Venting the liquid dispensing system / Maintenance

Venting the liquid dispensing system / Maintenance
Venting the liquid dispensing
system
Before using the machine for the first
time, or if the container had not been
filled in time and allowed to empty, the
dispensing system for liquid agents
(apart from rinsing agent) must be
vented.
Proceed as follows:
Press the I-0 (ON-Off) button.
A
B
Switch "A":
Dispensing system for neutralising
agent.
Switch "B":
Dispensing system for detergent (if
using liquid detergent).
Use a pointed object to press in
switch "A" and/or "B" at the left hand
side end of the control panel for
three and a half minutes.
Each time the system is vented any liq-
uid which runs into the wash cabinet
must be washed out as follows:
Select programme "A" (Cold).
After this has run the dispensing sys-
tem is filled and ready for use.
Checking the dispensing
system
To maintain the performance of the dis-
pensing system, it should be serviced
at regular intervals.
This work should only be carried out by
suitably qualified people or by a MIELE
Service Engineer.
Every 12-18 months
– Replace the dispensing hoses in
either the dispensing systems or the
DOS modules.