Chapter 4 Command Reference

:DIAGnostic Subsystem

:DIAGnostic Subsystem
This subsystem controls the remote calibration of the Counter.
All of the calibration values, with the exception of the interpolator values, are
stored in non-volatile memory and are unaffected by
power-on, save/recall, and *RST.
Any of the commands which perform a calibration, with the exception of the
interpolator calibration, will generate error -221 if the user tries to execute a
calibration while the Counter is secured. (Note, this will not occur in early
revisions of the Counter because calibration security does not exist.) Please refer
to the :CALibration:SECurity subtree for command specifics regarding calibration
security.
:DIAGnostic:CALibration:INPut[1|2]:GAIN:
AUTO ONCE | OFF
Calibrates the channel 1 or 2 input trigger GAIN when the ONCE parameter is
used.
Before sending this command, connect a +5V source to the appropriate input.
Query Response
A sequence of ASCII-encoded bytes: OFF
Comments
Related
Scale and Offset/POWER (Calibration menu).
Front-Panel
Keys
4-32
The calibration values are stored in non-volatile memory, and are unaffected
by power-on, save/recall, and *RST.
Use :DIAG:CAL:STAT? to check for successful calibration.
After calibration is completed, the state of this command' s parameter is OFF.
Programming Guide