GE LOGIQ P7 Service Manual
Hide thumbs Also See for LOGIQ P7:
Table of Contents

Advertisement

P R E L I M I N A R Y
GE Healthcare
LOGIQ P7/P8/P9/P10
Service Manual
COPYRIGHT GENERAL ELECTRIC COMPANY.
ADVANCED SERVICE DOCUMENTATION.
COPYRIGHT GENERAL ELECTRIC COMPANY.
GE CONFIDENTIAL. UNLICENSED USE BY NON-GE PARTIES IS
GE RESTRICTED. DISCLOSURE TO NON-GE PARTIES
© 2022 General Electric Company.
Document Number:
(MAY 24, 2022)
VE6_VE8_COVER.FM
BASIC SERVICE DOCUMENTATION.
STRICTLY PROHIBITED.
PROPRIETARY TO GE.
PROHIBITED.
All Rights Reserved.
5604324
Revision:
18

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LOGIQ P7 and is the answer not in the manual?

Questions and answers

Summary of Contents for GE LOGIQ P7

  • Page 1 Service Manual BASIC SERVICE DOCUMENTATION. COPYRIGHT GENERAL ELECTRIC COMPANY. ADVANCED SERVICE DOCUMENTATION. COPYRIGHT GENERAL ELECTRIC COMPANY. GE CONFIDENTIAL. UNLICENSED USE BY NON-GE PARTIES IS STRICTLY PROHIBITED. PROPRIETARY TO GE. GE RESTRICTED. DISCLOSURE TO NON-GE PARTIES PROHIBITED. © 2022 General Electric Company.
  • Page 3 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL Important Precautions Translation policy...
  • Page 4 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL...
  • Page 5 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL...
  • Page 6 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL...
  • Page 7 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL...
  • Page 8 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL...
  • Page 9 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL...
  • Page 10 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL viii...
  • Page 11 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL...
  • Page 12 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL...
  • Page 13 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL...
  • Page 14 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL...
  • Page 15 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL xiii...
  • Page 16 GE personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
  • Page 17 The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE. GE may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
  • Page 18 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL Revision History Revision Date Reason for change October 2014 Initial release December 2014 Enhancement feedback May 2015 Enhancement feedback March 2016 Initial release for R2 August 2016 Correction November 2016 Initial release for R2.5...
  • Page 19: Table Of Contents

    LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL Table of Contents CHAPTER 1 Introduction Overview ............1 - 1 Purpose of Chapter 1 .
  • Page 20 Electrical Requirements ........2 - 2 LOGIQ P7/P8/P9/P10 Power Requirements ....2 - 3 Inrush Current .
  • Page 21 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL CHAPTER 3 Setup Instructions Overview............3 - 1 The Purpose of Chapter 3 .
  • Page 22 Dataflow Examples ........3 - 62 Stand-alone LOGIQ P7/P8/P9/P10 ......3 - 63 LOGIQ? + PC within a “Sneaker Net”...
  • Page 23 ANUAL You will need one network cable to connect the LOGIQ P7/P8/P9/P10 to a wall outlet on the hospital’s network..........3 - 64 Connection using Wireless Option .
  • Page 24 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL CHAPTER 4 Functional Checks Overview ............4 - 1 Purpose of Chapter 4 .
  • Page 25 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL General Information ........4 - 33 B-Mode .
  • Page 26 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL CHAPTER 5 Components and Functions (Theory) Overview ............5 - 1 Purpose of Chapter 5 .
  • Page 27 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL Service Platform ........... 5 - 31 Introduction .
  • Page 28 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL CHAPTER 6 Service Adjustments Overview ............6 - 1 Purpose of Chapter 6 .
  • Page 29 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL CHAPTER 7 Diagnostics/Troubleshooting Overview............7 - 1 Purpose of Chapter 7 .
  • Page 30 GE Icon and Remote Connection ....... . 7 - 19...
  • Page 31 LOGIQ P7/P9 R2.5 to R3 upgraded system ........
  • Page 32 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL Settings - Backup and Restore ........8 - 69 General Information .
  • Page 33 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL Tools ............8 - 84 Removal Procedure .
  • Page 34 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL Replacement of Fixed Arm Cover and Fixed Arm Holder Grip ....8 - 105 Manpower ..........8 - 105 Tools .
  • Page 35 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL Replacement of Key Caps (by special native language keys) ....8 - 131 Manpower ..........8 - 131 Tools .
  • Page 36 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL Manpower ..........8 - 168 Tools .
  • Page 37 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL Note on EEPROM Replacement ....... . 8 - 183 Accessing EEPROM .
  • Page 38 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL Manpower ..........8 - 197 Tools .
  • Page 39 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL Manpower ..........8 - 215 Tools .
  • Page 40 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL Replacement of Front Fan Filter ........8 - 225 Manpower .
  • Page 41 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL Removal Procedure - R4/R4.5 ......8 - 233 Installation Procedure .
  • Page 42 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL Installation Procedure ......... 8 - 257 Functional Check .
  • Page 43 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL Check the IOBOX (or IOBOX-B) and replaced ACDC serial number ..8 - 283 Check the IOBOX (or IOBOX-B) and replaced ACDC serial number ..8 - 284...
  • Page 44 LOGIQ P7/P8/P9/P10 Software........
  • Page 45 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL CHAPTER 10 Care & Maintenance Overview............10 - 1 Preventive Maintenance Inspections .
  • Page 46 LOGIQ P7/P8/P9/P10 5604324, R IRECTION EVISION ERVICE ANUAL Grounding Continuity ......... 10 - 26 Meter Procedure .
  • Page 47: Overview

    Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the LOGIQ P7/P8/P9/P10 ultrasound system. The service provider must read and understand all the information presented in this manual before installing or servicing a unit.
  • Page 48: Typical Users Of The Basic Service Manual

    5560938 LOGIQ P9 5641296 LOGIQ P7 with High Cabinet 5641037 LOGIQ P9 with Articulating Monitor Arm 5724885 LOGIQ P7 R2 with High Cabinet 5724886 LOGIQ P9 R2 with Articulating Monitor Arm 5759916 LOGIQ P7 R2.5 5759915 LOGIQ P9 R2.5 5794264...
  • Page 49: Purpose Of Operator Manual(S)

    EVISION ERVICE ANUAL 1-1-5 Purpose of Operator Manual(s) The Operator Manual(s) should be fully read and understood before operating the LOGIQ P7/P8/P9/ P10 and also kept near the unit for quick reference. Section 1-2 Important Conventions 1-2-1 Conventions Used in this Manual...
  • Page 50: Standard Hazard Icons

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
  • Page 51: Product Icons

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. Table 1-5 Product Icons LABEL/SYMBOL...
  • Page 52 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 1-5 Product Icons (Continued) LABEL/SYMBOL PURPOSE/MEANING LOCATION “ATTENTION - Consult accompanying documents " is intended to alert the user to refer to the operator manual or...
  • Page 53 Unique Device Identification (UDI) Label. The UDI label consists of a series of alpha-numeric characters and barcode, which uniquely identify the LOGIQ P7/P9 system as a medical device manufactured by General Electric. Scan or enter the UDI information into the patient health record as required by country-specific laws.
  • Page 54 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 1-5 Product Icons (Continued) LABEL/SYMBOL PURPOSE/MEANING LOCATION Rating Plate Every system has a unique marking for identification, the Unique Device Identification (UDI) Label. The UDI label...
  • Page 55 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 1-5 Product Icons (Continued) LABEL/SYMBOL PURPOSE/MEANING LOCATION Rating Plate Every system has a unique marking for identification, the Unique Device Identification (UDI) Label. The UDI label consists of a series of alpha-numeric characters and barcode, which uniquely identify the LOGIQ P8/P9/P10 R4.5 system as...
  • Page 56: Safety Considerations

    Servicing should be performed by authorized personnel only. Only personnel who have participated in a LOGIQ P7/P8/P9/P10 Training are authorized to service the equipment. 1-3-3 Mechanical Safety The LOGIQ P7/P8/P9/P10 weighs 60 kg or more, depending on installed peripherals, (133 lbs CAUTION or more) when ready for use. !! CAUTION: Care must be used when moving it or replacing its parts.
  • Page 57: Electrical Safety

    • Never immerse the probe connector or adapter into any liquid. NOTE: For detailed information on handling probes, refer to the LOGIQ P7/P8/P9/P10 Basic User Manual and the care card supplied with the probe. Chapter 1 - Introduction 1-11...
  • Page 58: Auxiliary Devices Safety

    The ultrasound system is an extremely sensitive and complex medical system. Any unauthorized peripherals may cause system failure or damage! The LOGIQ P7/P8/P9/P10 may be used with an isolation transformer to provide the required separation from mains for both, the system and the auxiliary devices.
  • Page 59 After each installation, the leakage currents have to be measured according to WARNING WARNING IEC 60601-1 respectively ANSI/AAMI ES60601-1. !! WARNING: All peripherals mounted on the LOGIQ P7/P8/P9/P10 system chassis must be firmly secured in NOTICE position. !! NOTICE: Chapter 1 - Introduction...
  • Page 60: Battery Safety

    1-3-6 Battery Safety NOTE: The LOGIQ P7/P8/P9/P10 ultrasound system is supplied with a lithium ion battery in the power supply module, as option. The lithium ion technology used in the system’s battery is significantly less hazardous to the environment than the lithium metal technology used in some other batteries (such as watch batteries).
  • Page 61: Labels Locations

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 1-3-7 Labels Locations The LOGIQ P9/P7 ultrasound system comes equipped with product labels and icons. These labels and icons represent pertinent information regarding the operation of the unit.
  • Page 62 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 1-2 Label Location (R3) * Required for Asia. 1) LCD Caution Label 2) Gender Caution Label (Only for India, China, Korea) 3) Multi Caution Label 4) LOGIQ P9/P7 Rating Label (For China, Korea, Japan)
  • Page 63 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 1-3 Label Location (R4) 1) LCD Caution Label 2) Gender Caution Label (Only for India, China, Korea) 3) Multi Caution Label 4) LOGIQ P8/P9/P10 Rating Label (For China, Korea, Japan)
  • Page 64 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 1-4 Label Location (R4.5) 1) LCD Caution Label 2) Gender Caution Label (Only for India, China, Korea) 3) Multi Caution Label 4) LOGIQ P8/P9/P10 Rating Label (For China, Korea, Japan)
  • Page 65: Dangerous Procedure Warnings

    AC Outlet. 1-3-10 Returning/Shipping System, Probes and Repair Parts When returning or shipping the LOGIQ P7/P8/P9/P10 system in the original packaging: • system must be lowered to its minimum height with monitor flapped down (see Figure on...
  • Page 66: Electromagnetic Compatibility (Emc)

    EMC also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and disturbances in the electrical power supply. For applicable standards please refer to Chapter 2 in the Basic User Manual of the LOGIQ P7/P8/P9/ P10 ultrasound system. 1-4-2...
  • Page 67: Customer Assistance

    Latin America Applications support 1-262-524-5698 Europe OLC - EMEA (Europe, Middle East & Africa) GE Ultraschall Deutschland GmbH Phone: +49 (0) 212 2802 - 652 (-OLC) Beethovenstraße 239 +33 1 3083 1300 (English/German all segments incl. training) Postfach 11 05 60, D-42655 Solingen...
  • Page 68 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 1-22 Section 1-5 - Customer Assistance...
  • Page 69: Overview

    2-1-1 Purpose of Chapter 2 This chapter provides the information required to plan and prepare for the installation of a LOGIQ P7/ P8/P9/P10 ultrasound unit. Included are descriptions of the facility and electrical needs to be met by the purchaser.
  • Page 70: General Console Requirements

    !! CAUTION: 2-2-1-1 Cooling The cooling requirement for the LOGIQ P7/P8/P9/P10 is 680 BTU/hr. This figure does not include cooling needed for lights, people, or other equipment in the room. NOTE: Each person in the room places an additional 300 BTU/hr. demand on the cooling system.
  • Page 71: Logiq P7/P8/P9/P10 Power Requirements

    The LOGIQ P7/P8/P9/P10 ultrasound system is supplied with a main power plug, as standard. In the event that the unit arrives without a power plug, or with the wrong plug, contact your GE dealer. When necessary, the installation engineer will supply the appropriate power plug to meet the applicable local regulations.
  • Page 72: Emi Limitations

    Ultrasound machines also generate EMI. The LOGIQ P7/P8/P9/P10 complies with limits as stated on the EMC label. However, there is no guarantee that interference will not occur in a particular installation.
  • Page 73: Probe Environmental Requirements

    Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six weeks before delivery, for enough time to make necessary changes. Have two people available to deliver and unpack the LOGIQ P7/P8/P9/P10 Ultrasound System. CAUTION Attempts to move the unit considerable distances (or on an incline) by one person alone, could result in personal injury and/or damage to the system.
  • Page 74: System Specifications

    ANUAL 2-2-6 System Specifications 2-2-6-1 Physical Dimensions of LOGIQ P7/P8/P9/P10 The physical dimensions and weight (without Peripherals) of the LOGIQ P7/P8/P9/P10 unit are summarized in Table 2-6. NOTE: Physical dimensions (especially height and depth) depend on control console and monitor positioning.
  • Page 75: Facility Needs

    The products involved (and the accompanying electrical installations) are highly sophisticated and special engineering competence is required. All electrical work on these products must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
  • Page 76: Mandatory Site Requirements

    NOTE: All relevant preliminary network port installations at the prepared site must be performed by authorized contractors. The purchaser of GE equipment must utilize only qualified personnel to perform servicing of the equipment. 2-3-3 Site Recommendations The following are (optional) site recommendations.
  • Page 77: Recommended Ultrasound Room Layout

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2-3-3-1 Recommended Ultrasound Room Layout Figure 2-1 below shows a floor plan illustrating the recommended layout of the Ultrasound Room and depicting the minimal room layout requirements.
  • Page 78: Networking Setup Requirements

    With DICOM, images can be archived, stored, and retrieved faster, easier, and at a lower cost. 2-3-4-4 DICOM Option Pre-installation Requirements To configure the LOGIQ P7/P8/P9/P10 ultrasound unit to work with other network connections, the network administrator must provide some necessary information. Use the Agent Configuration (R3, R4 and R4.5) on page 3-75...
  • Page 79: Overview

    Setup Instructions Section 3-1 Overview 3-1-1 The Purpose of Chapter 3 This chapter contains information needed to setup the LOGIQ P7/P8/P9/P10 Ultrasound System. Included are procedures to receive, unpack and configure the equipment. A worksheet is provided (see: page 3-75 page 3-81) to help ensure that all the required information is available, prior to setup the system.
  • Page 80: Set Up Reminders

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 3-2 Set Up Reminders 3-2-1 Average Installation Time Once the site has been prepared, the average installation time required is shown in Table 3-2 below. Table 3-2...
  • Page 81: Installation Warnings

    ANUAL 3-2-2 Installation Warnings 1) Since the LOGIQ P7/P8/P9/P10 weighs approximately 90 kg (200 lbs.) without peripherals, two people are required to unpack it. 2) There are no operator serviceable components. To prevent shock, do not remove any covers or panels.
  • Page 82: Brake Pedal Operation

    (When covers are removed, EMI may be present). !! CAUTION: OPERATOR MANUAL(S) CAUTION The User Manual(s) should be fully read and understood before operating the LOGIQ P7/P8/ P9/P10. Keep manuals near the unit for reference. !! CAUTION: ACOUSTIC OUTPUT HAZARD...
  • Page 83 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 3-3 Environmental Labels (R4.5) Chapter 3 - Setup Instructions...
  • Page 84: Receiving And Unpacking The Equipment

    ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment Please read this section carefully before unpacking the LOGIQ P7/P8/P9/P10 LOGIQ P9/P7 CAUTION ultrasound system and its (optional) peripherals. !! CAUTION: The LOGIQ P7/P8/P9/P10 ultrasound system, together with peripherals, probes and accessories are shipped from the factory in a single durable shipping crate which is mounted on a raised wooden platform base.
  • Page 85 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL The envelope with delivery address, packing list and invoice is located on the front panel of the crate. Check whether delivery is complete (according to packing list) and check visual damage!
  • Page 86 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 3-5 Unpacking Procedure Step Task Remove the monitor pads from the monitor and remove the 4 pads of the box bottom assy. monitor pad Carefully remove foam packing material and plastic bag from the ultrasound unit and monitor.
  • Page 87: Packing Materials - Recycling Information

    IRECTION EVISION ERVICE ANUAL Section 3-4 Packing materials - recycling information The packing materials for LOGIQ P7/P8/P9/P10 are recyclable: • The Transportation Box is made of cardboard. • The inner reinforcements are made of Ethafoam (Polyethylene foam). • The plastic foil is made of LDPE (Low Density Polyethylene).
  • Page 88: Preparing For Set Up

    Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or any deterioration. Power Cord Verify continuity. Replace the power cord, as required. NOTE: Report any items that are missing, back-ordered, or damaged, to your GE sales representative. The contact address is shown in Contact Information on page 1-21.
  • Page 89: Emi Protection

    Verify that all EMI rules listed in the following table are followed: The LOGIQ P7/P8/P9/P10 ultrasound unit is approved for use in hospitals, clinics and other environmentally qualified facilities, in terms of the prevention of radio wave interference. Operation of the ultrasound unit in an inappropriate environment can cause electronic interference to radios and television sets situated near the medical equipment.
  • Page 90: Connection Of Auxiliary Devices

    1,5 m 2,5 m 1,5 m 1,5 m NOTE: For more detailed Safety Considerations when connecting auxiliary devices to the LOGIQ P7/P8/P9/ P10 system, please review: Chapter 1 - Auxiliary Devices Safety, on page 1-12. 3-12 Section 3-6 - Connection of Auxiliary Devices...
  • Page 91: Connecting The Black & White Printer

    There are three USB ports (stack) on rear panel. The BW printer may be connected to any of the 3x USB stack connector. 3) When all the cables are connected, press the Power ON switch on the Black & White printer. 4) Power ON/Boot up the LOGIQ P7/P8/P9/P10 system as described in Section 3-7-2 on page 3-22.
  • Page 92: Connection Scheme: B&W Printer (R1 To R3)

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-6-1-1 Connection Scheme: B&W Printer (R1 to R3) Figure 3-6 B&W Printer connection (R1 to R3) 3-14 Section 3-6 - Connection of Auxiliary Devices...
  • Page 93: Connection Scheme: B&W Printer (R4 And R4.5)

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-6-1-2 Connection Scheme: B&W Printer (R4 and R4.5) Figure 3-7 Connection Scheme: B&W Printer (R4 and R4.5) Chapter 3 - Setup Instructions 3-15...
  • Page 94: Connecting The Secondary "Patient" Lcd Monitor

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-6-2 Connecting the Secondary “Patient” LCD Monitor Secondary LCD “Patient” Monitor MUST NEVER be connected to the LOGIQ P9/P7 ultrasound CAUTION systems mains supply directly! Always connect it to the supplied Isolation Transformer! !! CAUTION: The Secondary Monitor is the only item to be connected to the Transformer.
  • Page 95 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL appropriate Transformer Secondary Monitor HDMI to DVI / HDMI to HDMI cable connected to the rear side of the 2nd Monitor Composite (BNC) Cable S-Video Cable CAUTION A Secondary “Patient”...
  • Page 96: Connecting The Footswitch

    ERVICE ANUAL 3-6-3 Connecting the Footswitch The footswitch should be directly connected to any accessible USB-port on the LOGIQ P7/P8/P9/P10 (e.g., on rear of the system). NOTE: Connection of the Footswitch is always the same (no differences between PC-Motherboard version of the system).
  • Page 97: Connecting The Usb Flash Memory Stick

    Devices on page 3-20. !! NOTICE: The USB Flash Memory Stick may be connected to an accessible USB port of the LOGIQ P7/P8/P9/ P10 system. Refer to Section 3-9-2 "External I/O Connectors" An external USB Flash Memory Stick can be connected once the system is powered ON, or after shutdown.
  • Page 98: General Remarks And Hints When Using External Usb-Devices

    When an external USB-storage device (such as a USB-memory stick or an external hard disk) is connected to the LOGIQ P7/P8/P9/P10, Windows detects the device and automatically installs a driver. During this process, several dialogs may pop up, starting with the “Found New Hardware“ dialog.
  • Page 99: Completing The Set Up

    LOGIQ P7/P8/P9/P10 System User Training. Use only the power cords, cables and plugs provided by or designated by GE to connect the unit to the power source. Prior to connect the LOGIQ P7/P8/P9/P10 unit to a power source, verify compliance with all CAUTION electrical and safety requirements.
  • Page 100: Power On / Boot Up

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-7-2 Power On / Boot Up 3-7-2-1 Scanner Power On 1) Ensure that the wall outlet is of appropriate type, and that the Circuit Breaker is turned off.
  • Page 101: Back End Processor Boot Up

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-7-2-2 Back End Processor Boot Up When AC power is applied to the scanner, the ON/OFF button on the control panel illuminates blue, NOTICE indicating that the System (including the Back-end Processor) is in Standby mode.
  • Page 102: During A Normal Boot, You May Observe

    Figure 3-12 below.Start-up progress bars indicating software loading procedures, Figure 3-12 GE Healthcare wallpaper with progress bar (R1.x.x) Figure 3-13 GE Healthcare wallpaper with progress bar (R2.x.x and later) 3-24 Section 3-7 - Completing the Set Up...
  • Page 103 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 3-14 GE Healthcare wallpaper with progress bar (R3.x.x and later) Figure 3-15 GE Healthcare wallpaper with progress bar (R4.x.x and later) 5) Startup GE Chapter 3 - Setup Instructions 3-25...
  • Page 104 ANUAL NOTE: Startup GE Healthcare wallpaper with progress bar may differ depending on product type and release. 6) The software initiates and sets up the Front-End electronics and the rest of the scanner (incl. the clicking sound of the relays on the PIL board).
  • Page 105: Power Off / Shutdown

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-7-3 Power Off / Shutdown When you power off the unit, the system performs an automatic shutdown sequence Figure 3-16 System Exit Window (R2 and R2.5) Figure 3-17 System Exit Window (R3, R4 and R4.5)
  • Page 106 Cancel Use this button to exit from the System-Exit menu and return to the previous operation. Exit (Only available when logged in as GE Service with Service Dongle) • Select this button when you want to exit to the Windows Desktop.
  • Page 107: Scanner Shutdown

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-7-3-1 Scanner Shutdown 1) If not already in read mode, freeze the image. 2) Press the ON/OFF button (see: Figure 3-11) on the control panel. Following dialog appears.
  • Page 108 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 3-20 System - Exit (R4 and R4.5) 3-30 Section 3-7 - Completing the Set Up...
  • Page 109 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Select the SHUTDOWN button. The system performs an automatic full shutdown sequence. 4) Switch OFF the Circuit at the rear of the system. NOTE: The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the ON/OFF button.
  • Page 110 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5) After complete power down, disconnect the main power cable from the system or unplug it from the AC wall outlet socket. Circuit Breaker Main Power Figure 3-21 Circuit and main power at rear of system 6) Press on the brakes to block the front caster wheels.
  • Page 111: Transducer Connection

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-7-4 Transducer Connection 3-7-4-1 Connecting the Probe Inspect the probe before and after each use for damage or degradation to the housing, strain CAUTION relief, lens, seal, cable and connector. DO NOT use a transducer which appears damaged until functional and safe performance is verified.
  • Page 112: Connecting The Cw Pencil Probe

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8) Carefully position the probe cord so it is free to move and is not resting on the floor. 3-7-4-2 Connecting the CW Pencil Probe Insert the probe connector into the probe port all the way seated in. Carefully position the probe cord so it is free to move and is not resting on the floor.
  • Page 113: Printer Installation

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 3-8 Printer Installation NOTE: For Connection schemes refer to Section 3-6 "Connection of Auxiliary Devices" on page 3-12. For further installation instructions see: • Section 3-8-1 "Installing Digital Black & White Printer Sony UP-D898" on page 3-35 •...
  • Page 114: Digital Color Printer Sony Up-D25Md

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-8-2 Digital Color Printer Sony UP-D25MD 1) Power off/Shutdown the system as described in: Section 3-7-3 on page 3-27. 2) Connect USB cable into the available USB port on the rear of the system. There are three available USB ports on rear of the system.
  • Page 115: Installing The Bluetooth Deskjet Printer (R2 And R2.5)

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-8-3 Installing the Bluetooth Deskjet Printer (R2 and R2.5) 1) Power off/Shutdown the system as described in: Section 3-7-3 on page 3-27. 2) Physically connect the Bluetooth dongle as described on Section 3-8-3-1 on page 3-37.
  • Page 116: Verifying Bluetooth Device Driver

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-8-3-2 Verifying Bluetooth device driver 1) Turn on the system. 2) Enter maintenance mode by using service dongle. 3) Go to Device Manager by pressing shortcut on the taskbar 4) Check that there is “Bluetooth Radios”...
  • Page 117: Prepare The Printer And Connect The Bluetooth Printer Adapter

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-8-3-3 Prepare the Printer and connect the Bluetooth Printer Adapter 3-8-3-3-1 HP Officejet 100 1) Connect the power cord to a wall outlet socket. 2) Turn on the printer.
  • Page 118: Adding Printer To The System

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-8-4 Adding Printer to the system 1) Select Utility -> Connectivity -> Service. 2) Add the Standard Print. 3) Highlight Standard Print in the Service list. 4) Select the printer from the Printer pull-down Properties menu. For the UP-D898 printer, select “Portrait”...
  • Page 119: Up-D898 - Printer Settings

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-8-5-1 UP-D898 - Printer Settings 1) Call up the ‘Printer Preferences’; operation see: Section 3-8-5 "Adjustment of Printer Settings" 2) Select the page (see: Figure 3-29 below) and select:...
  • Page 120: Up-D25Md - Printer Settings

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-8-5-2 UP-D25MD - Printer Settings 1) Call up the ‘Printer Preferences’; operation see: Section 3-8-5 "Adjustment of Printer Settings" Settings for Paper Size MUST match with the used Paper (large/small) and also the right color ink NOTICE cartridge has to be used.
  • Page 121 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL b) Color Correction: set Printer Hardware Color Correction1 Figure 3-31 Graphics page (Color Balance + Color Correction) c) Lightness: Sharpness = 7; Dark = 0; Gamma = 14; Light = 0...
  • Page 122: Installing Network Printer (R3, R4 And R4.5)

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-8-6 Installing Network Printer (R3, R4 and R4.5) 1) Confirm that your HP Network Printer is listed on HP Universal Print Driver (UPD) supported printers from https://support.hp.com/us-en/document/ish_4952109-2831856-16.
  • Page 123 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5) Select your HP Network Printer and Click ‘OK’. Figure 3-35 Chapter 3 - Setup Instructions 3-45...
  • Page 124: Setting Printer To Print Reports

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-8-7 Setting Printer to Print Reports 3-8-7-1 Setting up the Printer to Print Reports To set up the Off-Line Printer to print reports, 1) Press Utility -> System -> Peripherals.
  • Page 125: System Configuration

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 3-9 System Configuration 3-9-1 System/General Preset Menu The System/General screen allows you to specify hospital name and system date and time. Figure 3-38 System/General Preset Menu (R1.x.x) Figure 3-39 System/General Preset Menu (R2.x.x)
  • Page 126 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 3-40 System/General Preset Menu (R3, R4 and R4.5) Table 3-8 Location Preset Parameter Description Hospital Type the institution’s name. Department Type the institution's department name. Preset Region (restart needed) Select region (None, Americas, Asia, Europe or Japan).
  • Page 127 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 3-10 General User Interface Preset Parameter Description Color Level (restart needed) Select System Color according to the condition of the room. Table 3-11 Title Bar Preset Parameter...
  • Page 128 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 3-15 Scan Assistant Preset Parameter Description Auto Selection Off, Category, or Description. • Off. The Scan Assistant selection on the Patient screen is completely manual. It will say “None”...
  • Page 129: External I/O Connectors

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-9-2 External I/O Connectors Figure 3-41 External I/O Connectors - on Rear of System Table 3-19 External I/O Connector - Description Item Connector Name Table Number Description Table 5-2 USB port.
  • Page 130 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 3-20 External lI/O Connector - Description (R1/R2/R2.5) Item Connector Name Table Number Description Table 5-2 Connected to IO Box via USB HUB. Refer to Section 5-2-4 for details.
  • Page 131: Available Probes

    DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 3-10 Available Probes 3-10-1 Supported probes The LOGIQ P7/P8/P9/P10 supports the following types of probes: • Sector Phased Array • Linear Array • Convex Array • Micro convex Array • Matrix Array •...
  • Page 132 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Part Probe Part Name Illustration Application Feature Number Type B, CHI, CF, PDI, M, PW, B-Flow, B- 4C-RS Abdomen, OB/ Flow GYN, Color, CrossXBeam, LOGIQView, Urology, Vascular...
  • Page 133 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Part Probe Part Name Illustration Application Feature Number Type RAB2-6-RS Abdomen, OB/ B, CHI, CF, PDI, M, PW, GYN, CrossXBeam, Volume Urology LOGIQView, ATO/ASO,CTO, (Real KTZ303982 SRI-HD, 3D/4D real-time imaging,...
  • Page 134 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Part Probe Part Name Illustration Application Feature Number Type Small Parts, B, CHI, CF, PDI, M, PW, B-Flow, B- L10-22-RS Musculoskeletal, Flow Neonatal Color, Virtual Convex,CrossXBeam, LOGIQView, ATO/ASO, CTO, SRI-...
  • Page 135 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Part Probe Part Name Illustration Application Feature Number Type Cardiac, CW, ASO Vascular(No Pencil transcranial) Probe 5729437 Split Crystal) Cardiac, CW, ASO Vascular(No transcranial) Pencil TE100024 Probe...
  • Page 136 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Part Probe Part Name Illustration Application Feature Number Type Neonatal, B, CHI, CF, PDI, M, PW, 10C-D Pediatric, B-Flow, B-Flow Color, Vascular, (No CrossXBeam, transcranial) LOGIQView, ATO/ASO, CTO, SRI-...
  • Page 137 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL LP10 LP10 R4.5 R4.5 R4.5 R2.5 R2.5 (RS/ (RS/ (RS/ (RS/ (RS/ (RS/ Probes type) type) type) type) type) type) type) type) type) type) type) type) type)
  • Page 138: Software/Option Configuration

    Section 3-9-1 "System/General Preset Menu" on page 3-47 More detailed information pertaining System Setup and Measure Setup adjustments is found in the NOTICE LOGIQ P7/P8/P9/P10 Basic User Manual, which is available in different languages. !! NOTICE: 3-60 Section 3-11 - Software/Option Configuration...
  • Page 139: Connectivity Setup

    By utilizing data flows, the user can configure the LOGIQ P7/P8/P9/P10 ultrasound unit to optimally meet the needs of the facility, while keeping the user interface unchanged. Once the dataflow is selected, the actual location of the database is entirely transparent.
  • Page 140: Dataflow Examples

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-12-1-2 Dataflow Examples The local database is used for patient archiving. Images are stored to internal hard drive. The local database is used for patient archiving. Afterwards images are stored to a DVD/CD or external USB device, etc.
  • Page 141: Logiq? + Pc Within A "Sneaker Net

    3-12-1-4 LOGIQ? + PC within a “Sneaker Net” A PC is used for review and work on studies acquired on one or more LOGIQ P7/P8/P9/P10 scanners without being connected in a network. The images are first stored on the LOGIQ P7/P8/P9/P10 scanner’s hard drive (Archive) and then exported from the scanner’s hard drive to a sneaker device (e.g., DVD/CD), and finally imported from...
  • Page 142: Wired Ethernet From Logiq P7/P8/P9/P10 To A Workstation

    You will need a network switch and one network cable for each unit connected to the switch. 3-12-2-3 Connection via Hospital Network 3-12-2-4 You will need one network cable to connect the LOGIQ P7/P8/P9/P10 to a wall outlet on the hospital’s network. 3-12-3 Connection using Wireless Option Refer to: Chapter 5 for theory.
  • Page 143: Configuring Connectivity

    This greatly improves query capability. DICOM Structured Reporting creates coded clinical data that can be used for clinical research, outcomes analysis, and disease management. DICOM Structured Reporting is a standardized format for medical results. LOGIQ P7/P8/P9/P10 supports the following templates: •...
  • Page 144 The ultrasound system is pre-configured for many services, with default settings selected. You can change these services and settings as needed. You must restart the LOGIQ P7/P8/P9/P10 (shutdown) after making any changes to connectivity CAUTION settings in the Utility menus. This includes any changes on the TCPIP or dataflow setup screens.
  • Page 145: Tcpip

    4) Re-boot the ultrasound system. NOTE: TCPIP settings do not get restored when restoring backups. This is per system design. The LOGIQ P7/P8/P9/P10 IP address MUST BE unique. Figure 3-44 Connectivity TCPIP Preset Menu (R1, R2, R2.5) Chapter 3 - Setup Instructions...
  • Page 146 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 3-45 Connectivity TCPIP Preset Menu (R3, R4 and R4.5) Table 3-25 Computer Name Preset Parameter Description Computer Name Type the unique name for the Ultrasound system (no spaces in name).
  • Page 147 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 3-26 Wired Network Configuration Preset Parameter Description Default IPv6 Gateway (R3.0 or Type IPv6 Gateway address above) Prmary DNS Server (R3.0 or Type Primary DNS Server address above) Secondary DNS Server (R3.0...
  • Page 148: Device

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-13-5 Device To add a new device, 1) Press Add. 2) Type the device name in the Name field. 3) Type the device's IP address in the IP Address field.
  • Page 149 Properties: IP Address Type the device's IP address. Properties: AE Title AE Title of the LOGIQ P7/P8/P9/P10. NOTE: Only available for MyComputer. Properties: Port Number IP Port Number Used for DICOM, set by default to 104. NOTE: Only available for MyComputer.
  • Page 150: Service

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3-13-6 Service For each Device that you added to the system, you need to set up the service(s) that device supports (you must be an administrator to update these screens).
  • Page 151: Adding A Service To A Destination Device

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 3-51 Connectivity Services Preset Menu, New Device (R3, R4 and R4.5) The Services screen has the following sections of information: 1) Destination Device - lists information about destination devices. You can select from a list of currently existing devices.
  • Page 152 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 3-29 Service Parameters: Common Service Parameters Preset Parameter Description Timeout The amount of time after which the system will stop trying to establish a connection to the service.
  • Page 153: Agent Configuration (R3, R4 And R4.5)

    Agent Configuration (R3, R4 and R4.5) Overview System may be configured to connect to the back office server (FFA Site), which enables GE Service Personnel to connect to the system to monitor health of the system, perform diagnostics, exchange files, or view console desktop.
  • Page 154 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 3-53 Agent Configuration (R4 and R4.5) Use Agent Configuration to: • Edit and configure the following: - Enterprise server, Select “Production” for install base. Select “Staging”for test purpose...
  • Page 155: Verifying Back Office Connection - Csd

    Verifying Back Office Connection - CSD 1) Login to FFA Site from Remote PC using Chrome web browser As of 2018, FFA Site address is https://ffa.am.health.ge.com 2) In System ID field, type the CRM Number of the machine and select [Get Started] Figure 3-54 FFA Site 3) At CONNECT mode, Select [Connect] of the “Service Desktop”...
  • Page 156: Verifying Back Office Connection - Voc

    ANUAL 3-14-1-3 Verifying Back Office Connection – VOC 1) On Console, invoke “Disruptive Mode” and “VCO” by left click “GE ICON” and select “Connect Clinical Lifeline Figure 3-56 Connecting Clinical Lifeline 2) Login to FFA Site from Remote PC using Chrome web browser As of 2018, FFA Site address is https://ffa.am.health.ge.com...
  • Page 157 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) At CONNECT mode, Select [Connect] of the “Virtual Console” on Remote Console Figure 3-58 Remote Console 5) After selecting [Connect] button, operator may need to wait long time to launch the remote service...
  • Page 158: Connectivity Setup Worksheet

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 3-15 Connectivity Setup Worksheet Site System Information Comments: Floor: Site: Dept: Room: LOGIQ P7/ Type: P8/P9/P10 REV: CONTACT INFORMATION Name Title E-Mail Address Phone Remote Archive Setup...
  • Page 159 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL LOGIQ P7/P8/ P9/P10 IP Address Local Port Host Name AE Title Net Mask ROUTING INFORMATION GATEWAY IP Addresses Destination IP Addresses Default ROUTER1 ROUTER2 ROUTER3 DICOM APPLICATION INFORMATION...
  • Page 160: Paperwork

    Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. 3-82 Section 3-16 - Paperwork...
  • Page 161: Overview

    4-3-11 4-53 Touch Panel Calibration (R3, R4 and R4.5) 4-58 Board Diagnostics Most of the information pertaining to this Functional Checks chapter is found in the LOGIQ P7/P8/P9/ NOTICE P10 Basic User Manual; !! NOTICE: Chapter 4 - Functional Checks...
  • Page 162: Required Equipment

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 4-2 Required Equipment • An empty (blank) CD-R, DVD-R, and/or external USB devices (stick or hard disk drive for data storage/backup. A DVD+RW disc for DVR recording •...
  • Page 163: General Procedure

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 4-3 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes.
  • Page 164: System Exterior Visual Check

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-1 System Exterior Visual Check 4-3-1-1 Physical Abnormalities Figure 4-1 System Exterior (R1 to R3) 1. LCD MONITOR 10. BW printer 18. OPIO tray (option) 2. USB Port 11.
  • Page 165: Physical Abnormalities (R4 And R4.5)

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-1-2 Physical Abnormalities (R4 and R4.5) Figure 4-2 System Exterior (R4 and R4.5) 1. LCD MONITOR 10. BW printer (option) 19. Rear Handle 2. USB port 11. Drawer (option) 20.
  • Page 166: Appearance Inspection

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-1-3 Appearance Inspection Figure 4-3 Label Location (R1, R2, and R2.5) * Required for Asia. LCD CAUTION LABEL GENDER CAUTION LABEL (ONLY FOR INDIA, CHINA, KOREA) MULTI CAUTION LABEL...
  • Page 167: Appearance Inspection (R3)

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-1-4 Appearance Inspection (R3) Figure 4-4 Label Location (R3) * Required for Asia. LCD CAUTION LABEL GENDER CAUTION LABEL (ONLY FOR INDIA, CHINA, KOREA) MULTI CAUTION LABEL...
  • Page 168: Appearance Inspection (R4 And R4.5)

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-1-5 Appearance Inspection (R4 and R4.5) Figure 4-5 Label Location (R4 and R4.5) * Required for Asia. LCD CAUTION LABEL GENDER CAUTION LABEL (ONLY FOR INDIA, CHINA, KOREA)
  • Page 169: Mechanical Parts Functional Check

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-2 Mechanical Parts Functional Check Pay attention to moving parts and pinch point when performing mechanical parts functional CAUTION check. !! CAUTION: [R4 and R4.5] [R1, R2, R2.5 and R3]...
  • Page 170: Main Lcd

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-2-1 Main LCD 1) Tilt LCD Monitor forward 2) Verify can be tilted. Figure 4-7 Main LCD (R1, R2, R2.5 and R3) Figure 4-8 Main LCD (R4 and R4.5)
  • Page 171: Lcd Arm

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-2-2 LCD ARM 1) Loosen LCD Arm LOCK Figure 4-9 LCD ARM Lock (R1, R2, R2.5 and R3) Figure 4-10 LCD ARM Lock (R4 and R4.5) 2) Verify LCD Monitor can move up/down (vertically) and left/right (horizontally).
  • Page 172 LCD Monitor can move approximately 150mm vertically and 250mm horizontally. Figure 4-11 LCD Movement (R1, R2, R2.5 and R3) Figure 4-12 LCD Movement (R4 and R4.5) • a, b : LOGIQ P7/P9 with Flexible Arm • c : LOGIQ P7/P9 with Fixed Arm 3) Engage LCD Arm LOCK.
  • Page 173: Opio Swivel Lock

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-2-3 OPIO Swivel Lock Figure 4-13 OPIO Swivel (R1, R2, R2.5 and R3) Chapter 4 - Functional Checks 4-13...
  • Page 174 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 4-14 OPIO Swivel (R4 and R4.5) 1) Push the swivel button and hold it. 2) Rotate OPIO (while holding button). 3) Verify OPIO is free to rotate.
  • Page 175: Swivel/Brake Lock Caster

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-2-4 Swivel/Brake Lock Caster System weighs approximately 60kg. Pay attention for system stability when working with the CAUTION casters !! CAUTION: NOTE: This procedure applies to RIGHT REAR Caster only.
  • Page 176 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-2-5 Swivel/Brake Lock Caster System weighs approximately 90kg. Pay attention for system stability when working with the CAUTION casters. !! CAUTION: NOTE: This procedure applies to casters other than RIGHT REAR Caster.
  • Page 177: Power On/Off

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-3 Power On/Off After turning off a system, wait at least 10 seconds before turning it on again. NOTICE The system may not be able to boot if power is recycled too quickly.
  • Page 178 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5) Hold down the On/Off button on the control panel for ~3 seconds. NOTE: The power for on-board peripheral auxiliary equipments are commonly switched with the ON/OFF button.
  • Page 179 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 6) Verify On/Off button color changes to GREEN. 7) Verify on-board peripherals power is also turned ON. As soon as the software is loaded, the system enters 2D-Mode with the probe and application that were used before the system was shut down.
  • Page 180: Complete Power Down

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-3-2 Complete power down After turning off a system, wait at least 10 seconds before turning it on again. The system may not be NOTICE able to boot if power is recycled too quickly.
  • Page 181 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 4-20 Power OFF User Interface (R3, R4 and R4.5) 3) Select the SHUTDOWN button. The system performs an automatic full shutdown sequence. 4) Switch OFF the Circuit Breaker at the rear of the system.
  • Page 182: Power Off / Shutdown Check

    Use this button to exit from the System-Exit menu and return to the previous operation. • Exit (Only available when logged in as GE Service with Service Dongle) Select this button when you want to exit to the Windows Desktop.
  • Page 183: System Information

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-4 System Information 4-3-4-1 Software Version 1) From the touch panel, select Utility -> System -> About. 2) Note “Software Version” in Software Group. 3) Note "Software Revision" in Software Group.
  • Page 184 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 4-23 Software version (R4 and R4.5) Chapter 4 - Functional Checks 4-24...
  • Page 185: Service Platform Confirmation (R1, R2, R2.5 Only)

    4-3-4-2 Service Platform Confirmation (R1, R2, R2.5 only) 1) From the touch panel, select Utility -> Service. 2) On Service Browser, select GE Service, and enter the password. 3) Note “SvcPform” version. Figure 4-24 Software Platform version Chapter 4 - Functional Checks...
  • Page 186: Usb Port Test

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-4-3 USB Port Test Required Equipment : Service Dongle External Port Test 1) Plug service dongle to the following locations. Refer to Section 3-9-2 for locations. Front USB Port in OPIO...
  • Page 187: System Integration Checks

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-5 System Integration Checks 4-3-5-1 OPIO Test Figure 4-25 OPIO (R1, R2, R2.5 and R3) Figure 4-26 OPIO (R4 and R4.5) 1) Power ON the system. 2) Check backlight on AN Keyboard, B Buttons, Freeze Key.
  • Page 188 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5) Verify backlight luminance (intensity) changes when knobs are turned. Figure 4-27 Utility Figure 4-28 6) Connect any probe. (if not connected) 7) Go to Live CF Mode. (Mode Rotary Line Area in touch panel, Press “CF” Mode button, unfreezes.) 8) Turn “GAIN”...
  • Page 189: Dvd Drive Test

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 10) Press “COMMENT” key. 11) Select ‘KBD’ in touch panel and Type in arbitrary strings. Verify keyboard is functional 12) Type in arbitrary strings from A/N Keyboard(If installed). Verify keyboard is functional.
  • Page 190: System Integration Checks (R4 And R4.5 Only)

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-6 System Integration Checks (R4 and R4.5 only) 4-3-6-1 OPIO Test Figure 4-29 OPIO 1) Power ON the system. 2) Check backlight on AN Keyboard, B Buttons, Freeze key.
  • Page 191: Dvd Drive Test

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 4-31 4) Connect any probe. (if not connected) 5) Go to Live CF Mode. (Press “CF” Mode button, unfreezes.) 6) Turn “GAIN” encoder around CF Mode button and verify ROI gain change.
  • Page 192: Peripheral Checks

    8) Verify that Recorded image and sound can be played back. NOTE: LOGIQ P7/P8/P9/P10 support USBHDD file capture. and this menu can select in “Utility” “System” “Peripherals” tab. The recorded data can play on your target PC. The support video format mpeg-2 and h.264 and it depend on codec on your PC of operation system.
  • Page 193: Gel Warmer Test

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-7-6 Gel Warmer Test 1) Insert Gel Bottle into Gel Warmer. 2) Power ON the warmer. 3) Verify Gel gets warm. NOTE: Depending on environment, Gel requres 30 minute or more to warm up.
  • Page 194 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 9) Press "Save" and restart the system into Question window by pressing "OK" Button. Figure 4-32 User Configurable Key Page (R1.x.x) Figure 4-33 User Configurable Key Page (R2.x.x)
  • Page 195 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 4-34 User Configurable Key Page (R2.x.x) Figure 4-35 User Configurable Key page (R3.x.x) Chapter 4 - Functional Checks 4-35...
  • Page 196 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 4-36 User Configurable Key Page (R4.x.x or later) Figure 4-37 Hard key and Touch Panel for User interaction (R2.x.x or R3.x.x) Chapter 4 - Functional Checks...
  • Page 197: B-Mode

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 4-38 Hard key and Touch Panel for User interaction (R4.x.x or later) 4-3-8-2 B-Mode 1) Press B Mode button. 2) Verify that system scan with the activated probe.
  • Page 198: Color Flow-Mode

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-8-3 Color Flow-Mode 1) Press CF Mode button. (Active CF Mode button on "User Configurable Mode Rotary".(R1-R3 only)) 2) Increase CF Gain until noise floor shows up.
  • Page 199: Pulse Doppler Mode

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-8-4 Pulse Doppler Mode Figure 4-40 (R2.x.x or R3.x.x) Figure 4-41 (R4.x.x or later) 1) From Utility -> Imaging -> General, verify check on “Simultaneous”. 2) Press B Mode button.
  • Page 200: B/Cf/Pw Mode

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 4-42 Pulse Doppler Mode 4-3-8-5 B/CF/PW Mode 1) Press B Mode button. 2) Press PW Mode Button. 3) Press CF Mode Button. Chapter 4 - Functional Checks...
  • Page 201 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Verify that system scan with B/CFM/PWD Modes. Figure 4-43 B/CF/PW Mode Chapter 4 - Functional Checks 4-41...
  • Page 202: M-Mode

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-8-6 M-Mode 1) Press B Mode Button. 2) Press M Mode Button. (If not R4.X.X, active M Mode button on "User Configurable Mode Rotary".(R1-R3 only)) 3) Verify that system scan B/M with the activated probe.
  • Page 203 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 4-45 BF Setting (R2.x.x or R3.x.x) Figure 4-46 BF Setting (R4.x.x or later) 3) Press B-Flow Button. (If not R4.X.X, active M Mode button on "User Defined Hard Key".(R1-R3...
  • Page 204: Cw Mode (Optional)

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Verify BF Mode. Figure 4-47 B-Flow Mode 5) From the touch panel, select “COLOR” 6) Verify B-Flow Color is shown. Figure 4-48 B-Flow Color Mode 4-3-8-8...
  • Page 205: Wlan (Optional)

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-8-9 WLAN (Optional) NOTE: This test is only applied if the WLAN option has been installed 1) Connect a service dongle to USB port and press a power button on console panel.
  • Page 206: Opio Interface Check (R1, R2, R2.5 Only)

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-9 OPIO Interface Check (R1, R2, R2.5 only) NOTE: In above example, key “Alt” is pressed. NOTE: Alternatively, OP Panel Interface test may be executed from Common Service Desktop.
  • Page 207 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Click the "YES" on Diagnostic Mode page Figure 4-50 Diagnostic Mode page screen after enable the Disruptive mode 4) Select Diagnostics -> OP Panel Interface Figure 4-51 OP Panel Interface page screen...
  • Page 208 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5) Click the "Start" on OP Panel Interface page Figure 4-52 OP Panel Interface page screen 6) Select "Op-Panel Open" 7) Verify OP Panel button operation is reflected on diag window.
  • Page 209: Opio Interface Check (R3, R4 And R4.5)

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-10 OPIO Interface Check (R3, R4 and R4.5) NOTE: Alternatively, OP Panel Interface test may be executed from Common Service Desktop. However, performance / response of the diagnostics may be degraded when executed with scanner software still active.
  • Page 210 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Click the "Enable" on Diagnostic Mode Utility page Figure 4-55 Diagnostic Mode page screen after enable the Disruptive mode (R3) Figure 4-56 Diagnostic Mode page screen after enable the Disruptive mode (R4 and R4.5) 4) Select Diags ->...
  • Page 211 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 4-58 OP Panel Interface page screen (R4 and R4.5) 5) On “Maintenance Access” dialog, 1.) enter the password (R3 only) 2.) click [Maintenance] button. On “System halted” dialog, click [Maintenance] button. (R4 and R4.5)
  • Page 212 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 4-60 OP Panel Interface page screen (R4 and R4.5) 6) Select "Op-Panel Open" 7) Verify OP Panel button operation is reflected on diag window. 8) Check the “Display Touch Panel Layout” and Press the four Points in Touch Panel.
  • Page 213: Touch Panel Calibration (R3, R4 And R4.5)

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-11 Touch Panel Calibration (R3, R4 and R4.5) 1) Power ON the scanner, and wait until the main screen comes up. 2) Insert the service dong to the front USB port.
  • Page 214 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 9) Press the Reset button in Display tab -> Display options -> Details Figure 4-61 Tablet PC Settings dialog 10) Press the OK when pop-up the window.
  • Page 215 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 12) Calibrated carefully 16 calibration points in touch panel. Figure 4-63 Calibration Layout NOTE: Figure 4-35 is intended to be shown upside-down. 13) Press the Yes when pop-up the window.
  • Page 216: Monitor Sound Setting (R4.5)

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4-3-12 Monitor sound setting (R4.5) 1) Power On the scanner and wait until the main screen comes up. 2) Insert the service dongle to the front USB port 3) Press Power button on the operator panel once.
  • Page 217 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 12) Select Stereo Mix on “Recording” menu. Figure 4-66 Recording tab 13) Verify “Listen to this device” is checked and “Playback through this device” is “FW23****”. Figure 4-67 Listen tab...
  • Page 218: Board Diagnostics

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 4-4 Board Diagnostics Battery Category Item DC4D Other Module MCB Swept Demodulator Test MCB Front End Interface Test MCB Analog Test MCB Memory Access Test MCB Front End Interface FPGA Test...
  • Page 219 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Essential Test COMEXP HD Test Long HD Test Short Memory COMEXP COMEXP/ Network Interface EthernetPort COMEXP/ Network Interface Reliability Test EthernetPort RTC Clock COMEXP Video cards COMEXP...
  • Page 220 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Chapter 4 - Functional Checks 4-60...
  • Page 221: Components And Functions (Theory)

    Components and Functions (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter explains LOGIQ P7/P8/P9/P10’s system concepts, component arrangement, and subsystem function. It also describes the Power Distribution scheme and probes. Table 5-1 Contents in Chapter 5 Section Description...
  • Page 222: General Information

    EVISION ERVICE ANUAL Section 5-2 General Information LOGIQ P7/P8/P9/P10 is a digital beamforming curved-, linear- and phased array ultrasound imaging system. The system supports various operating mode such as • B-Mode, M-Mode, Color Flow (CFM) • Power Doppler Imaging (PDI), PW Doppler, CW Doppler •...
  • Page 223 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 5-2 System Exterior (R4 and R4.5) Chapter 5 - Components and Functions (Theory)
  • Page 224: Operator Panel

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-2-2 Operator Panel 5-2-2-1 Basic Block Diagram R1 to R4.5 Figure 5-3 Basic block diagram of LOGIQ P7/P8/P9/P10 R1 to R4 Section 5-2 - General Information...
  • Page 225 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 5-4 Operator Panel and Probe connector (R1, R2, R2.5 and R3) Chapter 5 - Components and Functions (Theory)
  • Page 226 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 5-5 Operator Panel and Probe connector (R4 and R4.5) Section 5-2 - General Information...
  • Page 227: System Options

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-2-3 System Options Figure 5-6 System Options Chapter 5 - Components and Functions (Theory)
  • Page 228 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 5-7 System Options (R4 and R4.5) Section 5-2 - General Information...
  • Page 229: System Ports

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-2-4 System Ports USB Ports Rear Panel Figure 5-8 System Ports (R1, R2, R2.5 and R3) System is equipped with 4 USB ports in front. These are customer use and Service port connected to USB controller via USB Hub.
  • Page 230 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 5-9 System Ports (R4 and R4.5) System is equipped with 2 USB ports in front. These are customer used and service port connected to USB controller via USB Hub.
  • Page 231: Front Usb

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-2-4-1 Front USB Customer use and Service port connected to USB controller via USB Hub. NOTE: Front USBs are non-isolated port and not fit to connect external powered USB Devices.
  • Page 232 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL LAN : 1000Base-T, 100Base-T, 10Base-T USB : USB 2.0, 480Mbps, 5V 500mA. NOTE: Rear USBs are non-isolated port and not fit to connect external powered USB Devices.
  • Page 233: Hdmi

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-2-4-3 HDMI NOTE: HDMI for external monitor Table 5-6 Pin Out Pin number Signal name TMDS Data 2+ TMDS Data 2 Shield TMDS Data 2- TMDS Data 1+...
  • Page 234: S-Video / Composite

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-2-4-4 S-Video / Composite There are S-video / Composite port for external display <S-Video 4Pin Mini-DIN Female> Table 5-7 S-Video 4Pin Mini-DIN Female Pin number Signal name...
  • Page 235: Ac Inlet

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-2-4-5 AC Inlet 100-120V/220-240V AC Figure 5-10 AC Inlet Chapter 5 - Components and Functions (Theory) 5-15...
  • Page 236: Power On/Down Sequence

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 5-3 Power ON/DOWN Sequence 5-3-1 Power On Sequence 1) Circuit breaker ON 2) AC/DC in IOBOX generates standby 5V 3) Standby 5V is supplied to OPIO, COM Express and each board...
  • Page 237: Software Options

    EVISION ERVICE ANUAL Section 5-4 Software Options 5-4-1 Options Following options are offered for LOGIQ P7/P8/P9/P10. Table 5-8 Software Options Software Options Description Auto IMT automatically measures the thickness of the Intima Media on the far and near vessel walls.
  • Page 238 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 5-8 Software Options Software Options Description Auto EF Automated Ejection Fraction (AutoEF) is a semi-automatic measurement tool used for measurement of the global EF (Ejection fraction).
  • Page 239 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 5-8 Software Options Software Options Description UGAP (R4 or later) UGAP( Ultrasound-Guided Attenuation Parameter) calculates attenuation coefficient for the quantification of liver steatosis. SonoAVC (R4 or later) A method for volume calculation available on 3D/4D mode.
  • Page 240: Smart Devices Applications (Logiq Apps)

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-4-2 Smart Devices Applications (LOGIQ Apps) The Smart Devices Applications are available for Android via Bluetooth: • Remote Control App • Camera App To link to the device: 1.) Insert Bluetooth dongle into USB slot...
  • Page 241: Hardware Options

    IRECTION EVISION ERVICE ANUAL Section 5-5 Hardware Options 5-5-1 Options Following options are offered for LOGIQ P7/P8/P9/P10. Table 5-9 Hardware Options Hardware Options Description ECG Option LP7/P8/P9/P10 ECG Module without ECG cables Provides interface to ECG cable ECG Cable - AHA...
  • Page 242 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 5-9 Hardware Options (Continued) Hardware Options Description LP7/P8/P9/P10 Probe Cable Hanger - Front LP7/P8/P9/P10 Front Probe Cable Hanger LP7/P8/P9/P10 Probe Cable Hanger - Rear LP7/P8/P9/P10 Rear Probe Cable Hanger...
  • Page 243: Regional And Peripheral Options

    Printers Description LOGIQ P7/P8/P9/P10 Sony UP-D898 Digital B/W Printer Kit B/W Printer Kit LOGIQ P7/P8/P9/P10 Sony UP-D25MD Digital Color A6 Printer Kit Digital Color A6 Printer Kit LOGIQ P7/P8/P9/P10 Color Inkjet Printer (Off-board) Inkjet Printer Chapter 5 - Components and Functions (Theory)
  • Page 244: Mechanical Descriptions

    • Depth: 740mm • Weight: 69kg (152 lb) 5-7-2 LCD Monitor The LOGIQ P7/P8/P9/P10 has a free adjustable LCD monitor in relation to the user interface. • position up/down: +/- 7.5 cm • position left/right: +/- 18cm • rotation up/down: +90°/-15°...
  • Page 245: Opio Positioning

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-7-3 OPIO Positioning The control console can be rotated, translated and adjusted in height. Highest 910mm Lowest 810mm Figure 5-12 Adjustable Control Console (R1, R2, R2.5 and R3)
  • Page 246 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 5-13 Adjustable Control Console (R4 and R4.5) 5-26 Section 5-7 - Mechanical Descriptions...
  • Page 247: Air Flow Distribution

    Clean the Air Filter periodically. Through the filter grid on the front of the system, air flow into the LOGIQ P7/P8/P9/P10 scanner. By means of the 2 FAN, air is blown through the nest-box, and the warm air exits the scanner through holes in the left side panel and rear of the system.
  • Page 248: Exiting To Windows Desktop

    ERVICE ANUAL Section 5-8 Exiting to Windows Desktop ® LOGIQ P7/P8/P9/P10 backend software operates on embedded version of Windows 7 operating ® system (R2 and R2.5) or ersion of Windows 10 Enterprise operating system (R3, R4 and R4.5). To access some of the functions, service personnel occasionally require to enter Windows desktop to access files, folders or applications.
  • Page 249 5) On “Maintenance Access” window, type in service password (user name not required). NOTE: Service Password is rotating password that changes on 6 months period. Only the GE service personnel or GE authorized dealers have access to maintenance password. NOTE: Maintenance Password is function of date/time.
  • Page 250 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 5-19 Windows Desktop (R4 and R4.5) 5-30 Section 5-8 - Exiting to Windows Desktop...
  • Page 251: Service Platform

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 5-9 Service Platform 5-9-1 Introduction The Service Platform will increase service productivity and reduce training and service costs. 5-9-2 Access / Security The Service Platform has different access and security user levels.
  • Page 252: How The Customer Enables/Disables Disruptive Mode And Vco

    5-9-2-2 How the Customer enables/disables Disruptive Mode and VCO 1) If not already in read mode, FREEZE the image. 2) Move the cursor to the GE icon and press the (= right-click). right trackball key 3) Select Connect Clinical Lifeline.
  • Page 253: Common Service Desktop (Csd) (R1, R2, R2.5 Only)

    5-10-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface. There are different possibilities to access the Common Service Desktop and its available features:...
  • Page 254: Csd Top Page

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-10-2 CSD Top Page 5-10-2-1 General Layout Table 5-12 System Information System ID, Serial, IP Address, InSite StatusÅc.etc Connected Probes Connected and System recognized probe listed Option...
  • Page 255: Decoding System Temperature And Voltage

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-10-2-2 Decoding System Temperature and Voltage NOTE: These are reference only values and therefore, parameters are not constant. Some parameters may be added/deleted depending on the system configuration, type, and variation.
  • Page 256 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-36 Section 5-10 - Common Service Desktop (CSD) (R1, R2, R2.5 only)
  • Page 257: Error Logs

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-10-3 Error Logs When the Error Logs page is selected, different log viewing options are available. Log Viewer is displayed in a separate window. 5-10-4 Diagnostics Figure 5-25 Common Service Desktop - Error Logs The Diagnostic page uses a web-controlled user interface to provide access to common service components and perform diagnostics.
  • Page 258: Image Quality

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-10-5 Image Quality In the Image Quality page, you can verify image quality. Figure 5-27 Common Service Desktop - Image Quality This page is not populated in this version.
  • Page 259: Configuration

    Figure 5-28 Common Service Desktop - Configuration Remote access is ONLY possible if the service platform is properly configured (either by the user NOTICE or a GE technician at site). Operation see: Section 7-5-6 "CSD: Configuration" on page 7-21. !! NOTICE:...
  • Page 260: Utilities

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-10-7 Utilities The Utilities page contains a variety of Windows utility tools to indicate the status of the system, in addition to various other tools. Figure 5-29 Note on Utilities: ®...
  • Page 261 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Select “Disk Defragmenter” from left pane Figure 5-31 4) Select Disk Drive to defragment on right (upper) pane 5) Press “Defragment” to execute defragmentation of the selected disk NOTE: Press “Analyze”...
  • Page 262: Replacement

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-10-8 Replacement There is no accessible parameters in Replacement page. 5-10-9 In the PM page, information about planned, proactive and preventive maintenance is displayed, as described in Chapter 10 - Care &...
  • Page 263: Common Service Desktop (R3, R4 And R4.5)

    • Local user – a user locally logged into the machine will be able to set the LOGIQ P7/P8/P9/P10 to Disruptive mode or allow a Disruptive mode request from a remote user through the Service desktop.
  • Page 264: Color Statuses

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-11-3 Color statuses Throughout the Service desktop, colors indicate the following: • Green – Status is normal • Orange – Status is a warning • Red – Status is an error...
  • Page 265 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL With Service Advanced (Class C), these are the available options: • HOME • Diags - Run Diags - Diags History • DICOM - DICOM Service Status - DICOM Spooler •...
  • Page 266 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL With GE Service access (Class M) and an SSA key, these are the available options: • HOME • Diags - Run Diags - Diags History • DICOM...
  • Page 267 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL With Remote access (Back office server), these are the available options: • HOME • Diags - Run Diags - Diags History • DICOM - DICOM Service Status - DICOM Spooler •...
  • Page 268: Advanced Information - Service Desktop And Ssa (Secure Service Access)

    ANUAL 5-11-5 Advanced Information - Service Desktop and SSA (Secure Service Access) SSA Key The SSA key provides secure access for GE service personnel to advanced tools to service the system • Access to all service features • Access to Windows Desktop •...
  • Page 269: Home

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-11-6 Home Home configurations vary depending upon the purchased service level. Figure 5-34 Home with Class C and Class M Access (R3 only) Figure 5-35 Home with Class C and Class M Access (R4 and R4.5 only)
  • Page 270 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 5-36 Home with Class A Access (R3 only) Figure 5-37 Home with Class A Access (R4 and R4.5 only) For more information, see: • System Information...
  • Page 271 IRECTION EVISION ERVICE ANUAL System Information System Information displays general information about the LOGIQ P7/P8/P9/P10. When the LOGIQ P7/P8/P9/P10 has been successfully configured with the back office, these elements will have the corresponding values: • Agent Registered will be Yes Agent Quarantine will be No •...
  • Page 272 • No – The agent is not verified in the back office. • Not Available – The agent is not running or has not been configured. GE part number for the LOGIQ P7/P8/P9/P10. The same number as listed on the Model Number rating plate.
  • Page 273 ERVICE ANUAL Software Status Use Software Status to view general information about the software installed on the LOGIQ P7/P8/P9/ P10. The information on Software Status is available to all service class licenses. To access Software Status, navigate to Utility > Service > Home.
  • Page 274 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 5-15 Software Status Element DESCRIPTION Date the base image software was installed. The base image software includes the Base Image Installation Date Windows operating system and other supporting software.
  • Page 275 IRECTION EVISION ERVICE ANUAL System Status Use System Status to view status information on the LOGIQ P7/P8/P9/P10. Specifically, do the following: • View the hardware configuration file • View temperatures for FRUs, fans, and a graphical representation of the monitored temperatures •...
  • Page 276 30 days or last status checkpoint whichever is sooner. Based on the latest occurrence, displays all abnormal boots of the LOGIQ P7/P8/ Bootup Issues P9/P10 for either the past 30 days or the latest check point.
  • Page 277 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Hardware Configuration Hardware Configuration displays the contents of the hardware configuration file which includes all the vital product data (VPD). To access Hardware Configuration, navigate to Utility > Service > Home, and then under System Status, select Hardware Configuration.
  • Page 278 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Temperature Use Temperature to troubleshoot problems with monitored FRUs and the fans. Specifically, use these pages: • FRU Status • Graphs To access Temperature, navigate to Utility > Service > Home, and then under System Status, select Temperature.
  • Page 279 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL FRU Status FRU Status displays a summary of the FRUs being monitored and the details for each FRU. To access this page, under System Status, select Temperature and then select FRU Status.
  • Page 280 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Graphs Graphs displays trend graphs for selected elements which have been selected as key indicators of overall temperature status inside the card rack. To access this page, under System Status, select Temperature and then select Graphs.
  • Page 281 IRECTION EVISION ERVICE ANUAL Voltage Use Voltage to troubleshoot problems with monitored FRUs of the LOGIQ P7/P8/P9/P10. Specifically, use these pages: • FRU Status To access Voltage, navigate to Utility > Service > Home, and then under System Status, select Voltage.
  • Page 282 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL This table shows all the elements available on FRU Status with descriptions. Table 5-19 FRU Status Element DESCRIPTION Name Name of the FRU. Status of the voltage for the FRU. Valid values are: •...
  • Page 283 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Bootup Issues Bootup Issues displays a summary of bootup issues detected by the system that prevented normal scanning, shutdowns, or abnormal statuses. To access this page, under System Status, select Bootup Issues.
  • Page 284 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL This table shows all the elements available on Bootup Issues with descriptions. Table 5-20 Bootup Issues Element DESCRIPTION Identifier for the system software boot. For example, 10 for the tenth boot and so on.
  • Page 285 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Alert Logs Alert Logs displays the details for each alert. Alerts are status logs typically reported as a result of an error. Alerts are sent to the back office to be managed and reported to service.
  • Page 286 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL USB Devices USB Devices displays the details for each USB device. To access this page, under System Status, select USB Devices. Figure 5-51 USB Devices This table shows all the elements available on USB Devices with descriptions.
  • Page 287 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Disk Drive Space Use Disk Drive Space to view the hard drive partitions including total size and available free space in GB. Includes the overall health of the drive in one of these colors: •...
  • Page 288 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 5-53 E: Drive Details 5-68 Section 5-11 - Common Service Desktop (R3, R4 and R4.5)
  • Page 289 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL This table shows all the elements available on Disk Drive Space with descriptions. Table 5-23 Disk Drive Space Element DESCRIPTION C: partition shows the graphical representation of the following: C: Drive •...
  • Page 290 Network Configuration Use Network Configuration to view network (wired and wireless) information (the full ipconfig details) for the network configured with the LOGIQ P7/P8/P9/P10. The information on Network Configuration is available to Class C and Class M licenses. To access Network Configuration, navigate to Utility > Service > Home.
  • Page 291 This table shows all the elements available on Network Configuration with descriptions. Table 5-24 Network Configuration Element DESCRIPTION Host Name Name of the local host for the LOGIQ P7/P8/P9/P10. More Details Displays the ipconfig details. Wired Connection lpV4 Address Local IP address for the wired network connection.
  • Page 292 ANUAL Connected Probes Connected Probes shows probes connected to the LOGIQ P7/P8/P9/P10. The order on the user interface is top down matching the left-to-right order on the LOGIQ P7/P8/P9/P10. The information on Connected Probes is available to all service class licenses.
  • Page 293: Diags

    Run Diags – Run diagnostics on hardware components. • Diag History – View a history of the diagnostics you have run and the results. NOTE: Reboot the system after performing any diagnostics before returning the LOGIQ P7/P8/P9/P10 to customer use. For more information, see: •...
  • Page 294 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL This table shows all the elements available on Run Diags with descriptions. Table 5-26 Run Diags Element DESCRIPTION Service Menu Available service diagnostics in a diagnostics tree.
  • Page 295 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL To run a diagnostic: 1.) Navigate to Utility (second page) > Service > Utilities > Disruptive Mode Utility and set Disruptive mode to On. 2.) Navigate to Utility (second page) > Service > Diags > Run Diags.
  • Page 296 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7.) View progress in the Total Progress area. 8.) View the results (Passed or Failed) in the Diagnostic Results Summary area. Figure 5-60 (R3) Figure 5-61 (R4 and R4.5) 10.)To view details about the diagnostics run, see Diag History...
  • Page 297 ANUAL Diag History Use Diag History to view a history of the diagnostics run on the LOGIQ P7/P8/P9/P10 and the results. The information on Diag History is available to Class C and Class M licenses. To access Diag History, navigate to Utility > Service > Diags > Diag History.
  • Page 298 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL To view a history of diagnostics: 1.) Navigate to Utility (second page) > Service > Diags > Diag History. 2.) Select the options that you want to view.
  • Page 299: Dicom

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-11-8 DICOM Use DICOM to troubleshoot DICOM components and connectivity. Specifically, use these pages: • DICOM Service Status • DICOM Spooler For more information, see: • DICOM Service Status...
  • Page 300 Application entity title of the LOGIQ P7/P8/P9/P10. Port No of Console Port number of the LOGIQ P7/P8/P9/P10. DICOM Connectivity Status Service Name Name of the DICOM services configured for the LOGIQ P7/P8/P9/P10. Status of the DICOM service. Valid values are: • OK • Not Ok Status •...
  • Page 301 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL DICOM Spooler DICOM Spooler provides a summary of the current and recent outgoing jobs. The information on DICOM Spooler is available to Class C and Class M licenses.
  • Page 302 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL This table shows all the elements available on DICOM Spooler with descriptions. Table 5-29 DICOM Spooler Element DESCRIPTION Job Status Displays the number of jobs for each type by status. For example, Pending.
  • Page 303 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-11-9 Utilities Utilities configurations vary depending upon the service class. For more information, see: • Change Password on page 5-89 • Checkpoints on page 5-103 • Clean Userdefs...
  • Page 304: Utilities

    Disk fragmentation can reduce the amount of disk space available, and slow computing speed. Use the disk defragmenter to restore optimum disk space and speed performance. If LOGIQ P7/ P8/P9/P10 performance would be significantly reduced when the disk is being defragmented, a warning message displays.
  • Page 305 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Gather Logs Gather Logs provides a way to collect system logs and place the log files in the D:\Service directory for retrieval by the online center. These log files do not include protected data such as crash dumps and keyboard shadow logs.
  • Page 306 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Delete Files Delete Files displays all the files and folders present in the D:\Service folder and allows for their deletion. Deleting unneeded files improves performance and reduces the need to defragment the disk drive.
  • Page 307 1.) Navigate to Utility (second page) > Service > Utilities, and then select Disable Remote Connectivity. 2.) Select Disable Remote Connectivity. 3.) To disable connectivity to the LOGIQ P7/P8/P9/P10 through Agent Configuration, SSH, and Virtual Console Observation, click OK. 4.) Once connectivity has been disabled, select OK again.
  • Page 308 3.) Select Start. When VCO starts, the Windows task bar and Windows desktop display. 4.) Modify settings and programs as needed. 5.) To stop virtual console observation, click Stop. When VCO is stopped, the Windows desktop is hidden. 6.) Restart the LOGIQ P7/P8/P9/P10. For more information, see: • Utilities on page 5-83 5-88 Section 5-11 - Common Service Desktop (R3, R4 and R4.5)
  • Page 309 RISK OF LOSING DATA. CAUTION IF THE PASSWORD IS LOST, GE WILL NOT BE ABLE TO RECOVER OR RESET IT. LOSS OF A PASSWORD MAY RESULT IN THE LOSS OF PATIENT DATA. The information on Change Password is available to all service class licenses. Change Password is not available through a remote connection.
  • Page 310 GE WILL NOT BE ABLE TO RECOVER OR RESET CHANGED PASSWORDS. SECURELY RECORD THE NEW PASSWORD. 4.) Click Update Password. 5.) When a SVCService user password has been changed, reboot the LOGIQ P7/P8/P9/P10 to reflect the password change. For more information, see: •...
  • Page 311 ANUAL Software Reload Software Reload displays versions of the base image, application software, base patch, service platform, and documentation packages currently installed on the LOGIQ P7/P8/P9/P10. Software Reload reimages the C:\ drive and installs the base image. • If the drives are encrypted, Software Reload will have a catastrophic effect if the customer does not have the recovery key.
  • Page 312 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL To access Software Reload, select Utility > Service (second page) > Utilities > Software Reload. Figure 5-71 Software Reload (R3 only) Figure 5-72 Software Reload (R4 and R4.5) 5-92 Section 5-11 - Common Service Desktop (R3, R4 and R4.5)
  • Page 313 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL This table shows all the elements available on Software Reload with descriptions. Table 5-34 Software Reload Element DESCRIPTION Base Image Version Displays the version number of the base image software.
  • Page 314 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL To reload software: 1.) If the drives are encrypted, obtain the recovery key. 2.) Navigate to Utility (second page) > Service > Utilities > Disruptive Mode Utility and set Disruptive mode to On.
  • Page 315 Once the operation is complete, the LOGIQ P7/P8/P9/P10 reboots with an empty database. Disruptive mode must be On to reset the patient database. If Disruptive mode is Off, a warning message displays asking to turn on Disruptive mode.
  • Page 316 On to clean user-defined data. If Disruptive mode is Off, a warning message displays asking to turn on Disruptive mode. Before cleaning user-defined data, the LOGIQ P7/P8/P9/P10 application software automatically shuts down. Once the operation is complete, the system reboots.
  • Page 317 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL To clean user-defined data: 1.) If you want to use the current settings later, back them up now. 2.) Navigate to Utility (second page) > Service > Utilities > Disruptive Mode Utility and set Disruptive mode to On.
  • Page 318 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Third Party Software Licenses Third Party Software Licenses displays the third-party software licenses used as part of the service platform. The information on Third Party Software Licenses is available to all service class licenses.
  • Page 319 Class M Key Status • Not Plugged In • Plugged In Drive Letter Drive where the SSA key is plugged into the LOGIQ P7/P8/P9/P10. Expire Date Date the SSA key is set to expire. SSO ID Identifier for the user assigned to the SSA key.
  • Page 320 2.) If open, close the Service desktop. 3.) Restart the LOGIQ P7/P8/P9/P10. 4.) Once the LOGIQ P7/P8/P9/P10 has restarted, plug in the SSA key. 5.) Navigate to Utility (second page) > Service > Utilities > SSA License. 6.) Click Restore SSA.
  • Page 321 System Shutdown With remote access, System Shutdown provides a way to shut down or restart the LOGIQ P7/P8/P9/ P10. Disruptive mode must be On to shut down or restart the LOGIQ P7/P8/P9/P10. If Disruptive mode is Off, a warning message displays.
  • Page 322 EVISION ERVICE ANUAL To remotely start, restart, or shut down the LOGIQ P7/P8/P9/P10: 1.) Navigate to Utility (second page) > Service > Utilities > Disruptive Mode Utility and set Disruptive mode to On. 2.) Navigate to Utility (second page) > Service > Utilities > System Shutdown and then select an operation: To shut down the LOGIQ P7/P8/P9/P10, select Shutdown System.
  • Page 323 Checkpoints allows for the creation of checkpoints and displays the history of those checkpoints. Checkpoints are used to indicate when the LOGIQ P7/P8/P9/P10 has been serviced, status has been reset, and information related to the service performed has been added.
  • Page 324 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL To create a check point: 1.) Navigate to Utility (second page) > Service > Utilities > Checkpoints. 2.) Under Create Checkpoints, from Action, select the action. 3.) In Comment, enter notes for the checkpoint.
  • Page 325 If No is selected, then Disruptive mode is not turned on and an error message displays saying that permission to change the mode was denied. When Disruptive mode is On, the GE icon in the status bar changes to red and displays this message: "Due to Service testing restart needed."...
  • Page 326 • View information about past transfers of (APM) information. • Set up automatic/scheduled transfer of allowed data files from the LOGIQ P7/P8/P9/P10 to the server. • Manually transfer data files allowed data files from the LOGIQ P7/P8/P9/P10 to the server.
  • Page 327 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL This table shows all the elements available on Data Transfer with descriptions. Table 5-40 Data Transfer Element DESCRIPTION Type of log file. For example, Incremental Logs or Full Logs. Monitoring Logs,...
  • Page 328 Transfer Machine Logs to GE – Default is checked Transfer Operator Information to GE Transfer Images (captured without patient information) to GE – Default is checked. Enable Automatic Probe Diagnostics – Default is unchecked. (R4.5 only) When Enable Automatic Probe Diagnostics, allows the Probe Check (VITA) Tool to gather information about probes used and put that information into a VITA folder.dump file.
  • Page 329 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2.) Click Save. 3.) Navigate to Utility (second page) > Service > Utilities > Data Transfer. 4.) On Data Transfer, select Scheduler, and then select the days to perform the data transfer.
  • Page 330 This protocol is used for remote login to the operational system. SSH allows remote login to the LOGIQ P7/P8/P9/P10 and remote file transfer from the LOGIQ P7/P8/P9/P10. Remotely log into the LOGIQ P7/P8/P9/P10 using the username “GEService”...
  • Page 331 ANUAL Network Capture Network Capture displays network traffic between the LOGIQ P7/P8/P9/P10 and configured devices. A network capture outputs two log files: one for main logging with no protected information and another including protected information. These log files are useful when debugging connectivity issues.
  • Page 332 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL To perform a network capture: 1.) Navigate to Utility > Service > Utilities > Network Capture. 2.) From Network Capture, do the following: Under Maximum Size, select the allowed size of the generated log file.
  • Page 333: Options

    Software Options Option Software options on the LOGIQ P7/P8/P9/P10. Status Status of the options on the LOGIQ P7/P8/P9/P10. Software Option Details Product Name of the product. Number for the hardware. The hardware number is the hash of the serial number Hardware Number that is used to generate the option key.
  • Page 334 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 5-43 Options Element DESCRIPTION Software Option Keys Available Keys List of the option keys installed on the LOGIQ P7/P8/P9/P10. 5-114 Section 5-11 - Common Service Desktop (R3, R4 and R4.5)
  • Page 335 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-11-11 Agent Configuration Use Agent Configuration to: • Edit and configure the following: Enterprise host name in the agent Enterprise port number in the agent Proxy server in the agent...
  • Page 336 Element DESCRIPTION Agent Configuration Phone number for the person at the customer site a GE remote service engineer Contact Details would contact. The phone number is entered during installation and reviewed at every service call to make sure the information is correct.
  • Page 337 LOGIQ P7/P8/P9/P10. Serial Number The serial number of the agent is tied to the serial number of the LOGIQ P7/P8/P9/ P10. Customer Relationship Management (CRM) number. System identifier assigned to the customer unit by the service region.
  • Page 338 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5-118 Section 5-11 - Common Service Desktop (R3, R4 and R4.5)
  • Page 339: Overview

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of Chapter 6 This chapter describes how to test and adjust the mechanical capabilities of a scanner that may be out of specification.
  • Page 340: Lcd Monitor Adjustment

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 6-3 LCD Monitor Adjustment 6-3-1 Brightness/Contrast (R1.x.x) 6-3-1-1 Brightness Adjusting the monitor’s contrast and brightness is one of the most important factors for proper image quality. If these controls are set incorrectly, the Gain, TGC, Dynamic Range and even Acoustic Output may have to be changed more often than necessary to compensate.
  • Page 341: Osd Control On The Utility Mode (R1.X.x)

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 6-3-2 OSD Control on the Utility mode (R1.x.x) Figure 6-2 Monitor 1) Go to Utility -> System -> General. 2) See Monitor category. 3) There are preset parameters below for monitor setting.
  • Page 342: Inch Wide Monitor Adjustment (R2.X.x And Later)

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 6-3-3 21.5 inch Wide Monitor Adjustment (R2.x.x and later) All Monitor settings have been optimized at the factory, so normally, there should be no need for any further adjustments.
  • Page 343 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Display scan screen while the Utility is ac- Display Test patterns tive while the Utility is active Select Color Profile Select Gamma Select Room Profile Adjust Brightness and Contrast for Room profile “User Defined”...
  • Page 344: Scan Screen

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 6-3-3-1 Scan Screen In order to view the monitor while adjusting the Contrast/Brightness and Color Profile, press the Scan Screen on the second Utility Page. 1) Press Scan Screen. The scan screen displays on the main display while the Utility touch panel is active.
  • Page 345: Advanced Monitor Adjustment (R4 And R4.5)

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 6-4 Advanced monitor adjustment (R4 and R4.5) 6-4-1 Review Gamma Setting Gamma settings can be adjusted in two ways, with or without DICOM Gray Scale Standard Display Function (GSDF) enabled.
  • Page 346: Test Patterns

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 6-4-2 Test Patterns Test patterns are available to aid in evaluating the display visually and quantitatively. Table 6-4 Test Patterns Test Patterns Description Corresponding Graphic Confirm that you can clearly see all of the...
  • Page 347 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 6-4 Test Patterns Test Patterns Description Corresponding Graphic Luminance for different grayscale levels. Usually, the Facility’s Biomed Engineer tests each gray level map on the monitor from LN-1 to...
  • Page 348 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 6-4 Test Patterns Test Patterns Description Corresponding Graphic Contrast. Confirm that you can detect the circles CT Used for Luminance response within the square grayscale boxes.
  • Page 349: Diagnostics/Troubleshooting

    7-19 7-5-2 How the Customer enables/disables Disruptive Mode and VCO 7-19 7-5-3 GE Icon and Remote Connection 7-19 7-5-4 Customer Granting Full Remote Access Permission to GE Service Technician 7-21 7-5-5 Common Service Desktop (CSD) 7-21 7-5-6 CSD: Configuration 7-22...
  • Page 350: Collect Vital System Information

    NOTE: This information is normally collected with the Alt+D or Gather Logs utility. - Product Name = LOGIQ P7/P8/P9/P10 1) From the touch panel, select Utility -> System-> About. Figure 7-1 About (R1, R2, R2.5 only)
  • Page 351 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 7-2 About (R3 only) Figure 7-3 About (R4 only) Figure 7-4 About (R4.5 only) Chapter 7 - Diagnostics/Troubleshooting...
  • Page 352 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Record software version and System Image version Applications Software - Software Version - Software Revision - Software Part Number - Build View - Build Date System Base Image Software...
  • Page 353: Request For Service (Rfs) (R1, R2, R2.5 Only)

    There are two possibilities to contact GE: • by means of the GE “Remote Status Icon” that is displayed on the bottom of the screen. Move the cursor to the GE icon and press the (= left-click) to display the left trackball key “Request For Service”...
  • Page 354 If the service platform is not configured an Error message is displayed. The request is NOT sent! NOTE: The request is saved in QUEUE for reviewing and/or resending. Figure 7-7 Contact GE - Queue Section 7-2 - Collect Vital System Information...
  • Page 355: Request For Service (Rfs) (R3, R4 And R4.5)

    Request for Service (RFS) (R3, R4 and R4.5) 7-2-3-1 Request for Service 1) Move the cursor to the GE icon and press the right-click, select “Contact GE” Figure 7-8 Contact GE (R3 only) Figure 7-9 Contact GE (R4 and R4.5) If Agent configuration is not configured correctly, "Contact GE"...
  • Page 356 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) The Request for Service (RFS) page is displayed Figure 7-10 Contact GE – Request for Service Section 7-2 - Collect Vital System Information...
  • Page 357 IRECTION EVISION ERVICE ANUAL 3) Click the “Submit” button to send the problem description to GE Service/Application representatives If Agent configuration is not configured correctly, the "Submit" button is disabled. Refer to the below. NOTICE !! NOTICE: Figure 7-11 Contact GE – Not configured correctly of the Agent configuration 4) Click the “RFS History”...
  • Page 358: Check Points Voltages

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 7-3 Check Points Voltages 7-3-1 LED for Power Status LEDs near the front cooling fan can provide information pertaining to the DC power distribution status. 1) Remove Front Cover (no tools required)
  • Page 359: Screen Captures And Logs

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 7-4 Screen Captures and Logs Sometimes, the customer or field engineer wants to capture a presentation on the screen to identify the issue may be recovered by the Online Center (OLC). This is accomplished by saving the log files.
  • Page 360: System Warning/Errors And Logs

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7-4-3 System Warning/Errors and Logs 7-4-3-1 Temperature warning 7-4-3-1-1 Cause CPU Temperature exceeded threshold (104°C). 7-4-3-1-2 What to do 1) Shut down the system and Clean the filter.
  • Page 361: System Voltage Failure

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7-4-3-3 System voltage failure 7-4-3-3-1 Cause System detected unexpected voltage level from voltage level monitoring circuits. 7-4-3-3-2 What to Do 1) Press OK and reboot the system.
  • Page 362: Hardware Configuration Error

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7-4-3-4 Hardware configuration error 7-4-3-4-1 Cause Hardware(s) not detected as valid system. System will boot up but will not allow scanning. 7-4-3-4-2 What to do 1) Press OK, and shutdown the system and reboot.
  • Page 363: System Error

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7-4-3-5 System Error 7-4-3-5-1 Cause Unrecoverable error occurred and unable to continue scan. Error dialog does not indicate the root cause of system error. To identify the root cause clearly, it is necessary to investigate the log files.
  • Page 364: Log Message Of System Error

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7-4-3-5-3 Log message of System error Log message format “Error ; EAErrorHandler(****); Error from” + ErrorModuleName + ErrorMessage 7-4-3-5-4 Types 7-4-3-5-4-1 Example1 Hardware Errors: Hardware error Error Module Name Exception Type “Error ;...
  • Page 365 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Error Descriptions: Message Description GFE Ch0 empty-GFE Ch1 empty-CPS_ERROR(FAULT_HVLV). Severity :FatalHW MDC Error. GFE Ch0 empty-GFE Ch1 empty-CPS_ERROR(FAULT_POWER). Severity :FatalHW MDC Error. GFE Ch0 empty-GFE Ch1 empty-CPS_ERROR(FAULT_STOP). Severity :FatalHW MDC Error.
  • Page 366: Remote Access To Service Platform

    OLC can access the customer’s ultrasound scanner remotely. “Disruptive Mode” can be selected by the customer directly on the LOGIQ P7/P8/P9/P10 ultrasound system (see: Section 7-5-2 on page 7-16), or remotely by the service technician or OLC.
  • Page 367: How The Customer Enables/Disables Disruptive Mode And Vco

    7-5-2 How the Customer enables/disables Disruptive Mode and VCO 1) If not already in read mode, FREEZE the image. 2) Move the cursor to the GE icon and press the (= right-click). right trackball key 3) Select Connect Clinical Lifeline This activates “Disruptive Mode”...
  • Page 368: Customer Granting Full Remote Access Permission To Ge Service Technician

    Customer Granting Full Remote Access Permission to GE Service Technician 7-5-4-1 If GE Service Technician requests Remote Access Permission If a GE Service technician requests remote access to your LOGIQ P7/P8/P9/P10 scanner, following “InSite Notification” appears on the systems screen. select [Yes] to enable “Disruptive Mode”...
  • Page 369: Common Service Desktop (Csd)

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7-5-5 Common Service Desktop (CSD) Refer to Section 3-13 for accessing Common Service Desktop remotely. 7-5-6 CSD: Configuration Refer to Section 3-13 for accessing Common Service Desktop remotely.
  • Page 370: Minimum Configuration To Boot/Scan

    24, 2022) IRECTION EVISION ERVICE ANUAL Section 7-6 Minimum Configuration to Boot/Scan LOGIQ P7/P8/P9/P10 will require at minimum MPI, MCB (with COMExpress),MDC, IO Box, HDD(Mounted on MBP) and OPIO to boot up. 7-22 Section 7-6 - Minimum Configuration to Boot/Scan...
  • Page 371: Troubleshooting Trees, Instructions And Tech Tips

    24, 2022) IRECTION EVISION ERVICE ANUAL Section 7-7 Troubleshooting Trees, Instructions and Tech Tips Table 7-2 below outlines LOGIQ P7/P8/P9/P10 troubleshooting trees and instructions shown in the sub- sections. Table 7-2 Troubleshooting Trees and Instructions Sub-section Description Page Number 7-7-1...
  • Page 372: System Does Not Boot Up

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7-7-1 System does not boot up System doesn’t boot Verify that Is Standby Switch - Power Cord is connected to Mains illuminated ? - Power Switch is turned on...
  • Page 373: Noise Disturbs The Image

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7-7-2 Noise disturbs the image Noise is shown in Image. Is there any electrical devices that affect the Scanner? Keep the Scanner away from those electrical devices.
  • Page 374: Trackball - Impaired Sensitivity

    EVISION ERVICE ANUAL 7-7-3 Trackball - Impaired Sensitivity The trackball has low sensitivity! Contact GE service representative Is it difficult to roll the to replace the trackball module. trackball? Set up the appropriate trackball response time (via [Trackball Speed] button in the System Setup - User Settings page).
  • Page 375: Printer Malfunction

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7-7-4 Printer Malfunction Printer malfunction Is printer Check all signal and power supply cable properly connections between Printer and Scanner connected? Load paper by using the correct format.
  • Page 376: Monitor Troubleshooting

    How to import patient data base from older consoles Customers should back up their dataset through DICOM Export from their older scanner to import back onto the LOGIQ P7/P8/P9/P10. The LOGIQ P7/P8/P9/P10 can only understand DICOM media and cannot understand GE-specific database or private xml.
  • Page 377: Use Sclogwindow

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7-7-8 Use ScLogWindow The ScLogWindow application helps identify errors while the system is running. Access this application from the Windows desktop or through Task Manager. From ScLogWindow, you can: •...
  • Page 378 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL To open ScLogWindow from Task Manager: 1.) Insert and activate the SSA key. 2.) Press Ctrl+Alt+Del and select Task Manager. 3.) In Task Manager, select File > Run New Task.
  • Page 379 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL To view logs in the ScLogWindow: 1.) With the ScLogWindow open, double-click the EchoLoader icon. 2.) Use Alt+Tab to put the ScLogWindow in front. 3.) While the application is booting up, observe logging in ScLogWindow.
  • Page 380: Ssa Key

    Expiration of an SSA key When you plug in an SSA key to the LOGIQ P7/P8/P9/P10 and enter the code to unlock it, a message should display on the screen acknowledging the connection. If the message does not appear, the SSA key may have expired.
  • Page 381: Log Analysis

    Service alert information is contained in /Log/SvcAlert*.txt. Use this log to troubleshoot issues with alerts generated by the LOGIQ P7/P8/P9/P10. Alerts can also be viewed on the Service desktop. Review log file content to determine the errors the LOGIQ P7/P8/ P9/P10 has generated in the past.
  • Page 382 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL System configuration logs Hardware configuration information is contained in /Log/SysCfg.csv. Use this file to troubleshoot issues with hardware configuration (also known as VPD). This log helps to determine the part numbers of boards installed along with versions and serial numbers.
  • Page 383: Automatic Probe Diagnostics (Probe Check / Vita)

    GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7-7-11 Automatic probe diagnostics (Probe Check / VITA) For R4.5.x and later, automatic probe diagnostics (PHT Probe Check /VITA) has been added. NOTE: To enable automatic probe diagnostics (Probe Check/VITA), see Section "To configure automatic...
  • Page 384: Max. Number Of Total Defect Elements

    GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7-7-11-1 Max. number of total defect elements Increasing defect transducer elements result in a more and more dark image and might reduce the lateral resolution. If the overall limit of allowed defect element is reached, the algorithm rates the connected probe as failed.
  • Page 385 Attempting do-it-yourself repairs invalidate warranty and are an infringement to regulations and are inadmissible acc. to IEC 60601-1. Energy Control and Power Lockout for LOGIQ P7/P8/P9/P10. NOTICE When servicing parts of the system where there is exposure to voltage greater than 30 Volts: 1.
  • Page 386 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 8-1 Chapter 8 Contents Section Description Page Number 8-10 8-84 Replacement of Side Tray & Foot Rest Cover 8-11 8-86 Replacement of Monitor and LCD Arm Plastic Covers...
  • Page 387 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 8-1 Chapter 8 Contents Section Description Page Number 8-44 8-232 Replacement of DCWD ASSY 8-45 8-234 Replacement of MCWD ASSY 8-46 8-234 Replacement of battery pack for LP7-P9 R2 ext battery pack...
  • Page 388 ERVICE ANUAL 8-1-2 Returning/Shipping System, Probes and Repair Parts When returning or shipping the LOGIQ P7/P8/P9/P10 system in the original packaging: • system must be lowered to its minimum height with monitor flapped down • the Control Console has to be centered and locked in “unextended” position Equipment being returned must be clean and free of blood and other infectious substances.
  • Page 389 SEPARATELY. PLEASE CONTACT THE MANUFACTURER OR OTHER AUTHORIZED DISPOSAL COMPANY TO DECOMMISSION YOUR EQUIPMENT. Energy Control and Power Lockout for LOGIQ P7/P8/P9/P10. NOTICE When servicing parts of the system where there is exposure to voltage greater than 30 Volts: 1. Turn off the scanner.
  • Page 390 Equipment being returned must be clean and free of blood and other infectious substances. GE Healthcare policy states that body fluids must be properly removed from any part or equipment prior to shipment. GE Healthcare employees, as well as customers, are responsible for ensuring that parts/ equipment have been properly decontaminated prior to shipment.
  • Page 391 Application Software version and the installed Options 8-3-5 Preparation 8-3-5-1 General Notes on LOGIQ P7/P8/P9/P10 Software Loading When preparing to load software, you must perform the following steps: • Back up the entire Patient Archive and Reports (you can import the database from previous version into this version).
  • Page 392 DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL • The Hard Disk Drive (HDD) on LOGIQ P7/P8/P9/P10 is divided into four visible (R1.x.x.) or five (R2.x.x or later) visible partitions. Table 8-2 Use of the Partitions (R1.x.x) Drive Letter Label SYSTEM...
  • Page 393 Peripherals (see: “System - Recording Peripheral settings”). About (see: “System - Recording software and hardware versions”). General, Record preset Region, Language and Regional Options. (See the latest revision of the LOGIQ P7/P9, LOGIQ P8/P9/P10 Basic User Manual)) 8-3-5-2-1-2 Connectivity - Recording the TCP/IP settings 1) From the touch panel, select Utility ->...
  • Page 394 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 8-1 Connectivity - Recording the TCP/IP settings (R1.x.x) Figure 8-2 Connectivity - Recording the TCP/IP settings (R2.x.x or R3.x.x) Figure 8-3 Connectivity - Recording the TCP/IP settings (R4.x.x or later)
  • Page 395 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Record all settings in Table 8-4. Table 8-4 Record settings from TCP/IP screen Parameter Value Computer Name IP Settings Enable DHCP ON / OFF IP-Address Subnet Mask...
  • Page 396 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 8-5 Record settings from WLAN screen PARAMETER VALUE IEEE 802.1x AUTHENTICATION INFORMATION EAP TYPE 8-12 Section 8-3 - System Software - Installation/Upgrade Procedure...
  • Page 397 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-5-2-1-4 Connectivity - Recording the AE Title and Port settings 1) From the touch panel, select Utility -> Connectivity -> Device. 2) Select My Computer in the List of Devices, if not already selected.
  • Page 398 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Record all settings in Table 8-7. Table 8-7 Record settings for each device listed in Device table Name IP Address 8-3-5-2-1-5 Connectivity - Recording Device Destination settings 1) From the touch panel, select Utility ->...
  • Page 399 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-5-2-1-6 Connectivity - Recording Dataflow settings 1) From the touch panel, select Utility -> Connectivity -> Dataflow. Figure 8-7 Connectivity - Recording Dataflow settings 2) Record all settings in Table 8-9 for each Destination Device and Service.
  • Page 400 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-5-2-1-7 Connectivity - Recording the Print Key Assignments 1) From the touch panel, select Utility -> Connectivity -> Buttons. Figure 8-8 Connectivity - Recording the Print Key Assignments...
  • Page 401 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-5-2-1-8 Connectivity - Recording other settings for Print 1) From the Touch Panel, select Utility -> Connectivity -> Buttons. Figure 8-9 Button settings 2) Record all settings in...
  • Page 402 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 8-10 Connectivity - Service settings 2) Record all settings in Table 8-13. Table 8-13 Record settings from Service screen Destination Services (Types) Properties Name Device Name...
  • Page 403 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-5-2-1-10 Connectivity - Recording Miscellaneous settings 1) From the touch panel, select Utility -> Connectivity -> Miscellaneous. Figure 8-11 Connectivity - Recording Miscellaneous settings (R1.x.x or R2.x.x) Figure 8-12 Connectivity - Recording Miscellaneous settings (R3.x.x or later)
  • Page 404 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Record all settings in Table 8-14. Table 8-14 Record settings from Miscellaneous screen Patient/Exam Menu Options ON/OFF or Value Use birthdate Auto search for patient Automatic generation of Patient ID...
  • Page 405 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 8-14 Record settings from Miscellaneous screen Add Titlebar Information to Multiframe Loops (R3.x.x or R4.x.x) Add Scan Parameter Information to Multiframe Loops (R3.x.x or R4.x.x) DICOM Multi-frame image resolution (R1.x.x or R2.x.x) Show progress bar while storing image (R1.x.x or R2.x.x)
  • Page 406 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 8-14 Admin - Recording the Software Option Keys (R3.x.x or later) 2) Record the Software Option Key(s) - alphanumeric string(s) - from the Installed Option Keys field in Table 8-15.
  • Page 407 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-5-2-1-12 Admin - Users 1) From the touch panel, select Utility -> Admin -> Users. Figure 8-15 Admin - Users (R1.x.x or R2.x.x) Figure 8-16 Admin - Users (R3.x.x or later)
  • Page 408 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 8-16 Record settings from Users screen Group User ID Prefix Last Name First Name Op Rights Membership 8-3-5-2-1-13 System - Data Store Management 1) From the touch panel, select Utility -> System-> Backup/Restore.
  • Page 409 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 8-18 System - Data Store Management (R4.x.x or later) 2) Record all settings in Table 8-17. Table 8-17 Record settings from Backup/Restore screen Move Files Older...
  • Page 410 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-5-2-1-14 System - Recording Peripheral settings 1) From the touch panel, select Utility -> System-> Peripherals. Figure 8-19 System - Recording Peripheral settings (R1.x.x.) Figure 8-20 System - Recording Peripheral settings (R2.x.x) Figure 8-21 System - Recording Peripheral settings (R3.x.x or later)
  • Page 411 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Record all settings in Table 8-18. Table 8-18 Record settings from Peripherals screen (R1.x.x) Parameter Value Video Settings - Format Standard Printer Default Printer Setup - Print Full Screen...
  • Page 412 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 8-19 Standard printer and Default printer (R2.x.x. and later) Parameter Value Standard Printer Default Printer Table 8-20 DVR (R2.x.x. and later) Parameter Value Media ____ DVR ____ USB...
  • Page 413 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-5-2-1-16 Location, Region and Language 1) From the touch panel, select Utility -> System-> General. Figure 8-23 Location, Region and Language (R1.x.x) Figure 8-24 Location, Region and Language (R2.x.x) Figure 8-25 Location, Region and Language (R3.x.x or later)
  • Page 414 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 8-22 Record settings from Location Parameter Value Units Machine Description1 (R2.x.x) Machine Description2 (R2.x.x) Dedicated Preset (R3.x.x or later) 8-30 Section 8-3 - System Software - Installation/Upgrade Procedure...
  • Page 415 EVISION ERVICE ANUAL 8-3-6 System Software - Installation Procedure Two separate methods of software installation is offered in LOGIQ P7/P8/P9/P10. A) Software Load from DVD(R1/R2) or USB media (R3/R4/R4.5) B) Software Re-Load from Repository Situation • Harddrive is new • Re-installation for application or operating system issues •...
  • Page 416 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-7 Software Installation from DVD 8-3-7-1 System Software - Installation Procedure (R1.x.x) 1) Prepare the Base System Software DVD. 2) Power ON the scanner and press “DEL”(If AN key is installed) or press left bottom key of trackball (If there’s no AN Key) to enter BIOS...
  • Page 417 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Go to “Boot” in BIOS menu and change the setting value from “SATA 1 Drive” to “SATA 2 Drive” in ‘1st Boot Device’, change the setting value from “SATA 2 Drive” to “SATA 1 Drive” in ‘2nd Boot Device’...
  • Page 418 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 9) The notification screen appears, and then hit the Enter key. NOTE: Use simulator A/N keyboard if customer’s system does not have external A/N keyboard 10) Select either [1] “Load the complete disk”, or...
  • Page 419 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 11) A confirmation window appears, and then hit the Enter key. 12) Approx. 40 minutes later, loading the software will be completed. 13) Remove Base System Software DVD, and type "exit" to reboot the system.
  • Page 420 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 17) Go to “Save & Exit” in BIOS menu and select “Restore Defaults”. 18) Go to “Boot” in BIOS menu and check that ‘1st Boot Device’ is set to “SATA 1” and ‘2nd Boot Device’...
  • Page 421 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 19) Go to “Save & Exit” in BIOS menu and select “Save Changes and Exit”. Then, the system will re- boot. Chapter 8 - Replacement Procedures 8-37...
  • Page 422 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-7-2 System Software - Installation Procedure (R2.x.x) 1) Prepare the Base System Software DVD. 2) Power ON the scanner and press “DEL”(If AN key is installed) or press left bottom key of trackball (If there’s no AN Key) to enter BIOS...
  • Page 423 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Go to “Boot” in BIOS menu and change the setting value from “SATA 1 Drive” to “SATA 2 Drive” in ‘1st Boot Device’, change the setting value from “SATA 2 Drive” to “SATA 1 Drive” in ‘2nd Boot Device’.
  • Page 424 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 9) The notification screen appears, and then hit the Enter key. NOTE: Use the trackball and left set key if the customer’s system does not have an external A/N keyboard.
  • Page 425 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 11) A confirmation window appears, and then hit the Enter key. 12) Approx. 30~40 minutes later, loading the software will be completed. 13) Remove Base System Software DVD, and hit the Enter key to reboot the system.
  • Page 426 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 17) Go to “Save & Exit” in BIOS menu and select “Restore Defaults”. 18) Go to “Boot” in BIOS menu and check that ‘1st Boot Device’ is set to “SATA 1” and ‘2nd Boot Device’...
  • Page 427 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 19) Go to “Save & Exit” in BIOS menu and select “Save Changes and Exit”. Then, the system will re- boot. Chapter 8 - Replacement Procedures 8-43...
  • Page 428 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-7-3 System Software - Installation Procedure (R3.x.x or later) 8-3-7-3-1 Set the system boot order (BIOS setting) 1.) Prepare the THUMB DRIVER R3.x.x, R4.x.x or R4.5.x.x (USB media) including Base Image Load Windows 10.
  • Page 429 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 6.) In “Boot” page, Change “Boot Option #1” to USB memory which is inserted (R3.x.x) or change “1st Boot Device” to USB Harddisk (R4.x.x or later). NOTE: The name of USB memory can be displayed different depend on the inserted USB memory (R3.x.x).
  • Page 430 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7.) After changing the “Boot Option#1” (R3.x.x) or "1st Boot Device" (R4.x.x), save the BIOS set and exit by pressing ‘F10’ key. Figure 8-30 “Boot” page 2 (R3.x.x) Figure 8-31 “Boot”...
  • Page 431 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-7-3-2 Base and Application Software installation • Base Image Software Load Procedure 1.) The system will reboot automatically. 2.) Wait few moments for USB drive to read USB contents.
  • Page 432 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 6.) The Installation progress screen appears like below. Figure 8-34 The notification screen 3 Do not interrupt the software loading at any time. WARNING WARNING 7.) Approximately 20~ 30 minutes (R3.x.x) or 10~ 15 minutes (R4.x.x or later) later, loading the software will be completed.
  • Page 433 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 11) After change the “1st Boot Device”, save the BIOS set and exit by pressing ‘F10’ key. (R4.x.x.) 12) At the system reboot after loading the software, the OS will reboot the system automatically.
  • Page 434 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-7-3-3 Application Software Load Procedure 1.) After the reboot again, the application software installation will start automatically. 2.) About few minutes, the following options will appear. Select your product. Click “Start Installation”...
  • Page 435 Base and Application Software installation for LOGIQ P7/P9 R2.5 to R3 upgraded system 1.) If LOGIQ P7/P9 R3.0.11 System Software or above is being installed to the R2.5 to R3 upgrade R2.5 to R3 upgrade system, follow the below instructions.
  • Page 436 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8.) Click “Save Changes” and then click “Yes”. Figure 8-41 9) MUST force the shutdown by pushing the power button for 5 seconds. 10) After rebooting by forcing the shutdown, reinstall R3.0.11 or upgrade higher version by referring section System Software - Installation Procedure (R3.x.x or later) on page 8-44...
  • Page 437 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-7-4 Application Software Installation Preparation (R1.x.x or R2.x.x) 1) Verify the system boot up to Windows Screen 2) Prepare the Application Software DVD. 3) Press DVD-Drive eject button, and place the Application Software DVD on the tray.
  • Page 438 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-7-5 Application Software Installation (R1.x.x or R2.x.x) 1) Application Software Installation process will take approximately 20 minutes 2) Select your product. Click “Start Installation” button. 3) After the installation completes, the system will be rebooted automatically.
  • Page 439 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5) If SW License window appears, then enter correct Option String and click “OK”. Chapter 8 - Replacement Procedures 8-55...
  • Page 440 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-7-6 Software DVD / USB Storage (R1 to R3) Remove the rear cover and store the software DVD / USB in the system. 1) Remove the rear cover.
  • Page 441 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Store the Software USB STICK in the pocket as below picture.(R3) USB STICK POCKET Figure 8-44 USB stick pocket 8-3-7-7 Software USB Storage (R4 and R4.5) 1) Simply pull out Front Cover.
  • Page 442 8-3-8 Software Installation from Common Service Desktop (R1.x.x or R2.x.x) NOTE: This procedure requires PC with LAN connection capable of reaching to to GE Network 8-3-8-1 Involking Software Reload 1) Select [SERVICE] from [UTILITY] Menu (R1.x.x) or Select [SERVICE] from [UTILITY] Menu (R2.x.x)
  • Page 443 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Verify entering to Common Service Desktop (CSD) 4) Select “Disruptive Mode” under Common Utility, in [Utilities] Tab 5) Acknowledge “Disruptive Mode” on Console Screen 6) Verify Disruptive Mode...
  • Page 444 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8) Click Setup software reload. 9) Select [SYSTEM SHUTDOWN] from Common Utility Menu in CSD 10) Select Restart System Retain Disruptive Mode 11) Press [SUBMIT] 12) On console side, wait until application software loading completes.
  • Page 445 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-9 Software Installation from Common Service Desktop (R3.x.x or R4.x.x) For detail information, please refer to 8-3-9 Software Reload on page 5-91 Chapter 8 - Replacement Procedures...
  • Page 446 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-10 System Setup Procedure to display the messages in Japanese/Chinese (Up to R3.0.8) To setup the system to display the messages in Japanese/ Chinese: NOTE: You cannot enter text in Japanese/Chinese.
  • Page 447 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Under "Current system locale:", select Japanese (Japan) or Chinese (Simplified PRC). Press OK. Figure 8-49 Set Language - example 4) The "Change System Locale" pop-up prompt appears. Select Cancel to not reboot the system yet.
  • Page 448 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7) Open the Administrative tab and check Japanese or Chinese displays under Current language for non-Unicode programs. Figure 8-51 Administrative tab - example 8-64 Section 8-3 - System Software - Installation/Upgrade Procedure...
  • Page 449 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8) Under "Format:", select "Japanese (Japan)"/"Chinese (Simplified PRC)". Press Apply, Press OK. Figure 8-52 Formats 9) In Utility--> System--> General, set the Language as Japanese or Chinese. Press Save and reboot the system using Power on/off button.
  • Page 450 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-3-11 Setting the System’s Language and Keyboard (R3.0.9 or later) 1) Press Utility -> System -> General, then set the Language (restart needed) to the language required. Press Save.
  • Page 451 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 8-56 6) Insert SSA key into console USB port. 7) Enter the Windows maintenance mode. 8) Select Control Panel > Region. The Region pop-up menu appears.
  • Page 452 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-4 Software and Functional Checks after Installation Procedure 8-4-1 Functional Check Table 8-23 Item Test Procedure Reference Note Power ON 4-3-3 Power On/Off [___] Passed [___] Failed...
  • Page 453 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-5 Settings - Backup and Restore This section briefly describes Image Setting backup and restore function. For detailed documentations and warnings, refer to Chapter.16, Section “System/Backup and Restore Preset Menu”...
  • Page 454 For the sake of simplicity, we have used the CD-R in the examples in this section. 8-5-1-2 General Notes NOTE: When you restore backup data from the Utility menu, the LOGIQ P7/P8/P9/P10 application usually restarts automatically when the restoring is complete. NOTE: To perform backup and restore procedures, you must login with administrator privileges.
  • Page 455 6) Select Restore. The system performs the restore. As it proceeds, status information is displayed on the Backup/ Restore screen. 7) The LOGIQ P7/P8/P9/P10 restarts automatically when Restore is done. 8-5-4 EZBackup and EZMove EZBackup or move do not backup patients without images, or reports. If the entire...
  • Page 456 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL EZBackup/EZMove can take up to 20 minutes (or longer, depending on the size of the backup). Make sure to schedule this at the same time daily, when no patients are scheduled.
  • Page 457 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-5-5 Option Keys 1) From the touch panel, select Utility -> System-> Backup/Restore. 2) ADD or REMOVE new option key at “Enter New Option Key” field. Figure 8-58 Option Keys NOTE: This screen shot and options may slightly differ between product type and releases.
  • Page 458 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-6 Replacement of Front Cover 8-6-1 Manpower 1 person, 1 min. 8-6-2 Tools None required. 8-6-3 Removal Procedure 1) Simply pull out Front Cover. [R1, R2, R2.5 and R3] Front Cover [R4 and R4.5]...
  • Page 459 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Front cover is attached using four “latches” as shown below. Figure 8-60 latches 3) Pull out the Air filter from Front Cover. AIR FILTER FRONT Figure 8-61 Removal of Air filter...
  • Page 460 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-6-4 Installation Procedure 1) Front cover and Air filter to be installed in reverse order of removal. 2) Align the edges and simply push to make sure all four “latches” are engaged.
  • Page 461 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-7 Replacement of Cabinet 8-7-1 Manpower 1 person, 5 min. 8-7-2 Tools None required. 8-7-3 Removal Procedure 1) Cabinet shapes differ for each configuration, but removal/installation procedure are the same.
  • Page 462 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-7-4 Installation Procedure 1) Cover to be installed in reverse order of removal. 2) Make sure “latches” on both sides are engaged. NOTE: For Mid and High Cabinet type, there is “Guide Rail” for installation.
  • Page 463 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-8 Replacement of Side Cover 8-8-1 Manpower 1 person, 15 min. 8-8-2 Tools Standard Phillips Screwdriver. 8-8-3 Removal Procedure 8-8-3-1 Removal Procedure - Side Cover L 1) Remove Cabinet.
  • Page 464 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Pull out Side Cover (L). There is mechanical hook on bottom of the part. Figure 8-68 Pull out Side Cover L 8-80 Section 8-8 - Replacement of Side Cover...
  • Page 465 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-8-3-2 Removal Procedure - Side Cover R 1) Remove Cabinet. Figure 8-69 Removal of Cabinet 2) Unscrew 3 or 4 screws of RS Knobs and remove RS Knobs.
  • Page 466 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Unscrew 3 screws for Side Cover R. [R4 and R4.5] [R1, R2, R2.5 and R3] Figure 8-71 Unscrew 3 screws 4) Pull out Side Cover R. There is mechanical hook on bottom of the part.
  • Page 467 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-9 Replacement of Rear Cover 8-9-1 Manpower 1 person, 1 min. 8-9-2 Tools No tool. 8-9-3 Removal Procedure 1) Pull the rear cover and remove it as following figure Figure 8-73 Removal of Rear Cover.
  • Page 468 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-10 Replacement of Side Tray & Foot Rest Cover 8-10-1 Manpower 1 person, 15 min. 8-10-2 Tools Standard Phillips Screwdriver. NOTE: “Short” type driver preferable for Foot Rest Cover removal/installation.
  • Page 469 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-10-3-2 Removal Procedure - Foot Rest Cover 1) Remove Front Cover (refer to section 8-6) 2) Remove Top Cover (refer to section 8-7) 3) Remove Side Cover (refer to section 8-8) 4) Remove 3 screws.
  • Page 470 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-11 Replacement of Monitor and LCD Arm Plastic Covers Be care that the monitor does not fall during the replacement process. WARNING WARNING 8-11-1 Replacement of Monitor and LCD Arm Plastic Covers (R1, R2, R2.5 and R3)
  • Page 471 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Unscrew 5 screws, then remove cable clamp. NOTE: There is cable position marker at red circle. Caution - This Clamp was only applied by opposite direction Figure 8-79 Unscrew cable clamp screws 3) Disconnect three Monitor cables.
  • Page 472 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Unscrew 2 screws, then separate the Monitor from the Monitor Stand Arm. Refer to the figure below. Figure 8-81 Unscrewing 2 screws & separating the Monitor...
  • Page 473 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Remove 2nd Axis Cover. Refer to the figure below. Figure 8-83 Removing 2nd Axis Cover 3) Remove 2nd Arm Cover. Refer to the figure below. Figure 8-84 Removing 2nd Arm Cover...
  • Page 474 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Unscrew 1 screw and Remove the 1st Arm Cover. Refer to the figure below. Figure 8-85 Removing the 1st Arm Cover 8-11-1-7 LCD Arm Plastic Covers Installation Procedure Parts to be installed in reverse order of removal.
  • Page 475 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-11-2 Replacement of Monitor and LCD Arm Plastic Cover (R4 and R4.5) 8-11-2-1 Manpower 2 person, 15 minutes 8-11-2-2 Tools Standard Phillips Screwdriver, Hex Wrench set, Nipper...
  • Page 476 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Unscrew 5 screws, then disconnect three monitor cables. Figure 8-88 Unscrewing 5 screws and disconnecting monitor cables 3) Unscrew 2 screws. Figure 8-89 Unscrewing 2 screws...
  • Page 477 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Unscrew 2 screws only halfway through and do not remove these screws. Figure 8-90 Halfway unscrewing points 5) Push the monitor so that screw holes meet the screw head as following figure and remove the monitor.
  • Page 478 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-11-2-5 Monitor Remove Procedure (R4 and R4.5 HD monitor) 1) Detach the Monitor Cable Cover. Figure 8-92 Removing the monitor cable cover 2) Disconnect the USB cable.
  • Page 479 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Unscrew 3 screws and unlock the cable clamp, then disconnect three monitor cables. Figure 8-94 Unscrewing 3 screws and disconnecting monitor cables 4) Unscrew 2 screws.
  • Page 480 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL monitor. Figure 8-97 Pushing the monitor to remove the monitor 8-11-2-6 Monitor Installation Procedure Parts to be installed in reverse order of removal. 8-96 Section 8-11 - Replacement of Monitor and LCD Arm Plastic Covers...
  • Page 481 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-11-2-7 LCD arm Plastic covers Remove procedure 1) Unscrew 2 screws, then detach Pan Arm Down Cover. Figure 8-98 unscrewing 2 screws and separating the Pan Arm Down Cover 2) Unscrew 2 screws, then detach Lift Arm Cover, L, R.
  • Page 482 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Detaching the Joint Cover. Figure 8-100 Detaching the Joint Cover 4) Unscrew 1 screw, then detach Extension Arm Cover. Figure 8-101 Unscrewing 1 screw and detaching the Extension Arm Cover...
  • Page 483 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-12 Replacement of Monitor Covers 8-12-1 Replacement of Monitor Covers (R1, R2, R2.5 and R3) 8-12-1-1 Manpower 2 persons, 10 minutes. 8-12-1-2 Tools Standard Phillips Screwdriver, Hex Wrench...
  • Page 484 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-12-1-5 Installation Procedure 1) Parts to be installed in reverse order of removal. Make sure not to remain any dust on Glass during Front Cover and Glass Filter Assy...
  • Page 485 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-12-2 Replacement of Monitor Covers (R4 and R4.5) 8-12-2-1 Manpower 2 persons, 10 minutes. 8-12-2-2 Tools Standard Phillips Screwdriver, Hex Wrench, Nipper, Hook Remover 8-12-2-3 Pre-Work 1) LCD Monitor and Monitor Cable Cover removed (refer to section 8-11-2-4)
  • Page 486 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Unscrew 8 screws and remove the Monitor front cover. Figure 8-105 Unscrewing 8 screws 8-12-2-5 Monitor Covers Installation Procedure Parts to be installed in reverse order of removal.
  • Page 487 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-12-3 Replacement of Monitor Covers (R4 and R4.5 HD monitor) 8-12-3-1 Manpower 2 persons, 10 minutes. 8-12-3-2 Tools Standard Phillips Screwdriver, Hex Wrench, Nipper, Hook Remover 8-12-3-3...
  • Page 488 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Unscrew 4 screws and remove the Monitor front cover by using Hook remover. Figure 8-107 Unscrewing 4 screws and remove the Monitor Front Cover 8-12-3-5 Monitor Covers Installation Procedure Parts to be installed in reverse order of removal.
  • Page 489 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-13 Replacement of Fixed Arm Cover and Fixed Arm Holder Grip 8-13-1 Manpower 2 person, 5 minutes. 8-13-2 Tools Standard Phillips Screwdriver, Hex Wrench. 8-13-3 Removal Procedure...
  • Page 490 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-13-4 Installation Procedure Parts to be installed in reverse order of removal. 8-13-5 Functional Check Visual Check only. 8-106 Section 8-13 - Replacement of Fixed Arm Cover and Fixed Arm Holder Grip...
  • Page 491 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-14 Replacement of OPIO and Related Parts NOTE: This section describes removal and installation of OPIO and related parts. OPIO Assy itself, Main PWA Assy, LCD Touch module Assy, HUB Board, Encoder, Freeze key Assy, OPIO AN Key &...
  • Page 492 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-14-1-3-1 Removal Procedure - Trackball Ring NOTE: Removal of Trackball Ring and Cleaning does not require tools. 1) Turn the ring counter clock-wise 2) Lift trackball from OP Panel...
  • Page 493 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Remove 2 screws on OPIO covers. Figure 8-114 Rear of OPIO Base 3) Remove USB cables on Touch panel and unscrew 2 screws. Refer to following figure.
  • Page 494 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Lift OP Panel. Figure 8-116 Lift OP Panel 5) Remove USB locking type Cable 6) Remove OPIO Power cable. 7) Remove USB locking type connector. USB locking type Cable OPIO Power cable GEL &...
  • Page 495 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-14-1-3-3 Removal Procedure - AN Keyboard (Option) 1) Remove 4 screws on AN keyboard Brkt. ( 2 screws on right side and 2 screws on left side) Figure 8-118 Removal of screws 2) Unscrew 2 screws on AN keyboard Brkt only half .
  • Page 496 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Disconnect USB cables as following figure. Figure 8-121 Disconnecting USB cable 8-14-1-3-4 Removal Procedure - Trackball 1) Remove a connector from back of OP Panel Assy.
  • Page 497 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-14-1-3-5 Removal Procedure - LCD Touch Panel Assy 1) Disconnect connectors as following figure. Figure 8-123 Disconnection of LCD Touch panel Assy 2) Remove 4 screws as following figure.
  • Page 498 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-14-1-3-6 Removal Procedure - Freeze key assy 1) Remove connector as following figure Figure 8-125 Disconnection of freeze key Assy connector 2) Remove 3 screws from Freeze key Assy Figure 8-126 Screw points of freeze key assy .
  • Page 499 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Place the new knobs. Figure 8-127 Position of knobs Figure 8-128 Removal of knobs Chapter 8 - Replacement Procedures 8-115...
  • Page 500 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Disconnect freeze key assy connector, trackball connector and Touch panel connectors. Refer to following figure. Touch panel connector Freeze key connector Trackball connector Figure 8-129 Removal of connectors on Main PWA assy...
  • Page 501 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Remove 23 screws around PWA. Refer to following figure. Figure 8-130 Screw points on Main PWA Assy 8-14-1-3-8 Removal Procedure - Encoder on PWA 1) Remove OP Panel Main PWA Assy.
  • Page 502 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Remove 12 screws on 3 BRKT and remove 3 BRKT Refer to following figure. Figure 8-132 Screw points to remove 3 BRKT 8-118 Section 8-14 - Replacement of OPIO and Related Parts...
  • Page 503 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Remove 10 screws on BRKT and remove 1 BRKT Refer to following figure. Figure 8-133 Screw points to remove 3 BRKT 5) Remove flexible connectors and carefully lift up Touch panel.
  • Page 504 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL NOTE: Make sure to connect Touch panel connector to hub board as blow figure. Figure 8-136 Installation of Touch panel connector Make sure not to remain any dust on Glass during OPIO Upper Cover and Touch Screen Assy...
  • Page 505 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-14-2 Replacement of OPIO and Related Parts (R4 and R4.5) 8-14-2-1 Manpower 1 person, 5 minutes 8-14-2-2 Tools Standard Phillips Screwdriver 8-14-2-3 Replacement Procedure Nesting of the parts are as follows.
  • Page 506 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-14-2-3-1 Removal Procedure - Trackball Ring NOTE: Removal of Trackball Ring and Cleaning does not require tools. 1) Turn the ring counter clock-wise 2) Lift trackball from OP Panel...
  • Page 507 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Unscrew 2 screws on OPIO Top Covers Figure 8-140 Unscrewing 2 screws on OPIO Top Covers 3) Disconnect USB cables on Touch panel and unscrew 2 screws.
  • Page 508 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5) Disconnect 4 cables and separate the OPIO assy Figure 8-143 Disconnecting cables 6) Assembly Procedure - OPIO Assy Parts to be installed in reverse order of removal.
  • Page 509 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Unscrew 4 screws and separate OPIO Top assy. Figure 8-145 Unscrewing 4 screws and separating OPIO Top assy 3) Assemble Procedure - OPIO Top assy Parts to be installed in reverse order of removal.
  • Page 510 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-14-2-3-7 Removal Procedure - OPIO Main board 1) Detach 11 knobs from the OPIO Bottom assy. Figure 8-147 Detaching 11 knobs 2) Disconnect freeze key assy connector, trackball connector and Touch panel connectors.
  • Page 511 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Unscrew 20 screws on Main board and detach the board. Figure 8-149 unscrewing 20 screws 4) Assemble Procedure - OPIO main board Parts to be installed in reverse order of removal.
  • Page 512 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-14-2-3-9 Removal Procedure - Joystick encoder on main board 1) Remove OPIO main board (Refer to the section 8-14-2-3-7) 2) Unfasten nut and detach the encoder. Figure 8-151 Unfastening nut and detaching the Joystick encoder 3) Assemble Procedure - Joystick Encoder on main board Parts to be installed in reverse order of removal.
  • Page 513 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-15 Replacement of the Caps for Hardkeys 8-15-1 Manpower 1 person, approx. 1 minute/cap. 8-15-2 Tools Small-sized slotted screwdriver or tweezers. 8-15-3 Removal Procedure 1) By means of a small slotted screwdriver, carefully push against the hardkey cap.
  • Page 514 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-15-5 Functional Check Test Refer to Note System Exterior Visual Check Control panel visual check only section 4-3-1 Power On/Off section 4-3-3 System Integration Checks section 4-3-5-1 OPIO Test only...
  • Page 515 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-16 Replacement of Key Caps (by special native language keys) 8-16-1 Manpower 1 person, 30 minutes. 8-16-2 Tools • Regular Standard Phillips Screwdriver • Screwdriver (Diameter : 3mm)
  • Page 516 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Carefully pull out key caps. Refer to below Figure 8-77 Figure 8-156 Detaching key caps. 8-16-5 Installation Procedure 1) Parts to be installed in reverse order of removal.
  • Page 517 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Run the OPIO Test tool (ScFpApiTest.exe) in the desktop screen and then click the “Op-panel Open” button of test tool. (Program path: C:\scanner\target\bin\ScFpApiTest.exe) Figure 8-157 5) Check on the “Keyboard Test on” check box.
  • Page 518 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-17 Replacement around Nest Box NOTE: This section covers replacement of components related to Nest Box. : MPI,MCB,MDC,MIO,CWD,DCWD,EEPROM,DC4D,Battery,HDD and SOM 8-17-1 Manpower 1 person, 15 minutes for each part.
  • Page 519 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Remove 5 screws and Cable Cover Bracket. Cable Cover Bracket Figure 8-160 Cable Cover Bracket 4) Disconnect Cables. Figure 8-161 Cable Disconnecting points Chapter 8 - Replacement Procedures...
  • Page 520 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-17-5 Removal Procedure - Nest Box NOTE: Before working with Nest box, be sure to remove probes. 1) Remove front cover. (refer to section 8-6) 2) Disconnect 1 fan connector of bottom fan.
  • Page 521 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8) Remove 2 screws on rear side. Figure 8-164 Rear side screw points 9) Pull out Nest Box. Figure 8-165 Nest Box 8-17-6 Installation Procedure - Nest Box.
  • Page 522 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-17-7 Removal Procedure - Board Assy in Nest Box 1) Get Next Box access. 2) Remove 7 screws on EMI bracket. Figure 8-166 Screw points on Nest Box...
  • Page 523 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 8-168 Board assys in Nest Box Chapter 8 - Replacement Procedures 8-139...
  • Page 524 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-17-8 Removal Procedure - Separating MPI and MCB 1) Unscrew 6 screws. Refer to following figure. Figure 8-169 Screw points on MPI assy 8-140 Section 8-17 - Replacement around Nest Box...
  • Page 525 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Carefully separate MPI assy from MCB assy. Refer to following figure. [R1, R2, R2.5 and R3] [R4 and R4.5] Figure 8-170 Lifting up MPI and position of Stacking up connector on MPI assy...
  • Page 526 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-17-9 Removal Procedure - Separating DCWD Board(Option) from MPI 8-17-9-1 R1, R2, R2.5 and R3 Philips screw-driver required to add this part. 1) Remove 3 screws on MPI assy.
  • Page 527 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-17-9-2 R4 and R4.5 1) Separating MPI from MCB assy : refer to section 8-17-8 Figure 8-173 Separated MPI assy 2) Remove 3 screw on bottom side of MPI assy...
  • Page 528 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Lift up DCWD board from MPI assy Figure 8-175 Lift up DCWD board Figure 8-176 Separated DCWD board 8-144 Section 8-17 - Replacement around Nest Box...
  • Page 529 The latest version of the BIOS is “GEMARUR9”. 2) Preparing BIOS update application and MARU BIOS binary file 1.) Need to download BIOS update application and MARU BIOS binary file. Access “SOM_BIOS_Revision_tool & MARU BIOS binary file” in the GE BOX. https://ge.box.com/s/2yr0onbu7z8iq0fyd5wgs580vgi2tr5y 2.) Download two files - BIOS update application : CGUTL_CGOS.zip...
  • Page 530 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3.) Copy the BIOS update application and the latest BIOS file. Figure 8-178 4.) Extract BIOS update application file, and run “Setup.exe” file in the folder Cgos.
  • Page 531 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5.) If the installation is complete, the program will be shown in the installed program list under “Program and Features”. The program name is 'congatec CGOS API”.
  • Page 532 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2.) Select “Board (CGOS)” and “BIOS Update”. Figure 8-182 3.) Select “Select BIOS file”. Figure 8-183 8-148 Section 8-17 - Replacement around Nest Box...
  • Page 533 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4.) Find the latest BIOS file and select it and “Open”. Figure 8-184 5.) Select “Update BIOS”. Figure 8-185 Chapter 8 - Replacement Procedures 8-149...
  • Page 534 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 6.) After updating the BIOS, it will be shown like below. Check if the Current OEM BIOS version is “GEMARU 9”. Then, select “Close”. Figure 8-186 7.) Select “Close”.
  • Page 535 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5) Deleting BIOS update application and the BIOS file. 1.) Enter “Programs and Features” in Control Panel. Select “Congatec CGOS API” and “Uninstall/Change”. Figure 8-188 2.) Delete copied files permanently. Select those files and push [Shift] + [Ctrl] + [Delete] keys all together at the same time.
  • Page 536 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3.) Turn the system off. Select ‘Start’ and ‘Shut down’ 6) Checking BIOS update completion 1.) Turn the system on by pushing the power button. 2.) Enter the BIOS mode by pushing Delete button or Trackball left-side button.
  • Page 537 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-17-10-2 Replacement Procedure 1) Remove 7 screws around MDC assy. Refer to following figure. Figure 8-191 Screw points around MDC assy 2) Carefully separate MDC assy from MCB assy. Refer to following figure.
  • Page 538 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-17-11 Handling EEPROM on MCB Board EEPROM on MCB board contains vital information. A) System Serial Number that is matched against stored software option key. B) ComExpress Functional REV that is monitored for sub-system compatibility.
  • Page 539 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Remove EEPROM from replacement MCB board if present. 4) Verify direction of the EEPROM Slot by aligning EEPROM “notch” mark to that of MCB board. print mark.
  • Page 540 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-17-12 Separating MCB assy from Brackets and COM Express 1) Remove 8 screws around MCB assy. Refer to following figure. Figure 8-196 Screw points around MCB assy...
  • Page 541 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Remove MCWD board from MCB assy. Refer to following figure. NOTE: Be careful not to damage CWD connector. MCWD board assy Figure 8-197 Removal of MCWD board.
  • Page 542 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-17-13 Separating DC4D Unit from IO BOX 1) Unscrew 12 screws to remove the IO BOX Cover. Figure 8-199 Screw points of IO BOX cover 2) Unscrew five screws to remove the DC4D from the MIO Assy. Refer to following figure.
  • Page 543 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-17-14 Separating Battery Unit from IO BOX 1) Unscrew 12 screws to remove the MIO Cover. (refer to section 8-17-13) 2) Unscrew 8 screws and 2 Battery connectors. Refer to following figure.
  • Page 544 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-17-15 Separating HDD (R1, R2, R2.5) or SSD (R3) 1) Open the front cover. (refer to section 8-6) 2) Remove 4 screws around HDD Bracket Refer to following figure.
  • Page 545 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-17-16 Separating M.2 SSD (R4 and R4.5) 1) Take the board assy out of the nest box (Refer to section 8-17-7) 2) Remove eight screws (check red-designating circles out) and separate MCB from the assy.
  • Page 546 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-17-17 Separating COM Express 8-17-17-1 Separating COM Express (R1, R2, R2.5 and R3) 1) Unscrew 5 screws on bottom side of MCB board assy. Refer to following figure.
  • Page 547 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-17-17-2 Separating COM Express (R4 and R4.5) 1) Disconnect fan cable. Figure 8-209 Disconnecting fan cable 2) Unscrew five male-type-screws. Figure 8-210 Unscrewing five male type screws...
  • Page 548 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Detach COM-Express module. Figure 8-211 Detaching COM-Express 4) Unscrew four male-type-screws in the side of COM-Express module. And remove SOM Bracket from the COM-Express module. Figure 8-212 Unscrewing 4 male-type-screws...
  • Page 549 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-17-18 Installation Procedure - MPI, MCB, MDC, DC4D, HDD, Battery and COM Express 1) Parts to be installed in reverse order of removal. 8-17-19 Functional Check - MPI, MCB, MDC, IO Box...
  • Page 550 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-17-20-2 Removal Procedure - MBP assy (R4 and R4.5) 1) Remove Nest Box from system 2) Remove MCB, MDC< MPI boards assy and IO BOX 3) Remove four screws and fan connector...
  • Page 551 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5) Remove stud of CSI and six screws (red-designating circle) 6) Pull MBP4 out of system Figure 8-217 8-17-21 Installation Procedure - MBP assy Parts to be installed in reverse order of removal.
  • Page 552 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-18 Replacement of Front fan 8-18-1 Manpower 1 Person, 10 minutes. 8-18-2 Tools Standard Phillips Screwdriver. 8-18-3 Removal Procedure 1) Remove the Front Cover from system. (refer to section 8-6) 2) Remove 4 screws and disconnect fan connector.
  • Page 553 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-19 Replacement of Bottom fan 8-19-1 Manpower 1 Person, 10 minutes. 8-19-2 Tools Hex Wrench. 8-19-3 Removal Procedure 1) Remove the Front Cover from system. (refer to section 8-6) 2) Remove the nest Box from system.
  • Page 554 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Place bottom fan cable not to be entangled on system frame as following figure Figure 8-220 Placing bottom fan connector 5) Pull out bottom fan assy as following figure...
  • Page 555 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 6) Remove 4 screws on the fan. Figure 8-222 Screw points of bottom fan. 8-19-4 Installation Procedure Parts to be installed in reverse order of removal. 8-19-5 Functional Check 1) After turn on the system, confirm the fan rotating by feeling wind at the bottom fan dust filter.
  • Page 556 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-20 Replacement of Rear Handle 8-20-1 Manpower 1 Person, 10 minutes. 8-20-2 Tools Standard Phillips Screwdriver and 6mm Hex Wrench. 8-20-3 Removal Procedure 1) Remove 1 screw and remove the Link Arm Top Cover as following figure.
  • Page 557 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-20-5 Functional Check Visual Check only. Chapter 8 - Replacement Procedures 8-173...
  • Page 558 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-21 Replacement of Link Arm Cable Hook 8-21-1 Manpower 1 Person, 10 minutes. 8-21-2 Tools Standard Phillips Screwdriver. 8-21-3 Removal Procedure 1) Remove 1 screw and remove the Link Arm Top Cover as following figure.
  • Page 559 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-22 Replacement of Caster 8-22-1 Manpower 2 persons, 15 minutes. 8-22-2 Tools 24mm Socket or adjustable Wrench, Standard Phillips Screwdriver NOTE: “Short” type driver preferable for Foot Rest Cover removal/installation.
  • Page 560 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL NOTE: This does not require any tools. Be careful not to damage Hand during pulling hook of the cap Figure 8-228 Caster Cap and hook points 4) Remove Caster Cap by slightly pulling hook of the cap to outside and then move upward.
  • Page 561 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 6) Remove Nut and replace Caster. Figure 8-230 Removable Nut of Caster Proceed with extreme caution and avoid system tipping over. The system weighs CAUTION approximately 60kg.
  • Page 562 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-23 Replacement of the COM Express 8-23-1 Manpower 1 Person, 5 minutes. 8-23-2 Tools Standard Phillips Screwdriver. 8-23-3 Removal Procedure 8-23-3-1 Removal Procedure (R1, R2, R2.5 and R3) 1) Separate MCB Assy and MDC Assy.
  • Page 563 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Unscrew 5 screws on bottom side of MCB board assy. Refer to following figure. Figure 8-232 Screw points for removal of COM Express. 4) Disconnect fan cable and carefully detach COM Express module. Refer to following figure.
  • Page 564 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-23-3-2 Removal Procedure (R4 and R4.5) 1) Disconnect fan cable. Figure 8-235 Disconnecting fan cable 2) Unscrew five male-type-screws. Figure 8-236 Unscrewing five male type screws 8-180...
  • Page 565 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Detach COM-Express module. Figure 8-237 Detaching COM-Express 4) Unscrew four male-type-screws in the side of COM-Express module. And remove SOM Bracket from the COM-Express module. Figure 8-238 Unscrewing 4 male-type-screws...
  • Page 566 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-23-4 Installation Procedure Install components in reverse order of removal. When replacing with new SOM, Push the SOM in both side of MCB Assy like as below Figure CAUTION 8-130.
  • Page 567 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Type “maru” for password. 3) After successfully entering BIOS, verify the following parameters Table 8-24 Configuring BIOS. top menu Main System Time System Date 4) Change date/time parameters as needed, and save the changes.
  • Page 568 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-23-6-1 Accessing EEPROM 1) Exit to Desktop (Refer to section 5-10 for details) 2) Start VPD Editor by clicking icon on the desktop. (“Vpdedit”) 3) Select COMEXP from device grid.
  • Page 569 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL through the VPD contents), then press “NO” before continuing. 8-23-6-2 Initializing EEPROM When a new EEPROM is attached to a system, it will require initialization. 1) Select “COMEXP” from device grid. (right pane).
  • Page 570 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL The file can be found under c:\Scanner\target\resources\Acqusition\Housekeeping 4) Click “Open” to load the selected template. NOTE: This procedure can be applied to other components with VPD, for example, for initializing MPI VPD.
  • Page 571 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-23-7 Functional Check Test Refer to Note Board Diagnostics section 4-4 Chapter 8 - Replacement Procedures 8-187...
  • Page 572 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-24 Replacement of the HDD/SSD 8-24-1 Manpower 1 Person, 5 minutes. 8-24-2 Tools Standard Phillips Screwdriver. 8-24-3 Removal Procedure 1) Open the front cover. (refer to section 8-6) 2) Remove 4 screws around HDD/SSD Bracket Refer to following figure.
  • Page 573 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Unscrew 4 screws to remove HDD/SSD bracket. Refer to following figure. Figure 8-243 Removal of HDD/SSD bracket 8-24-4 Installation Procedure Install components in reverse order of removal.
  • Page 574 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-25 Replacement of the M.2 SSD 8-25-1 Manpower 1 person, 20 minutes 8-25-2 Tools Standard Phillips Screwdriver. 8-25-3 Removal Procedure 1) Take the board assy out of the nest box (Refer to section 8-17-7) 2) Remove eight screws (check red-designating circles out) and separate MCB from the assy.
  • Page 575 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-25-4 Installation Procedure Install components in reverse order of removal. 8-25-5 Functional Check Refer to section 4-3-3 for Power On/Off test. Chapter 8 - Replacement Procedures 8-191...
  • Page 576 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-26 Replacement of Battery 8-26-1 Manpower 1 person, 1 hour NOTE: LP9/P7 BUM 5504583 provides the information, "Refreshing the battery" in Chapter 3 Preparing to System for Use. The recommended procedure can be executed by user for maintaining battery life.
  • Page 577 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Unscrew 4 screws and disconnect 2 connectors from Battery. Refer to following figure. Figure 8-247 Screw points and Connectors on battery assy 3) Unscrew 4 screws to remove Battery bracket...
  • Page 578 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-27 Replacement of 4D 8-27-1 Manpower 1 Person, 1 hour. 8-27-2 Tools Standard Phillips Screwdriver. 8-27-3 Removal Procedure 1) Unscrew 12 screws to remove the IO BOX Cover.
  • Page 579 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Unscrew five screws to remove the DC4D from the IO BOX. Refer to following figure. Figure 8-250 Screw points of DC4D Assy 3) Carefully lift up the DC4D unit.
  • Page 580 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-28 Replacement of Wireless LAN 8-28-1 Manpower 1 Person, 5 minutes. 8-28-2 Tools None Required. 8-28-3 Removal Procedure 1) Remove Top Cover (refer to section 8-7) 2) Attach loop sheet on Wireless LAN dongle.
  • Page 581 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-29 Replacement of the Harness Cable and OPIO Cable Assy 8-29-1 Replacement of the Harness Cable and OPIO Cable Assy (R1, R2, R2.5 and R3) 8-29-1-1...
  • Page 582 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-198 Section 8-29 - Replacement of the Harness Cable and OPIO Cable Assy...
  • Page 583 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 8-253 Removal of Harness Cable Assy 8-29-1-5 Removal Procedure (OPIO Cable Assy) 1) Unscrew 1 screw, remove cable clamp, and then loosen cable through hole. NOTE: There is cable position marker at this location.
  • Page 584 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-29-1-7 Functional Check Test Refer to Note System Exterior Visual Check Control panel visual check only section 4-3-1 Power On/Off section 4-3-3 Visual check on LCD monitor to see if image displays without color distortion Check OSD Control function operates.
  • Page 585 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-29-2-4 Removal Procedure (Monitor cables) 1) Cut the 6 cable ties Figure 8-255 Cutting 6 cable ties 2) Extract 3 monitor cables through the LCD Arm hole.
  • Page 586 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Unscrew 1 screw and detach the LCD arm. Figure 8-257 Unscrewing 1 screw and detach the LCD arm 8-202 Section 8-29 - Replacement of the Harness Cable and OPIO Cable Assy...
  • Page 587 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Extract 3 monitor cables through the LCD Arm hole. Figure 8-258 Extracting 3 monitor cables 5) Extract 3 monitor cables as the picture below. Untying the cable clamp...
  • Page 588 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-29-2-5 Removal Procedure (OPIO cables) 1) Unscrew 1 screw and extract the 2 cables. Figure 8-259 Unscrewing 1 screw and extract 2 cables through the hole 2) Unscrew 2 screws and remove OPIO cables.
  • Page 589 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Assemble Procedure - OPIO Cables Parts to be installed in reverse order of removal. 8-29-2-6 Functional Check Test Refer to Note System Exterior Visual Check Control panel visual check only...
  • Page 590 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-30 Replacement of Peripheral Cable Assy 8-30-1 DVD Cables 8-30-1-1 Manpower 1 person, 5 minutes. 8-30-1-2 Tools Nipper, 2 Cable ties (GEPN: 2160955). 8-30-1-3 Pre work 1) Remove Rear Cover.
  • Page 591 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Disconnect 2 DVD cables from the rear panel. Refer to following figure. Figure 8-263 DVD cable connectors 4) Disconnect 2 DVD cables from the DVD. Figure 8-264 Removal of DVD cables from the DVD 5) Carefully pull out the cables.
  • Page 592 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-30-1-5 Installation Procedure Parts to be installed in reverse order of removal. But, refer to below procedure for cable ties. 2 holes for fixing cable tie 2 holes for fixing cable tie...
  • Page 593 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-30-2 B/W Printer Cables 8-30-2-1 Manpower 1 person, 5 minutes. 8-30-2-2 Tools Standard Phillips Screwdriver. 8-30-2-3 Pre work 1) Remove Rear Cover. (refer to section 8-9) 2) Remove Top Cover (refer to section 8-7) 3) Remove Cable Cover Bracket.
  • Page 594 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Unlock cable clamp. Refer to following figure. Figure 8-268 Cable clamp 8-210 Section 8-30 - Replacement of Peripheral Cable Assy...
  • Page 595 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Disconnect 2 BW printer cables from the rear panel. Refer to following figure. Figure 8-269 BW printer connectors 5) Disconnect 2 BW printer cables from printer and carefully pull out the cables.
  • Page 596 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-30-2-5 Installation Procedure Parts to be installed in reverse order of removal. But, refer to below procedure for cable ties. 2 holes for fixing cable tie Align the head of cable tie as shown...
  • Page 597 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-31 Replacement of the Probe Holder 8-31-1 Manpower 1 Person, 5 minutes. 8-31-2 Tools No tool is required. 8-31-3 Removal Procedure 1) Simply pull out Probe Holder from its location.
  • Page 598 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-32 Replacement of the Cable Hook 8-32-1 Manpower 1 Person, 5 minutes. 8-32-2 Tools Standard Phillips Screwdriver. 8-32-3 Removal Procedure 1) Remove screws as following figure and carefully remove cable hook.
  • Page 599 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-33 Replacement of the OPIO Tray 8-33-1 Manpower 1 Person, 5 minutes. 8-33-2 Tools Standard Phillips Screwdriver. 8-33-3 Removal Procedure 1) Remove screw as following figure and carefully remove OPIO Tray.
  • Page 600 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-34 Replacement of the Paper tray 8-34-1 Manpower 1 Person, 1 minutes. 8-34-2 Tools None Required. 8-34-3 Removal Procedure 1) Remove Paper tray from Top Cover.
  • Page 601 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-35 Replacement of AC Cable Holder 8-35-1 Manpower 1 Person, 5 minutes. 8-35-2 Tools Standard Phillips Screwdriver. 8-35-3 Removal Procedure 1) Remove 2 Screws from AC Cable Holder...
  • Page 602 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-36 Replacement of the BW printer 8-36-1 Manpower 1 person, 10 minutes 8-36-2 Tools Standard Phillips Screwdriver. 8-36-3 Pre work 1) Remove Rear Cover. (refer to section 8-9) 2) Remove Top Cover (refer to section 8-7) 3) Remove Cable Cover Bracket.
  • Page 603 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Pull out BW printer as following figure. Figure 8-279 Pulling out the BW printer 3) Remove 4 screws on BW printer bottom part. Refer to following figure.
  • Page 604 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-37 Replacement of the Speaker 8-37-1 Manpower 1 Person, 10 minutes. 8-37-2 Tools Standard Phillips Screwdriver, Hex Wrench, Nipper, 1 Cable ties (GEPN: 2160955). 8-37-3 Pre work 1) Remove Rear Cover.
  • Page 605 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Remove the cable tie. Figure 8-282 Cable tie point 3) Unlock cable clamp. Refer to following figure. Figure 8-283 Cable clamp 4) Remove 4 screws around speaker assy. Refer to following figure.
  • Page 606 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5) Remove 4 screws, and then remove speaker brkt. Refer to following figure. Figure 8-285 4 screw points to remove speaker brkt 6) Remove 8 screws and carefully pull out speaker assy. Refer to following figure.
  • Page 607 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-37-5 Installation Parts to be installed in reverse order of removal. But, refer to below procedure for cable ties. 2 holes for fixing cable tie Cable tie...
  • Page 608 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-38 Replacement of the Monitor Arm 8-38-1 Manpower 1 person, 15 minutes. 8-38-2 Tools Standard Phillips Screwdriver, Nipper, 4 mm Hex Wrench, 3 Cable ties (GEPN: 2160955).
  • Page 609 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-39 Replacement of Front Fan Filter 8-39-1 Manpower 1 Person, 2 minutes. 8-39-2 Tools None Required. 8-39-3 Removal Procedure 1) Remove Front Cover. (refer to section 8-6) 2) Pull out Filter upward.
  • Page 610 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-40 Replacement of Bottom Fan Filter 8-40-1 Manpower 1 Person, 3 minutes. 8-40-2 Tools None Required. 8-40-3 Removal Procedure 1) Pull out bottom fan filter as following figure.
  • Page 611 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-41 Replacement of SW CD set (R1, R2, R2.5 only) and SW USB (R3, R4 and R4.5 only) 8-41-1 Manpower 1 Person, 3 minutes. 8-41-2 Tools None Required.
  • Page 612 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-42 Replacement of Drawer 8-42-1 Manpower 1 Person, 5 minutes. 8-42-2 Tools Standard Phillips screwdriver. 8-42-3 Removal Procedure 1) Pull out the Top cover. Figure 8-290 Drawing out Top cover.
  • Page 613 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-42-4 Installation Procedure Parts to be installed in reverse order on removal. 8-42-5 Functional Check Visual Check Only Chapter 8 - Replacement Procedures 8-229...
  • Page 614 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-43 Replacement of ECG 8-43-1 Manpower 1 Person, 5 minutes. 8-43-2 Tools Standard Phillips screwdriver. 8-43-3 Removal Procedure 1) Remove OPIO covers (refer to section 8-14) 2) Remove cables on ECG Assy (square) and unscrew 4 screws (circle).
  • Page 615 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5) Unscrew 4 screws and remove ECG BRKT Figure 8-294 ECG screw points. 8-43-4 Installation Procedure Parts to be installed in reverse order on removal. 8-43-5 Functional Check...
  • Page 616 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-44 Replacement of DCWD ASSY 8-44-1 Manpower 1 person, 15 minutes 8-44-2 Tools Standard Phillips screwdriver. 8-44-3 Removal Procedure 8-44-3-1 Removal Procedure - R1 to R3 1) Separate MPI assy from MCB and MDC.
  • Page 617 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-44-3-2 Removal Procedure - R4/R4.5 1) Separate MPI assy from MCB and MDC. Refer to the section 8-17-8 Removal Procedure - Separating MPI and MCB. 2) Remove 3 screws on MPI assy.
  • Page 618 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-45 Replacement of MCWD ASSY 8-45-1 Manpower 1 person, 15 minutes 8-45-2 Tools Standard Phillips Screwdriver. 8-45-3 Removal Procedure 1) Separate MCB Assy and MDC Assy. (Refer to section 8-17-7)
  • Page 619 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-46 Replacement of battery pack for LP7-P9 R2 ext battery pack 8-46-1 Manpower 1 person, 30 minutes 8-46-2 Tools Standard Phillips Screwdriver, Hex-Wrench 8-46-3 Removal Procedure 1) Unscrew 2 screws and then pull out the BATTERY PACK COVER BRKT LP7_P9 from 6 points of LATCH.
  • Page 620 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Disconnect BATTERY SIG Harness and PWR Harness. PWR Harness SIG Harness Figure 8-302 Disconnecting BATTERY Harnesses 4) Unscrew 4 screws. Figure 8-303 Unscrewing 4 screws 8-236...
  • Page 621 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5) Disassemble LP7-P9 R2 EXT BATTERY PACK from extrude shape of BATTERY PACK FRAME BRKT. Figure 8-304 Removing LP7-P9 R2 EXT BATTERY PACK 6) Unscrew 8 screws to remove BATTERY PACK TOP BRKT.
  • Page 622 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7) Lift up POLARIS BATTERY PACK and disconnect BATTERY Harness from PWA. Then remove POLARIS BATTERY PACK. NOTE: Once replacing battery pack, all of battery packs (4 pcs) need to be replaced in same time.
  • Page 623 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Boot-up the console using power button on OPIO ASSY. Figure 8-308 Booting-up using power button 3) Check that battery icon is displayed on left bottom side.
  • Page 624 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-47 Replacement of battery pack for LP7-P9 R2 Battery option 8-47-1 Manpower 1 person, 30 minutes 8-47-2 Tools Standard Phillips Screwdriver, Hex-Wrench, Socket driver 8-47-3 Removal Procedure 1) Unscrew 2 screws to remove POWER CORD FIX BRKT.
  • Page 625 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Unscrew 7 screws to remove the NEST EMI BRKT and pull out MIO ASSY from system. Figure 8-312 Unscrewing 7 screws to remove NEST EMI BRKT and MIO ASSY 4) Unscrew 12 screws to remove the MIO BOX BRKT.
  • Page 626 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5) Disconnect 2 connectors of BATTERY and unscrew 4 screws to remove BATTERY ASSY. Figure 8-314 Removing BATTERY ASSY 6) Unscrew 4 screws to remove BATTERY TOP BRKT. Then remove POLARIS BATTERY PACK.
  • Page 627 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-47-5 Functional Check 1) Turn on the circuit breaker. Figure 8-316 Turning on circuit breaker 2) Boot-up the console using power button on OPIO ASSY. Figure 8-317 Booting-up using power button 3) Check that battery icon is displayed on left bottom side.
  • Page 628 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-48 Replacement of MBOB 8-48-1 Manpower 1 person, 20 minutes 8-48-2 Tools Standard Phillips Screwdriver 8-48-3 Removal Procedure 1) Remove POWER CORD FIX BRKT and POWER CORD.
  • Page 629 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Unscrew 5 screws and remove REAR CABLE COVER BRKT. REAR CABLE COVER BRKT Figure 8-321 Unscrewing 5 screws 4) Remove CABLES (RED CIRCLE) and unscrew 7 screws, then remove NEST EMI BACK BKRT.
  • Page 630 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5) Draw IO BOX from NEST FRAME. Figure 8-323 Drawing IO BOX from NEST FRAME 6) Unscrew 12 screws and remove MIO BOX BRKT. Figure 8-324 Unscrewing 12 screws...
  • Page 631 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7) Disconnect PWR and SIG connectors. Figure 8-325 Disconnecting PWR and SIG connectors 8) Disassemble BATTERY ASSY. Figure 8-326 Disassembling BATTERY ASSY 9) Disassemble MBOB PWR HARNESS (5728317) and MBOB SIG HARNESS (5728318)
  • Page 632 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 10) Disassemble MBOB PWA from MIO-B PWA Figure 8-328 Disassembling MBOB PWA 8-48-4 Installation Procedure All components to be installed in reverse order of removal. 8-48-5 Functional Check Refer to section 4-4 for Functional Diag Test List.
  • Page 633 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-49 Replacement of MIO-B 8-49-1 Manpower 1 person, 20 minutes 8-49-2 Tools Standard Phillips Screwdriver 8-49-3 Removal Procedure 1) Unscrew 12 screws and remove MIO BOX BRKT.
  • Page 634 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Disassemble ACFL BOARD FIX BRKT(5538682). Figure 8-330 Disassembling ACFL BOARD FIX BRKT(5538682) 3) Disassemble Nut from the IOBOX-B Chassis Bracket Figure 8-331 Disassembling Nut from the IOBOX-B Chassis Bracket...
  • Page 635 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Disassemble the following screws Figure 8-332 Disassembling the following screws 8-49-4 Installation Procedure All components to be installed in reverse order of removal. 8-49-5 Functional Check Refer to section 4-4 for Functional Diag Test List.
  • Page 636 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-50 Replacement of BBC 8-50-1 Manpower 1 person, 30 minutes 8-50-2 Tools Standard Phillips Screwdriver, Hex-Wrench 8-50-3 Removal Procedure 1) Unscrew 2 screws to remove COVER BRKT.
  • Page 637 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Disassemble CONNECTOR BRKT using 2 screws. (2327596, SCREW WH M3X6) Figure 8-335 Disassembling CONNECTOR BRKT using 2 screws 4) Disconnect BATTERY SIG HARNESS and PWR HARNESS.
  • Page 638 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 5) Unscrew LP7-P9 R2 EXT BATTERY PACK from BATTERY PACK FRAME BRKT LP7_9 using 4 screws (5431313, SCREW HSH M5X10). Figure 8-337 Unscrewing LP7-P9 R2 EXT BATTERY PACK from BATTERY PACK FRAME BRKT LP7_9...
  • Page 639 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 6) Pull out LP7-P9 R2 EXT BATTERY PACK from BATTERY PACK FRAME BRKT LP7-P9 R2 EXT BATTERY PACK Figure 8-338 Pulling out LP7-P9 R2 EXT BATTERY PACK from BATTERY PACK FRAME BRKT...
  • Page 640 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 7) Unscrew 8 screws to remove BATTERY PACK TOP BRKT. BATTERY PACK TOP BRKT Figure 8-339 Unscrewing 8 screws to remove BATTERY PACK TOP BRKT 8) Remove 4 BATTERY PACKS.
  • Page 641 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 9) Disassemble BBC PWA Figure 8-341 Disassembling BBC PWA 8-50-4 Installation Procedure All components to be installed in reverse order of removal. 8-50-5 Functional Check Refer to section 4-4 for Functional Diag Test List.
  • Page 642 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-51 Replacement of ROTATION MODULE KIT 8-51-1 Replacement of LOGIQ P9 ROTATION MODULE KIT (R1 to R3) 8-51-1-1 Manpower 1 person, 30 minutes 8-51-1-2 Tools Standard Phillips Screwdriver, Nipper, Torque wrench: Needed HEX wrench - 4mm HEX wrench (For M5 HEX Screw), Needed force - 60 Kgf·cm (For screw tighten force), Monkey Spanner (For NUT M4),...
  • Page 643 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4.) Unscrew nut and unfix the cable wire from the UP DOWN LINK BRKT. 5.) Unscrew nut and unfix the cable wire from the CABLE WIRE FIX BRKT.
  • Page 644 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8.) Unscrew 3 screws and remove the TURN BUCKLE SET and SPRING from the ROTATION MODULE ASSY. ROTATION MODULE TURN BUCKLE SET SPRING 9.) Unscrew 3 screws and remove the ROTATE LEVER BRKT from OPIO base.
  • Page 645 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL When reassemble the ROTATION MODULE ASSY, HSH screw should be tightened proper CAUTION torque (60kgf-cm) When reassemble the screws, please use SCREW HSH M5X16 NYLOK(5729013). CAUTION 12.) Unscrew 6 screws and carefully remove the OPIO base.
  • Page 646 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 14.) Each screws should be heated for 5 minutes by HOT AIR BLOWER or ELECTRIC HAIR DRYER to melt LOCTITE on screw before unscrew the screw. ( Caution hot surfaces 15.) Unscrew 7 screws and then remove the ROTATION PLATE.
  • Page 647 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 16.) Remove the THRUST NEEDLE BEARING WASHER 5578, THRUST NEEDLE BEARING 5578, OPIO SWIVEL SHAFT from the UP DOWN LINK ARM in order. Do not wipe out grease from part.
  • Page 648 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-51-2 Replacement of LOGIQ P8/P9/P10 ROTATION MODULE KIT (R4 and R4.5) 8-51-2-1 Manpower 1 person, 30 minutes 8-51-2-2 Tools Standard Phillips Screwdriver, Nipper, Torque wrench : needed HEX wrench – 4mm HEX wrench (For M5 HEX Screw), Needed force –...
  • Page 649 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Unscrew 1 screw and disassemble the cable wire from the bracket. Figure 8-343 Unscrewing 1 screw and disassemble the cable wire 3) Unscrew nut and disassemble the cable wire from the bracket with the arrow direction.
  • Page 650 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4) Unlock cable clamp, and then extract 5 cables and cable wire from the Hole. Figure 8-345 5) Unscrew 3 screws and remove the Turn Buckle Set and Spring.
  • Page 651 When reassemble the Rotation Module Assy, screw should be tightened with 60Kgf·cm torque CAUTION and use the Screw HSH M5X16 Nylok (GE Part no.: 5729013) for reassembly. 7) Unscrew 6 screws and detach the OPIO Base. Figure 8-348 Unscrewing 6 screws and detaching OPIO Base When reassemble the OPIO Base, screw should be tightened with 60Kgf·cm torque and use...
  • Page 652 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8) Detach the IGUS Thrust Bush. Figure 8-349 Detaching IGUS Thrust Bush DO NOT wipe out grease from part. CAUTION 9) Each screw should be heated for 5 minutes by Hot Air Blower or Electric Hair Dryer to melt LOCTITE on screw before unscrewing 6 screws.
  • Page 653 Figure 8-351 Unscrewing 7 screws and removing Rotation Plate When reassemble the ROTATION PLATE, screw should be tightened with 60Kgf·cm torque and CAUTION use SCREW FSH M5X16 NYLOK (GE Part no. : 5729014) for reassembly. Carefully unscrew to prevent wear of screw head. CAUTION 11) Remove the Thrust Needle Bearing Washer, Thrust Needle Bearing, OPIO Swivel Shaft.
  • Page 654 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-51-2-5 Installation Procedure Parts to be installed in reverse order on removal. But Rotation Plate, OPIO Swivel Shaft. Rotation Module Assy should be reassembled using new parts.
  • Page 655 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-52 Replacement of ODD Dummy Cover R3 8-52-1 Manpower 1 person, 5 minutes 8-52-2 Tools Standard Phillips Screwdriver 8-52-3 Pre work 1) Remove Rear Cover (refer to section 8-9)
  • Page 656 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-53 Replacement of Probe Light assy 8-53-1 Replacement of Probe Light Assy (R3) 8-53-1-1 Manpower 1 person, 20minutes 8-53-1-2 Tools Standard Phillips Screw driver 8-53-1-3 Pre work...
  • Page 657 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Untie clamp for release Probe light cable Figure 8-355 Un-tie of Probe light cable from OPIO base 8-53-2 Replacement of Probe Light Assy (R4 and R4.5)
  • Page 658 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-53-2-4 Removal Procedure 1) Disconnect the probe light cable. Figure 8-356 Disconnecting the probe light cable 2) Unscrew 1 screw for untying the probe light cable. Figure 8-357...
  • Page 659 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Unscrew 3 screws and detach the probe light assy Figure 8-358 Unscrewing 3 screws and detaching the probe light assy Chapter 8 - Replacement Procedures 8-275...
  • Page 660 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 8-54 Replacement of ACDC 8-54-1 Manpower 1 person, 15 minutues 8-54-2 Tools Standard Phillips screwdriver 8-54-3 Removal Procedure 1) Remove 2 screws and Power Cord Fix Bracket.
  • Page 661 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3) Remove 5 screws and Cable Cover Bracket. REAR CABLE COVER BRKT Figure 8-361 Unscrewing 5 screws and removing the rear cable cover brkt 4) Disconnect Cables and unscrew 7 screws. Then remove NEST EMI Back Bracket.
  • Page 662 IRECTION EVISION ERVICE ANUAL 6) Remove 12 screws: [LOGIQ P7/P9 R1 to R3] 1.) from IOBOSX or IOBOX-B. Figure 8-364 Removing 12 screws 2.) Install Artesyn ACDC-4 (5728320-4). 3.) Connect ‘Artesyn ACDC Input Harness (5728321)’ and ‘ACDC Output Harness (5728321)’...
  • Page 663 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 4.) Assemble Artesyn ACDC Output Harness (5728321) with ACDC Assy. Figure 8-366 5.) Tighten the harness using cable tie (2160954). Figure 8-367 Tightening the harness 6.) Assemble Artesyn ACDC heatsink. Red box is the ACDC heatsink position. Remove the adhesive tape from the heatsink.
  • Page 664 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL NOTE: Attach the heatsink exactly at the same position like red square. Figure 8-369 Attaching Artesyn ACDC heatsink 8.) Check the Internal harnesses. Figure 8-370 Checking the Internal harness...
  • Page 665 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL [LOGIQ P8/P9/P10 R4 and R4.5] 1.) from IOBOX. Figure 8-371 2.) Remove the TDK-L ACDC Input Harness (5856083), TDK-L ACDC Output Harness (5856084) then uninstall ACDC Module (5831552) with removing 4 screws on IOBOX.
  • Page 666 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 3.) Prepare ACDC Module (5831552). Then remove the 2screws on ACDC Module. Figure 8-373 TDK ACDC Module 4.) Connect ‘TDK-L ACDC Input Harness (5856083)’ to ACFL and ‘TDK ACDC Module (5831552)’.
  • Page 667 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 6.) Check the Internal harnesses Figure 8-376 Checking the Internal harness 8-54-4 Assemble procedure All components to be installed/assembled in reverse order of removal. Please refer to step 1) ~ 6).
  • Page 668 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 8-54-8 Check the IOBOX (or IOBOX-B) and replaced ACDC serial number Figure 8-378 ACDC serial number 8-284 Section 8-54 - Replacement of ACDC...
  • Page 669 Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives an overview of replacement parts available for the LOGIQ P7/P8/P9/P10. LCD Monitor OPIO Nest Box Parts Figure 9-1 LOGIQ P7/P9 R1 to R2.5 Mechanical Part...
  • Page 670 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL LCD Monitor OPIO Nest Box Parts Figure 9-2 LOGIQ P7/P9 R3 Mechanical Part Plastics Covers (Front/Sides/Rear) Section 9-1 - Overview...
  • Page 671 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL LCD Monitor OPIO Nest Box Parts Figure 9-3 LOGIQ P8/P9/P10 R4 and R4.5 Mechanical Part Plastics Covers (Front/Sides/Rear) Chapter 9 - Renewal Parts...
  • Page 672 IRECTION EVISION ERVICE ANUAL Table 9-1 Contents in Chapter 9 Section Description Page Number Overview List of Abbreviations LOGIQ P7/P8/P9/P10 Software Parts List Groups 9-10 Plastics Covers (Front/Sides/Rear) 9-16 LCD Monitor 9-19 OPIO 9-29 Nest Box Parts 9-37 Mechanical Part...
  • Page 673 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 9-2 List of Abbreviations Alternating Current Analog to Digital Converter A/N Key Alpha Numeric Keys ASIC Application Specific Integrated Circuit Assy Assembly MARU Core board(Merged Radio frequency module and Beaforming board)
  • Page 674 5760487-2 5781578 5781578-2(Skylake) R2.5.5 5760487-3 5766409-2(Ivybridge) 5781578-3(Skylake) R2.5.6 5760487-4 5766409-3(Ivybridge) 5781578-4(Skylake) R2.5.7 5760487-5 5766409-4(Ivybridge) Table 9-3 LOGIQ P7/P8/P9/P10 software (R3 to R4.5) Item SW Version SW FRU Number R4.5 R3.0.2 5795390 R3.0.3 5795390-2 R3.0.4 5795390-3 R3.0.5 5795390-4 R3.0.6 5795390-5 R3.0.8 5795390-7 R3.0.9...
  • Page 675 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-3 LOGIQ P7/P8/P9/P10 software (R3 to R4.5) Item SW Version SW FRU Number R4.5 R3.0.12 5795390-11 R4.0.3 5851137 R4.0.4 5851137-2 R4.0.5 5851137-3 R4.0.7 5851137-5 R4.0.6 5851137-4 R4.5.1...
  • Page 676 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 9-4 Parts List Groups Figure 9-4 Console Views (R1 to R2.5) Figure 9-5 Console Views(R3) Section 9-4 - Parts List Groups...
  • Page 677 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 9-6 Console Views (R4 and R4.5) Table 9-4 Mechanical and user accessible parts Item Part Group Name Table Number Description 101- Table 9-5 on page 9-10...
  • Page 678 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 9-5 Plastics Covers (Front/Sides/Rear) 9-5-1 Plastics Covers (Front/Sides/Rear) (R1,R2 and R2.5) Table 9-5 Plastic Covers Part Item Part Name Number Description R2.5 R4.5 Assy Plastic 5499849...
  • Page 679 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-5 Plastic Covers Part Item Part Name Number Description R2.5 R4.5 High Cabinet Cover Top Cover - High 5499645 Cabinet Low Cabinet Cover Top Cover - Low...
  • Page 680 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-5 Plastic Covers Part Item Part Name Number Description R2.5 R4.5 OPIO Top Cover L, OPIO Top Cover R, OPIO USB COVER 5582886 OPIO Top Cover Set...
  • Page 681 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 9-5-2 Plastics Covers (Front/Sides/Rear) (R3 and R4) Table 9-6 Plastic Covers (R3 and R4) Part Item Part Name Number Description R2.5 R4.5 Assy Plastic 101-1 5794893 Side Cover R Assy R3...
  • Page 682 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-6 Plastic Covers (R3 and R4) Part Item Part Name Number Description R2.5 R4.5 Assy Plastic 105-1 5794899 High Cabinet Assy R3 Assy Plastic 106-1 5794892...
  • Page 683 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-6 Plastic Covers (R3 and R4) Part Item Part Name Number Description R2.5 R4.5 Assy Plastic 111-1 5848755 Side Cover R Assy R4 Assy Plastic 112-1...
  • Page 684 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 9-6 LCD Monitor Table 9-7 User Interface Part Item Part Name Number Description R2.5 R4.5 Full 21.5" LCD Monitor set with hinge 5501560-122 21.5" LCD Full 21.5" LCD Monitor set with hinge 21.5 INCH LCD...
  • Page 685 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-7 User Interface Part Item Part Name Number Description R2.5 R4.5 Plastics, Front, Rear and Cable cover, Monitor USB cap 21.5 INCH LCD Cover 5727721 SET w/o glass filter...
  • Page 686 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-7 User Interface Part Item Part Name Number Description R2.5 R4.5 Plastics, Front, Rear, Cable Cover and Hook Remover HD Monitor Cover Set S5862906 23.8inch Bezelless 23.8 inch LCD monitor Bezelless 23.8 inch...
  • Page 687 R2.5 R4.5 OPIO ASSY LOGIQ P7 SET OPIO ASSY LOGIQ P7 S5781418-2 Full OPIO set, including Main and Sub OPIO sets for LOGIQ P7 R3(Different Emblem) OPIO ASSY LOGIQ P7 7503848 OPIO ASSY LOGIQ P9 SET OPIO ASSY LOGIQ P9...
  • Page 688 Description R2.5 R4.5 OPIO Main PWA Assembly 5503847 OPIO Main PWA Assy Maru OPIO 10.4 LCD Set LOGIQ P7/P9 OPIO 5561038 10.4 inch LCD module LP7 Upper ASSY SET S5788926 LP7 Upper ASSY SET LP9 Upper ASSY SET LP9 Upper ASSY SET...
  • Page 689 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-8 OPIO Part Item Part Name Number Description R2.5 R4.5 LP9 Upper ASSY SET R3 LP9 Upper ASSY SET S6788927 Trackball Assy (In case it is not...
  • Page 690 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-8 OPIO Part Item Part Name Number Description R2.5 R4.5 OPIO AN Key Gel Board 5503844 OPIO AN Key Gel Board LP7P9 OPIO ANKEY GEL BOARD PWA - H...
  • Page 691 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-8 OPIO Part Item Part Name Number Description R2.5 R4.5 LP7P9 OPIO MAIN BOARD PWA LP7P9 OPIO MAIN S5781465 BOARD PWA SET LP7P9 OPIO FREEZE AND P1...
  • Page 692 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-8 OPIO Part Item Part Name Number Description R2.5 R4.5 COMMON ROTARY ENCODER SET FOR LP9_LP7 COMMON ROTARY S5718726-2 ENCODER SET FOR LP9_LP7 LP7P9 R3 COMMON ROTARY...
  • Page 693 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-8 OPIO Part Item Part Name Number Description R2.5 R4.5 OPIO HUB BOARD PWA FOR LP9_LP7 OPIO HUB BOARD 5503846-3 PWA FOR LP9_LP7 Probe Light Assy...
  • Page 694 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-8 OPIO Part Item Part Name Number Description R2.5 R4.5 OPIO Assy LP8P9P10 R4 8603848 Operator Panel 10.4 R4 LP8/P9/P10 OPIO Upper Assy LP8P9P10 R4 OPIO...
  • Page 695 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-8 OPIO Part Item Part Name Number Description R2.5 R4.5 Joystick Encoder set R4 Joystick Encoder set for S5856454 Operator Panel 10.4 R4 B Mode Rotary Encoder set R4...
  • Page 696 S5781425 lower Assy - H NOTE: In case LOGIQ P7/P9 OPIO 10.4 inch LCD module set (5561038) is not available: For LP7 SW R1.0.5 or above systems, LP7 Upper ASSY SET(S5788926) can be used instead. For LP7 SW R1.0.4 or below systems, LP7 Upper ASSY SET(S5788926), Base SW(5507300-4) and Application SW(5507301-6) can be used instead.
  • Page 697 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 9-8 Nest Box Parts Table 9-10 Nest Box Parts Item Part Name Part Number Description R2.5 R4.5 MBP Assy 5503745 MBP Assy MBP R4 PWA 401-1...
  • Page 698 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-10 Nest Box Parts Item Part Name Part Number Description R2.5 R4.5 MCB-A PWA 402-3 8494666-2 MCB-A MDC Assy 5552220-2 MDC Assy MDC-R3 Assy 403-1 5552220-3...
  • Page 699 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-10 Nest Box Parts Item Part Name Part Number Description R2.5 R4.5 MPI-R4-PWA 405-1 5822930 MPI-R4 IOBOX Assy 406-1 5554921-5 IOBOX-B Assy IOBOX-B Assy 406-2 5991830-4...
  • Page 700 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-10 Nest Box Parts Item Part Name Part Number Description R2.5 R4.5 2.5inch SSD 500GB 5839947 M.2 SSD 409-1 5832283 M.2 SSD Without Cables DVD Drive...
  • Page 701 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-10 Nest Box Parts Item Part Name Part Number Description R2.5 R4.5 Nest Bot FAN with Cable Assy MARU Nest Bot FAN 5549309 with Cable Assy MARU...
  • Page 702 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-10 Nest Box Parts Item Part Name Part Number Description R2.5 R4.5 COM Express MODULE,Type6,i3- 6100E, DDR4 8G Memory, Heatsink with FAN for MARU R4 414-4...
  • Page 703 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-10 Nest Box Parts Item Part Name Part Number Description R2.5 R4.5 Artesyn LPS363 ACDC Assy for Artesyn 5737480-5 LPS363 ACDC Assy for FRU TDK ACDC with cable R4...
  • Page 704 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-10 Nest Box Parts Item Part Name Part Number Description R2.5 R4.5 Peripheral Cable Assy LP8P9P10 Peripheral 422-1 S5857337 Cable Assy LP8P9P10 R4 LP9P7 Cable Set R3...
  • Page 705 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-10 Nest Box Parts Item Part Name Part Number Description R2.5 R4.5 HDMI male to HDMI female adaptor cable HDMI male to H44942LW HDMI female (5761422)**...
  • Page 706 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-11 Mechanical Parts Item Part Name Part Number Description R2.5 R4.5 Rear-Right Caster (1 per console) Castor Dual Lock 5746121 Front Air Filter, Bottom Air Filter...
  • Page 707 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-11 Mechanical Parts Item Part Name Part Number Description R2.5 R4.5 Full Assy for LCD Monitor 5507554 Fixed Arm set Full Assy for LCD Monitor 5794897...
  • Page 708 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-11 Mechanical Parts Item Part Name Part Number Description R2.5 R4.5 Gel holder 5508055-2 Gel Holder R2 Gel holder 5794072 Gel Holder R3 Link Arm Cable Hook...
  • Page 709 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-11 Mechanical Parts Item Part Name Part Number Description R2.5 R4.5 LOGIQ P9 Rotation module kit LOGIQ P9 ROTATION 5728965 MODULE KIT RS KNOB SET R3...
  • Page 710 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 9-10 Options Table 9-12 Options Part Item Part Name Number Description R2.5 R4.5 Multi Purpose Holder and Gel Cup Multi Purpose 5582894 Holder set 3 button switches operated by the...
  • Page 711 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Part Item Part Name Number Description R2.5 R4.5 LCD Arm Assy 704-1 5841783 LCD Arm R4 LCD Arm Cover Set LCD Arm Cover Set 704-2 S5857259 Plastics...
  • Page 712 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Part Item Part Name Number Description R2.5 R4.5 Side Tray L, R 5582897 Side Tray Assy Drawer Assy 5537774 Drawer Assy Drawer Assy 5794898 Drawer Assy R3...
  • Page 713 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Part Item Part Name Number Description R2.5 R4.5 Rear Cable Hook OPIO Side Rear 5730661 Hook R3 Card Reader BRKT, Self adhesive hook and loop Card Reader BRKT...
  • Page 714 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Part Item Part Name Number Description R2.5 R4.5 Bottom of Gel Warmer Gel Warmer Cap 5692045 LS8 R3 Multi Purpose Holder and Gel Cup Multi Purpose S5794159...
  • Page 715 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Part Item Part Name Number Description R2.5 R4.5 Battery 5450061-3 Battery Universal Video Converter S300, UVC-S300 Universal Video Converter 5750263 S300, UVC-S300 DCWD assy 5507749 DCWD assy...
  • Page 716 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Part Item Part Name Number Description R2.5 R4.5 Norav ECG Patient cable AHA Type Norav ECG Patient 5178978 cable AHA Type Norav ECG Patient Cable IEC Type...
  • Page 717 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Part Item Part Name Number Description R2.5 R4.5 ECG LEADWIRE SET, 3-LD GROUPED, GRABBER, AHA, 74 CM/ 29 IN 2106390- ECG LEADWIRE SET AHA ECG LEADWIRE SET, 3-LD...
  • Page 718 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Part Item Part Name Number Description R2.5 R4.5 AN KEYBOARD AN KEYBOARD ASSY ENGLISH 5582898 ASSY AN KEYBOARD ASSY GREEK AN KEYBOARD ASSY AN KEYBOARD 5611226 NORWEGIAN...
  • Page 719 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 9-11 Optional Peripherals and Accessories Table 9-13 Optional Peripherals and Accessories - Printers Item Part Name Part Number Description Digital B/W Video Printer, USB-Port Digital B/W Video Printer...
  • Page 720 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 9-12 Probes Refer to the Table 3-24 Probe Description 9-52 Section 9-12 - Probes...
  • Page 721 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 9-13 Power Cord Table 9-14 Power Cord Part Item Name Part Number Description R2.5 R4.5 Power Cord for India Power Cord 6736108-3 for India Power Cord for Switzerland...
  • Page 722 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-14 Power Cord Part Item Name Part Number Description R2.5 R4.5 Power Cord for Italy Power Cord 6736110-3 for Italy Power Cord for Argentina Power Cord...
  • Page 723 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-14 Power Cord Part Item Name Part Number Description R2.5 R4.5 Power Cord for China Power Cord 6736104-3 for China Power Cord for Brazil Power Cord...
  • Page 724 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 9-14 Power Cord Part Item Name Part Number Description R2.5 R4.5 LP7P9 POWER CORD FIX BRKT BACK FRONT LP7P9 CORD FIX 5802024 BRKT Power Cord Fix Brkt R4...
  • Page 725 10-1-1 Preventive Maintenance Inspections It has been determined by engineering that your LOGIQ P7/P8/P9/P10 system does not have any high wear components that fail with use, therefore no Preventive Maintenance Inspections are mandatory. However, some Customers Quality Assurance Programs may require additional tasks and/or inspections at a different frequency than listed in this manual.
  • Page 726 The service procedures and recommended intervals shown in the Customer Care Schedule assumes that you use your LOGIQ P7/P8/P9/P10 for an average patient load (10-12 per day) and not use it as a primary mobile unit which is transported between diagnostic facilities.
  • Page 727 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 10-3-1 How often should care & maintenance tasks be performed? (cont?) Table 10-2 Customer Care Schedule Item Service at Indicated Time Notes • Recommend to clean filter at Air Filter Grid Remove the filter grid and clean the air filter.
  • Page 728 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 10-2 Customer Care Schedule Item Service at Indicated Time Notes Also after corrective • Surface Probe Leakage maintenance or as required by Current Checks your facilities QA program.
  • Page 729 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 10-4 Tools Required 10-4-1 Special Tools, Supplies and Equipment 10-4-1-1 Specific Requirements for Care & Maintenance Table 10-3 Overview of Requirements for Care & Maintenance Tool...
  • Page 730 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 10-5 System Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary. Table 10-4 System Checks...
  • Page 731 24, 2022) IRECTION EVISION ERVICE ANUAL 10-5-2 Functional Checks The functional checks take about 60 minutes to perform. Refer to the LOGIQ P7/P8/P9/P10 Basic User Manual whenever necessary. 10-5-2-1 System Checks Table 10-5 System Functional Checks Step Item Description Verify basic B Mode (2D) operation.
  • Page 732 Inlet Connector Inlet connector retainer is functional. 10-5-4 Cleaning 10-5-4-1 General Cleaning Frequent and diligent cleaning of the LOGIQ P7/P8/P9/P10 ultrasound system reduces the risk of spreading infection from person to person, and also helps to maintain a clean work environment. Table 10-8 General Cleaning Step...
  • Page 733 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 10-5-4-2 Air filter cleaning Clean the system's air filters to ensure that a clogged filter does not cause the system to overheat and reduce system performance and reliability. It is recommended the filters be cleaned every two weeks, but the requirements will vary with environment.
  • Page 734 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Pull out the Bottom air filter. Figure 10-2 Bottom Air filter location 3) Dust the filter with a vacuum cleaner and/or wash it with a mild soapy solution. If washed, rinse and dry the filter before re-installation.
  • Page 735 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 10-5-5 Physical Inspection Table 10-9 Physical Checks Step Item Description Labeling Verify that all system labeling is present and in readable condition. Scratches & Dents Inspect the console for dents, scratches or cracks.
  • Page 736 Failure to follow the prescribed cleaning or disinfection procedures will void the probe’s CAUTION warranty. DO NOT soak or wipe the lens with any product not listed in the LOGIQ P7/P8/P9/P10 Basic User Manual and/or care card. Doing so could result in irreparable damage to the probe !! CAUTION: and/or LOGIQ P7/P8/P9/P10 system.
  • Page 737 After you press a software download button, it may take up to three (3) seconds for the system to respond. NOTE: Software Updates through the GE service platform may not be available in all countries. Chapter 10 - Care & Maintenance 10-13...
  • Page 738 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 10-6-1 Software download and installation You receive notification of available software updates via the envelope icon that appears in the status bar on the bottom of the monitor.
  • Page 739 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL NOTE: If you are not logged in as an Administrator and there is a software package to download, you will see the following dialog when shutting down the system: Figure 10-5 Software Download Dialog If Not an Administrator If you get a dialog with “Unknown status,”...
  • Page 740 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 2) Press Download. Figure 10-7 System - Exit Dialog Window The download process starts. The progression of the download process is displayed. NOTE: The download step can be paused. While paused, you can return to normal operation. However, once the software installation has begun, the system is not useable until the software installation is done.
  • Page 741 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Press Download. Figure 10-8 Download in Progress Options • Stop: Stops the download process; no software upgrade is performed. If you select Stop, you WILL NOT be given an opportunity to download this package again.
  • Page 742 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Figure 10-10 Paused Software Download • Resume: Press Resume to continue the software download. 3) If auto-install was not checked, you will see the following dialog when the software download has finished (the message, “SW is ready to be installed,”...
  • Page 743 New Software Verification checklist. !! CAUTION: These verification results are tracked for regulatory purposes, sent back to GE for tracking, and approved with your signature. Chapter 10 - Care & Maintenance...
  • Page 744 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL NOTE: Press the Question mark to get information on how to check each feature. Figure 10-13 New Software Verification Checklist NOTE: You can move this dialog, but DO NOT close it. You can also minimize this dialog by clicking on the “<<“...
  • Page 745 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL When there is a failed feature noted on the New Software Verification Checklist, the following Warning dialog displays: Figure 10-15 Re-Install Previous Software Version Dialog When the New Software Verification Checklist contains a ‘Failed’ feature, you have two options available: 1) Re-Install the previous version of software (“OK”), or 2) Revert to the Checklist (“Cancel”)
  • Page 746 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 10-7 Using a Phantom Refer to the User Manual of the Phantom for information on using a phantom and quality assurance tests. For measurement accuracy of the system review chapter 13.5 of the Basic User Manual of LOGIQ P7/P8/P9/P10.
  • Page 747 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 10-8-2 GEHC Leakage Current Limits The following limits are summarized for ANSI/AAMI ES60601-1(For USA), IEC 60601-1 Medical Equipment Safety Standards and IEC 62353 Medical Electrical Equipment - Recurrent test and test after repair of medical electrical equipment.
  • Page 748 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 10-14 Type BF Patient Auxiliary Limits Country Conditions Others Close Protective earth and close neutral with normal polarity 0.1mA* 0.1mA*** Close Protective earth and close neutral with reverse polarity 0.1mA*...
  • Page 749 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 10-8-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
  • Page 750 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 10-8-4 Grounding Continuity Electric Shock Hazard! CAUTION The patient or operator MUST NOT come into contact with the equipment during this test !! CAUTION: Measure the resistance from the third pin of the attachment plug to the exposed protectively - earthed metal parts of the case.
  • Page 751 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 10-8-5 Enclosure Leakage Current Test 10-8-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible non- protectively-earthed parts of equipment if the ground wire were to break. The test verifies the isolation of the power line from the chassis.
  • Page 752 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL When using Dale 600/601 or a similar test instrument, its power plug may be inserted into the wall outlet and the equipment under test is plugged into the receptacle on the panel of the meter. This places the meter in the grounding conductor and the current flowing from the case to ground will be indicated in any of the current ranges.
  • Page 753 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 10-8-6-2 Generic Procedure The testing meter is connected from accessible protectively-earthed parts of the equipment to ground. Measurements should be made with power line polarity normal and reversed, and with the neutral open and closed.
  • Page 754 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 10-8-7 Type BF Patient Leakage Current Test - Probe 10-8-7-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
  • Page 755 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Table 10-18 Typical Data Sheet for Type BF Patient Leakage Current Test Tester Neutral Tester Polarity Switch Ground Switch Measured leakage current Open Reversed Closed Chapter 10 - Care & Maintenance...
  • Page 756 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 10-8-8 Type BF Patient Leakage current - Mains to Probe Reference the procedure in the IEC 60601-1. Measure leakage current flow from mains to Probe. Electric Shock Hazard.
  • Page 757 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 10-9 When There’s Too Much Leakage Current... 10-9-1 Earth and/or Enclosure Fails Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent.
  • Page 758 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 10-9-5 New Unit If the leakage current measurement tests fail on a new unit and if situation cannot be corrected, submit a Safety Failure Report to document the system problem. Remove unit from operation.
  • Page 759 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Type BF Patient Leakage Current Test - Probes (This test should be done for all probe) Table 10-23 Tester Tester Polarity Ground Max Value Value Comments Neutral...
  • Page 760 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL Section 10-10 Ultrasound INSPECTION CERTIFICATE Customer Name: System ID: Dispatch Number/Date Performed: Warranty/Contract/HBS System Type Model Number: Serial Number: Manufacture Date: Probe 1: Frequency: Scan Format*: Model Number:...
  • Page 761 GEDRAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground & wiring conflg.) System Ground Continuity Chassis Source Leakage Current - Probe Chassis Source Leakage Current - Wheel...
  • Page 762 GERAFT LOGIQ P7/P8/P9/P10 5604324, R DRAFT (M 24, 2022) IRECTION EVISION ERVICE ANUAL 10-38 Section 10-10 - Ultrasound INSPECTION CERTIFICATE...
  • Page 763 This page was intentionally left blank.
  • Page 764 2022 General Electric Company. © GE Ultrasound Korea, Ltd. 9, Sunhwan-ro 214beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea www.gehealthcare.com...

This manual is also suitable for:

Logiq p8Logiq p9Logiq p10

Table of Contents

Save PDF