation of the Rad-8 device. General knowledge of pulse oximetry and an understanding of the features and functions of the Rad-8 are a prerequisite for its proper use. Do not operate the Rad-8 without completely reading and understanding the instructions in this manual.
  • Page 5 ■ As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement or strangulation. ■ Do not place the Rad-8 or accessories in any position that might cause it to fall on the patient. Do not lift the Rad-8 by the power cord or any other cable.
  • Page 6 SAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES ■ If using Rad-8 during full body radiation, keep the sensor out of the radiation fi eld. If the sensor is exposed to the radiation, the reading might be inaccurate or the device might read zero for the duration of the active irradiation period.
  • Page 7: Table Of Contents

    SpO 2 General Description....................1-4 Principle of Operation ....................... 1-4 Functional Saturation ......................1-5 Rad-8 vs. Drawn Whole Blood Measurements ..............1-5 Signal Extraction Technology (SET) for SpO 2 Measurements ........1-6 FastSat..........................1-6 Masimo SET Parallel Engines ..................1-6 Masimo SET DST ®...
  • Page 8 SECTION 5 - ALARMS AND MESSAGES Alarm Identification ........................5-1 Alarm Indication ..........................5-1 Alarm Limits ..........................5-2 Alarm Silence ........................5-3 Alarm Bell ........................... 5-3 Alarm On/Off ........................5-3 Messages ............................. 5-4 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 9 M-LNCS ™ DCI & DCIP Sensors ................... 8-4 Sensor Accuracy ......................8-5 Cleaning And Reuse Of Masimo Reusable Sensors and Cables ........8-5 Reattachment of Single Use Adhesive Sensors .............. 8-5 SECTION 9 - SERVICE AND MAINTENANCE Introduction ........................... 9-1 Cleaning ............................
  • Page 10: About This Manual

    A b o u t T h i s M a n u a l This manual explains how to set up and use the Rad-8 Pulse Oximeter. Important safety information relating to general use of the Rad-8 Pulse Oximeter appears before this introduction.
  • Page 11: Warnings, Cautions And Notes

    Sample of Caution: CAUTION: THIS IS A SAMPLE OF A CAUTION STATEMENT. A NOTE is provided when extra general information is applicable. Sample of Note: NOTE: This is a sample of a Note. Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 12: Product Description

    The device consists of two models: the vertical Rad-8 and the horizontal Rad-8. FEATURES AND BENEFITS These features are common to the Rad-8: ■ Masimo SET is clinically proven to be the highest sensitivity and specifi city pulse oximeter technology in the world.
  • Page 13: Pulse Oximetry

    Deoxyhemoglobin 1000 Wavelength (nm) 2. The amount of arterial blood in tissue changes with your pulse (photoplethysmography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well. Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 14: Functional Saturation

    The maximum of the skin surface temperature is measured at an ambient temperature of less than 106º F (41º C). This is verified by Masimo sensor skin temperature test procedures.
  • Page 15: Signal Extraction Technology (Set) For Spo 2 Measurements

    fi ltering. Adaptive fi lters are powerful because they are able to adapt to the varying physiologic signals and/or noise and separate them by looking at the whole signal and breaking it down to its fundamental components. The Masimo ®...
  • Page 16: Masimo Set Dst

    MASIMO SET DST Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 18: Introduction

    ■ Full Masimo SET technology in a small compact device ■ Supports the full line of Masimo sensors and patient cables (see Section 8, Sensors and Patient Cables) ■ Supports standardization of sensors and pulse oximetry technology throughout the hospital Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 19: Rad-8 Front Panel (Horizonatal Model)

    RAD-8 FRONT PANEL (HORIZONTAL MODEL) 8 8 8 8 8 8 8 8 8 8 8 8 ® RAD-8 FRONT PANEL (VERTICAL MODEL) 8 8 8 8 8 8 ® Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 20 4 levels of brightness. Each press of the button Brightness Button increases the brightness one level. Once level 4 is accessed, an additional press of the button returns the brightness to level 1. Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 21: Rad-8 Rear Panel

    Call Connector must be IEC-60950 compliant. Use the Serial Output Connector to connect a serial device, SERIAL OUTPUT including a serial printer, to the Rad-8. See Section 7, Serial Interface Specifications. All external device connections to CONNECTOR the Serial Output Connector must be IEC-60950 compliant.
  • Page 22: Symbols

    5%-95% RH +70 C Storage temperature range: +70˚C to -40˚C -40 C Storage altitude range: +1600hPa to +500hPa +1060 hPa - +500 hPa 795 mmHg - 375 mmHg Keep dry Fragile/breakable, handle with care Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 24: Introduction

    Always use a hospital grade, AC power cable to connect the pulse oximeter to an AC power source. CAUTION: DO NOT CONNECT THE RAD-8 PULSE OXIMETER TO AN AC OUTLET CONTROLLED BY A SWITCH. Verify the AC power voltage and frequency before use. Verify that the power source can provide adequate power rating as indicated on the rear panel.
  • Page 25: Initial Battery Charging

    PULSE OXIMETER IS CONNECTED TO THE AC MAIN POWER SUPPLY. INITIAL INSTALLATION Place the Rad-8 on a stable hard flat surface near the patient. Always place the Rad- on a dry surface. Maintain a minimum of 1 inch (2.54 cm) free space around the device.
  • Page 26: Introduction

    4. Refer to the Directions for Use of the sensor before attaching the sensor to the patient. 5. Attach the Masimo sensor to the Patient. Connect the sensor to the patient cable with the logos lining up; make sure it is a firm connection.
  • Page 27: Default Settings

    18. After monitoring is complete, remove the sensor from the patient and store or dispose of the sensor according to local laws. See the Directions for Use of the sensor. 19. Press and hold the Power Button for 2 seconds to turn the Rad-8 off (3 seconds in the Home Mode). DEFAULT SETTINGS The Rad-8 Pulse Oximeter stores two types of default values that the device automatically retains after a power cycle.
  • Page 28: Factory Default And User Configurable Settings

    User Defaults. Refer to Setup Menu Level 3 in this section to save user defaults. ** Home default setting will not change when power is turned off and on. Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 29: Successful Monitoring

    ■ Read the sensor Directions for Use for proper sensor application. MASIMO PULSE OXIMETRY SENSORS Before use, carefully read the Masimo sensor Directions for Use. Use only Masimo sensors for pulse oximetry measurements. Tissue damage can be caused by incorrect application or use of a sensor, for example by wrapping the sensor too tightly.
  • Page 30: Numeric Display - Pulse Rate

    The pulsations from intra-aortic balloon support can cause the pulse rate displayed on the Rad-8 to be significantly different than the ECG heart rate. NUMERIC DISPLAY - (PI) The perfusion index (PI) bar graph indicator provides a relative numeric indication of the pulse strength at the monitoring site.
  • Page 31: Actions To Be Taken

    ■ If possible, ensure that the sensor is placed in a location with low ambient light. Although the pulse oximeter integrated with Masimo SET technology has significant immunity to ambient light, excessive ambient light may cause readings to be incorrect.
  • Page 32: Sensor Placement

    ■ If possible, remove electrical noise sources such as electro-surgical devices or other electrical/electronic equipment. If these solutions are not possible, operate the Rad-8 on battery power, or try plugging the device into a different electrical outlet. ■...
  • Page 33: Low Battery Audible Alarm

    When a low battery condition occurs, immediately discontinue patient monitoring and plug the Rad- into AC power. The AC Power Indicator on the Rad-8 illuminates and remains illuminated while the battery is charging, however, the Battery Charge Level Indicator does not illuminate. Once the battery is fully charged all Battery Charge Level Indicators illuminate green when unplugged.
  • Page 34: Normal Patient Monitoring

    FRONT PANEL CONTROL OPERATION SETUP MENU This section gives an overview of the Rad-8 menu selections available. To access the menu levels and navigate through the menu selections, use the front panel buttons, Enter Button and Up/Down Buttons as indicated in the following sections. Sub-sections describe each menu item in more detail.
  • Page 35: Led Brightness

    Press Up 2x Alarm Volume Level 1 let the device time out for 10 seconds to exit without saving the new setting. Press Up 3x Alarm Volume Level 2 4-10 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 36: Alarm Silence

    Press Up 2x Alarm Delay 15 seconds let the device time out for 10 seconds to exit without saving Press Up 3x the new setting. Alarm Delay 0 seconds Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-11...
  • Page 37: Clear Trend

    SETUP MENU LEVEL 2 (CONTINUED) CLEAR TREND The Rad-8 only stores data in the trend memory while the device is turned on. Trend data saves to the memory until the memory is full or cleared by the user. NOTE: It is recommended that you clear the trend prior to performing a new patient data collection procedure.
  • Page 38: Sensitivity

    Press Up Once FastSat On let the device time out for 10 seconds to exit without saving the new setting. Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-13...
  • Page 39: Setup Menu Level 3 (Enter Button + Down Button Menu Settings)

    NOTE: If the device is in the 16 seconds Sleep Mode, averaging time automatically reverts to 2 Press Up 5x Averaging Time seconds. 2 seconds Press Up 6x Averaging Time 4 seconds 4-14 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 40: Rapid Desat Limit

    **When Alarm On/Off is set to "Off rE", the audible alarm "beeps" twice every three minutes to remind the user that the Rad-8 is currently in alarm status but the audible alarm is muted. Visual alarms are active in this mode. If an alarm limit is violated, the associated parameter/measurement label and value flash, the alarm bell flashes red for high priority alarms.
  • Page 41: Default Settings

    SmartTone Press Up Once let the device time out for 10 (Smart Tone) seconds to exit without saving the new setting. 4-16 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 42: Year

    Press Up once Month and press Enter let the device time out for 10 (defaults to Yes) seconds to exit without saving the new setting. Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-17...
  • Page 43: Day

    Press Up Once Hour and press Enter (defaults to Yes) let the device time out for 10 seconds to exit without saving the new setting. 4-18 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 44: Minute

    Software Version: Press Up Once SET Software let the device time out for 10 seconds to exit without saving the new setting. Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-19...
  • Page 45: Serial Output

    When the Interface Alarms feature is set to “Off”, SpO and BPM audible alarms are muted at the Rad-8. Visual alarms are active in this mode, and the Alarm Bell indicator light flashes. Once the SpO and/or BPM alarm condition is resolved and there are no other system or parameter/measurement alarms, the Alarm Bell stops flashing.
  • Page 46: Nurse Call

    Polarity to the next menu option Normal let the device time out for 10 Press Up 2x Nurse Call seconds to exit without saving Polarity the new setting. Inverse Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-21...
  • Page 47: Set Mode

    Philips/Agilent monitor. Please see the Philips/Agilent Operator’s manual for complete instructions. 6. The Rad-8 Pulse Oximeter can be set up to audibly indicate all patient alarms while communicating with the Philips/ VueLink module. Use the Interface Alarms setting in the Output menu to enable and disable audible alarms on the Rad-8.
  • Page 48: Trend Setup And Use

    T r e n d S e t u p a n d U s e INTRODUCTION The Rad-8 can store up to 72 hours of trend data captured at 2 second intervals. The trend data can then be transferred to a PC for evaluation.
  • Page 49: Trend Data Format

    200 = reserved 400 = Low Signal IQ 800 = Masimo SET. This flag means the algorithm is running in full SET mode. It requires a SET sensor and needs to acquire some clean data for this flag to be set SAMPLE TREND OUTPUT 09/21/08 09:56:08 SpO2=000 PR=000 PI=00.00 EXC=820:OffPat,SET...
  • Page 50: Home Mode Operation

    CAUTION: ALARMS ARE DISABLED IN THIS MODE. NOTE: If the unit is connected to a Patient Safety Net system, there will be no communication with the Patient Safety Net in Home and Sleep mode. Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-25...
  • Page 52: Alarm Identification

    Sensor off and no sensor Low battery High saturation CAUTION: FOR HOME USE, ENSURE THAT THE RAD-8 ALARM CAN BE HEARD FROM OTHER ROOMS IN THE HOUSE ESPECIALLY WHEN NOISY APPLIANCES SUCH AS VACUUM CLEANERS, DISHWASHERS, CLOTHES DRYERS, TELEVISIONS, OR RADIOS ARE OPERATING.
  • Page 53: Alarm Limits

    When a sensor is not connected to its cable, “NO SEN” message will show on the display. NOTE: An audible alarm will accompany the visual indicators unless the Rad-8 has been set to Interface Alarms "Off" (only SpO and BPM alarms muted) or to Sleep Mode (all alarms muted).
  • Page 54: Alarm Silence

    Alarm Off (re): Audible alarms are muted, but the alarm “beeps” twice every three minutes to remind the user that the Rad-8 is currently in an alarm status but that the audible alarm is muted. Visual alarms are active in this mode. If the alarm is violated, the associated parameter/measurement label and value flash, and the alarm bell flashes red for high priority alarms.
  • Page 55: Messages

    MESSAGES The Rad-8 will indicate other data or system errors. Message conditions are as follows: DISPLAY TYPE SOLUTION Wait for found pulse. (This search Calibration and “---”) should occur whenever a sensor is...
  • Page 56 SINGLE BATTERY LEVEL Battery level Connect device to AC Power to INDICATOR FLASHES too low charge the battery. (WITH AUDIBLE ALARM) No Cable Connect appropriate cable to unit. Connected Defective cable Replace cable Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 58: Troubleshooting

    T r o u b l e s h o o t i n g The following chart describes what to do if the Rad-8 system does not operate properly or fails.
  • Page 59 Verify proper sensor for patient size. or sensor measurement Verify proper sensor site. location. Connect the AC Power Cord to the Rad-8 Low battery/not plugged and to an AC outlet. Make sure that the into AC power supply. AC Power Indicator light is on.
  • Page 60: Rad-8 Specifications

    41°F to 104°F (5°C to 40°C) Transportation/Storage Temperature: -40°F to 158°F (-40°C to +70°C) Storage Humidity: 5% to 95%, non-condensing 500 mbar to 1060 mbar pressure, Operating Altitude: -1000 ft to 18,000 ft (-304 m to 5,486 m) Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 61: Physical Characteristics

    0.5 and 4.25 kgs. Seventy-nine (79) data samples were collected over a range of 70 - 100% SaO2 with a resultant accuracy of 2.9% SpO 2. . Contact Masimo for testing specifications. only applies to LNOP ® Blue 2.
  • Page 62: Serial Interface Specifications

    68% of the population. 5 Masimo SET technology has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2 simulator and Masimo’s simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5% for saturations ranging from 70 to 100%.
  • Page 63: Serial Printer Setup

    NURSE CALL The nurse call feature on the Rad-8 Pulse Oximeter is based on the relay closing or opening depending on alarm, Low Signal IQ events or both. In addition the nurse call polarity can be inverted to accommodate various nurse call stations requirements.
  • Page 64: Introduction

    This section covers the use and cleaning of Masimo sensors and patient cables. Before use of any sensor, carefully read the sensor’s Directions for Use. Use only Masimo sensors and cables with the Rad-8 Pulse Oximeter. Other transducers, sensors and cables may affect Rad-8 performance.
  • Page 65: Masimo Spo Sensors

    DIRECT CONNECT SENSORS Masimo Red Direct sensors can be used with the Rad-8 to enable measurement of SpO and pulse rate only. Red Direct Connect sensors will only function with oximeter devices equipped with Masimo SET technology. Red Direct Connect sensors connect to the device directly.
  • Page 66: Lnop ® Reusable Sensors

    ± 3 bpm 70 - 100% ± 3.3% ± 3 bpm ± 3% ± 3 bpm LNOP Blue 2.5 - 30 kg 80 -100% ± 3% ± 3 bpm ± 3% ± 3 bpm Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 67: Lncs ® Reusable Sensors

    ± 3 bpm No Motion Pulse Rate Accuracy, Motion ± 5 bpm ± 5 bpm SpO 2 ± 2% SpO 2 ± 2% Low Perfusion Accuracy Pulse ± 3 bpm Pulse ± 3 bpm Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 68: Sensor Accuracy

    Complete accuracy specifications are located in the sensor Directions For Use (DFU) and are specific for the type of Masimo sensor used. CLEANING AND REUSE OF MASIMO REUSABLE SENSORS AND CABLES Reusable sensors and patient cables can be cleaned per the following procedure: 1.
  • Page 70: Introduction

    C l e a n i n g The outer surface of the Rad-8 Pulse Oximeter can be cleaned with a soft cloth dampened with a mild detergent and warm water solution. Do not allow liquids to enter the interior of the instrument.
  • Page 71: Battery Service

    To test the performance of the Rad-8 after repairs or during routine maintenance, follow the procedure outlined in this section. If the Rad-8 fails any of the described tests, discontinue its use and correct the problem before returning the device back to the user.
  • Page 72 TESTING THE RAD-8 WITH MASIMO SET TESTER (OPTIONAL): 1. Turn the Rad-8 off and then on again. 2. Connect the Masimo SET Tester to the Pulse Oximeter Patient Cable Connector. 3. Verify that within 20 seconds all available pulse bars display.
  • Page 73: Service And Repair

    Section 9, Cleaning. Make sure the equipment is fully dry before packing. To return the Rad-8 Pulse Oximeter for service, please follow the Return Procedure. RETURN PROCEDURE Please clean contaminated/dirty equipment before returning and make sure it is fully dry before packing the equipment.
  • Page 74: Sales & End-User License Agreement

    W a r r a n t y Masimo warrants to the initial Purchaser for a period of one (1) year from the date of purchase that: (i) each new Product and the Software media as delivered are free from defects in workmanship or materials, and (ii) the Product and Software will perform substantially as labeled in the directions for use.
  • Page 75: End-User License

    2. Ownership of Software: The Software is licensed not sold; all rights and interests in the Software and all copies thereof remain at all times vested in Masimo, and do not pass to Purchaser. Any references in this Agreement to the purchase or sale of the Software shall be deemed the purchase or sale of a Software License as set forth herein.
  • Page 76: Restrictions

    License, in whole or in part, by operation of law or otherwise without Masimo's prior written consent; except that the Software and all of Purchaser’s rights hereunder shall transfer automatically to any party that legally acquires title to the Product with which this Software is included.
  • Page 78: Accessories

    Rad-8 Operators Manual, Spanish 33140 Rad-8 Operators Manual, Swedish 33138 Rad-8 Operators Manual, Dutch 33139 Rad-8 Operators Manual, Danish 33141 Rad-8 Operators Manual, Portuguese 33142 Rad-8 Operators Manual, Chinese 33143 Rad-8 Operators Manual, Japanese Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 10-1...
  • Page 79 Instruments and sensors containing Masimo SET technology are identified with the Masimo SET logo.