V-ACTOR HF HANDPIECE OPERATIONS, SET-UP, CLEANING, AND MAINTENANCE

REPROCESSING THE HANDPIECE AND

VIBRATION TRANSMITTERS
After each use, the parts of the handpiece that come into
contact with the patient must be cleaned and disinfected
thoroughly before they are used again.
The instructions must be strictly observed in order to
prevent damage to the parts and malfunctions.
Make sure that you have the following agents and
equipment available to perform the cleaning and
disinfection work:
• Clean, soft and lint-free cleaning cloths
• Cleaning agent
• Alcohol-based surface disinfectant
• Ultrasonic bath (if desired)

V25 AND V10

• Unscrew the vibration transmitter from the handpiece.
• Remove the vibration transmitter insert from the front cap.
– The spring element (1) on the vibration transmitter insert
does not need to be removed.
• Clean all of the parts under running water.
• Alternatively, you can clean and disinfect the vibration
transmitter insert and the sealing ring in an ultrasonic bath.
INTELECT
RPW 2 USER MANUAL
®

V40

• Unscrew the vibration transmitter from the handpiece.
• Disassemble the threaded two-part shock transmitter screw
• Remove the vibration transmitter insert from the front cap.
• Clean all of the parts under running water.
• Alternatively, you can clean and disinfect the vibration
1
cap.
– Remove the front sealing ring (2).
– This is cut open so that it can be removed more readily.
– The spring element on the vibration transmitter insert (1)
does not need to be removed.
transmitter insert and the sealing ring in an ultrasonic bath.
1
E N
2