HP 53131A/132A 225 MHz Programming Manual page 204

225 mhz universal counter
Table of Contents

Advertisement

Chapter 4 Command Reference
:DISPlay Subsystem
:DISPlay[:WINDow]:TEXT:FEED "[:]CALCulate2" |
"[:]CALCulate3"
Sets or queries what data flow is fed into the display.
Choose from the following <data_handle> strings:
Query Response
A string is returned: "CALC2" or "CALC3."
Comments
Related
Stats
Front-Panel
Keys
4-38
"[:]CALCulate2"— should be used to direct any result other than the statistics
to the result display
Specifically this would select one of the following results for the result
display:
– raw measurement (if math is disabled— :CALC:MATH:STAT OFF)
– the scaled/offset measurement (if math is enabled— :CALC:MATH:STAT
ON)
– the limit graph (if limit testing is enabled with the graphic display—
:CALC2:LIM:STAT ON, :CALC2:LIM:DISP GRAPh)
"[:]CALCulate3"— should be used to direct the statistical result
(if statistics are enabled, :CALC3:AVER[:STAT] ON) to the result display;
the particular statistic displayed is determined by :CALC3:AVER:TYPE.
*RST: ":CALCulate2"
Refer to the section titled "How to Program the Counter to Display Results"
in Chapter 3 of this guide.