DIAGnostics:INSTrument:PRINter:ADDRess <num>
l
DIAGnostics:INSTrument:PRINter:ADDRess?
l
Sets the HP-IB address of the printer to use during some of the
calibration procedures when the synthesizer assumes HP-IB control.
*RST and power on do not effect this command. The default is 1.
The default value is set at memory initialization only.
DIAGnostics:IORW <num>,<num>
l
Performs a write to the I/O Device number specified in the first
<num> and sets it to the value in the second <num>.
DIAGnostics:IORW? <num>
l
Reads from the specified I/O device number and returns the response
data.
DIAGnostics:OUTPut:FAULt?
l
Returns a string of 16, 1s and OS that are equivalent to the fault
display.
Bit 0 = PEAK
Bit 1 = TRACK
Bit 2 = RAMP
Bit 3 = SPAN
Bit 4 = V/GHz
Bit 5 = ADC
Bit 6 = EEROM
Bit 7 = PWRON
Bit 8 = CALCO
Bit 9 = PLLZERO
Bit 10 = PLLWAIT
Bit 11 = FNXFER
Bit 12 = CAL YO
Bit 13 = CAL MAN
Bit 14 = TMR CNFLCT
Bit 15 = SEARCH
DIAGnostics:RESult?
l
Returns the following information:
<result>, <test failure>, <manual entry point>
where, <result> is one of the following:
0
Diagnosis successful.
1
Cannot diagnose; full selftest must be executed first.
2
No failures found-all selftests passed.
- 1
Cannot diagnose; diagnosis routine failed to isolate failure.
(software fault)
<test failure> is the test number of the most relevant failure
(-999 if parameter is not used, as in <result> of 1).