Extech Instruments HD750 User Manual

Extech Instruments HD750 User Manual

Heavy duty differential pressure manometer
Hide thumbs Also See for HD750:

Advertisement

USER GUIDE 
 
 
 
 
 
Heavy Duty 
Differential Pressure Manometer 
 
Model HD750 
 

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HD750 and is the answer not in the manual?

Questions and answers

Summary of Contents for Extech Instruments HD750

  • Page 1 USER GUIDE            Heavy Duty  Differential Pressure Manometer    Model HD750   ...
  • Page 2: Meter Description

    Introduction  Congratulations on your purchase of the Extech Manometer Model HD750. This device measures  gauge and differential pressure in the range of ±5.000 psi. The HD750 offers eleven (11) units of  measure (user selectable by front panel pushbutton) and has a differential input that uses  convenient quick‐disconnect fittings. Advanced features include DATA HOLD, MIN‐MAX‐AVG  RECORD mode, ZERO/OFFSET, AUTO POWER OFF, and USB PC interface. This meter is shipped fully  tested and calibrated and, with proper use, will provide years of reliable service. Please visit our  website (www.extech.com) to check for the latest version of this User Guide, Product Updates, and  Customer Support.    Meter Description  P1 input    AC adaptor jack  P2Input   LCD Display – Indicates the measurement data, unit of measure,  elapsed timer, and function symbols as described in this manual  USB PC interface jack (3.5mm) – For use with data acquisition  software  UNIT button – Press to select the unit of measure  DIF button– Press to display differential pressure. Also used to  OFFSET the displayed readings.  RECORD button– Press to access the MIN/MAX/AVG recording  mode  HOLD / ZERO button– Press to freeze the displayed reading. Also  used to zero the display (press and hold until display shows all  zeroes).  10. BACKLIGHT  button – Press to switch the display backlight ON.  The backlight will automatically turn OFF after 40 seconds.  11. POWER button – Press to turn the meter ON or OFF  12. Battery compartment (on rear)  HD750-EU-EN V1.7 7/13...
  • Page 3 Display  USB data output active  Differential Pressure Mode  Data Hold Mode  Primary display  Record Mode  MAX/MIN/AVG indicators  Low Battery Indicator  Pressure unit of measure  Elapsed Timer    HD750-EU-EN V1.7 7/13...
  • Page 4: Basic Operation

    Press the UNIT button to select the unit of measure. This meter offers eleven (11) units of  measure as listed below.    Unit of Measure  Display icon  Pounds per square inch  psi  Inches of Mercury  inHg  Milli‐bar  mbar   Millimeters of Mercury  mmHg  Kilo‐Pascals  kPa  Kilograms per centimeter squared  Kgcm   Inches of water  O  Feet of water  O  Centimeters of water  O  Bars  Bar  Ounces per square inch  ozin     Fittings Connection  Connect tubing to the P1, P2 (or both) input port(s). If both inputs are used (differential mode),  the meter displays a positive pressure reading if the P1 pressure is greater than P2 and a  negative reading if P2 is greater than P1.        HD750-EU-EN V1.7 7/13...
  • Page 5 Zero Adjust and Offset function  The unit must be kept in the upright position when ZEROING or when taking measurements. To  ZERO the meter: Before use and without fittings attached to the meter, press and hold the HOLD  button for 3 seconds or until the display shows all zeroes.    To OFFSET meter readings: Press the DIF button while taking a measurement. Subsequent  readings will represent the difference between the current measurement and the measurement  that was on the display when the DIF button was pressed.    MIN‐MAX‐AVG Record Mode with Elapsed Timer  The RECORD function tracks the maximum (MAX), minimum (MIN), and average (AVG) readings  with a time stamp. The elapsed timer is displayed on the lower right‐hand corner of the LCD and  informs the user of the time a MAX or MIN reading was recorded.  Press the RECORD button once. The REC indicator will appear on the display and the  displayed elapsed timer will start (HH:MM:SS format [hours:minutes:seconds]).  Press the RECORD button again to view the highest (MAX) reading recorded since the  RECORD button was first pressed. The MAX indicator and maximum reading will appear on  the display. The elapsed timer will show the time of the MAX event.  Press the RECORD button again, the MIN indicator and the minimum value will appear on  the display. The elapsed timer will show the time of the MIN event.  Press the RECORD button again, the AVG indicator and the average value will appear on the  display. The elapsed timer will continue to run.  To return to normal operation, press and hold the RECORD button until the display  indicators REC, MAX, etc. switch off.  Note that in the RECORD mode, the backlight and the power buttons are the only available  functions.  USB PC Interface  The meter has a built‐in USB port for use with the Manometer Data Acquisition software. The  software allows the user to view, save, export, and print readings from the HD750. Refer to the  documentation included with the software for complete details.  Displayed Error Codes  The following codes are displayed when the meter encounters an error.  ERR 1: Pressure input exceeds specified limits  ERR 2: Pressure input is too low to read  ERR 3: The Differential Pressure input exceeds specified limits  ERR 4: The Differential Pressure input is too low to read  HD750-EU-EN V1.7 7/13...
  • Page 6: Specifications

    0.5s typical  Input circuit  Differential inputs (P1 and P2)  Input Ports  Two metal 4.8mm, barbed for 1/8" (3.17mm) ID tubing  Data Hold  Freezes displayed reading  Sensor  Built‐in piezoelectric sensors  Zero / Offset function  Pushbutton activation  Sample rate  0.5 sec. (approx.)  Compatibility  Air or non‐corrosive and non‐ionized gas (not intended for use with  liquids)  Auto Power OFF  Meter turns off after approx. 20 minutes to conserve energy (can  be deactivated)  PC Interface  Built‐in USB serial interface  Storage conditions  0 to 60 C (14 to 140 F); < 80% RH  Operating conditions  0 to 50 C (32 to 122 F); < 80% RH  Battery power  9V alkaline battery (BAT low battery indicator)  Battery life  125 hours typical  Dimensions  210 x 75 x 50mm (8.2 x 2.9 x 1.9”)  Weight  280g (9.8oz)  HD750-EU-EN V1.7 7/13...
  • Page 7 0.01  ± 138.3inches H O  0.1  ± 351.5cm H O  0.1  ± 80.00oz in   0.01    Warning: Pressure inputs of 10 psi or higher will damage the instrument. Do not apply pressure higher than the maximum pressure listed in the specification table HD750-EU-EN V1.7 7/13...
  • Page 8: Battery Replacement

    Disposal: Follow the valid legal stipulations in respect of the disposal of the device at the  end of its lifecycle    Other Battery Safety Reminders   Never dispose of batteries in a fire. Batteries may explode or leak.   Never mix battery types. Always install new batteries of the same type.      Conversions  Useful conversion Factors  From  To  Multiplier  in of H O  in of Hg  0.07355  in of H O  cm of H O  2.54  mm of Hg  in of H O  0.03937          Copyright © 2013 FLIR Systems, Inc.  All rights reserved including the right of reproduction in whole or in part in any form  ISO‐9001 Certified  www.extech.com    HD750-EU-EN V1.7 7/13...

Table of Contents