NI 449x Calibration Procedure Français Deutsch ni.com/manuals This document contains information about calibrating the National Instruments 449x devices. For more information about calibration, visit ni.com/calibration Contents Software Requirements......................1 Documentation Requirements ....................2 Calibration Interval........................2 Test Equipment......................... 3 Test Conditions......................... 4 Calibration Procedures ......................
Calibration Interval National Instruments recommends an external calibration interval of one year for the NI 449x. Additionally, self-calibration can be performed as desired, but is recommended when the operating temperature exceeds 5 °C from the operating temperature of the previous calibration.
Test Equipment National Instruments recommends that you use the equipment in Table 1 for calibrating the NI 449x. Table 1. Recommended Equipment Recommended Parameter Equipment Model Measured Minimum Requirements Calibrator Fluke 5700A AI AC Frequency Range: coupled 20 Hz to 92.2 kHz...
Analog Input Performance Verification This section verifies the analog input (AI) performance of all NI 449x devices. Refer to the NI 449x Specifications for the number of channels and performance specifications. AI Offset Verification Complete the following steps to verify the AI offset: Connect a BNC shorting cap to channel AI 0 on the device and wait 5 seconds for the inputs to settle.
If verifying an NI 4492/4497/4499, separately verify AC and DC coupling for each supported gain setting including 0 dB at sample rates of 102.4 kS/s and 204.8 kS/s. Refer to Table 6 for the gain settings supported by supported by NI 449x devices. Repeat steps 1 through 8 for all input channels on the device.
Table 10. NI 4492/4496/4497/4498/4499 Supported Gain Settings Device Gain Settings NI 4492 NI 4496 0 dB, 20 dB NI 4497 NI 4498 0 dB, 10 dB, 20 dB, 30 dB NI 4499 If verifying an NI 4492/4497/4499, repeat steps 2 through 8 for each gain setting that your device supports.
Table 13. Maximum AI Flatness Specifications (Relative to 1 kHz) Frequency Band Flatness (dB) 20 Hz to 20 kHz ±0.005 20 kHz to 45 kHz ±0.016 45 kHz to 92.2 kHz ±0.08 Table 14. AI Flatness Limits Test Limit Calibrator Output Frequency (Hz) 0.99942 * V 1.00058 * V...
If verifying an NI 4495, repeat steps 2 through 7 for each gain setting that your device supports. Refer to Table 17 for the gain settings supported by NI 449x devices. In step 2, change the output of the voltage source based on the gain setting that is being verified. Table 15 lists the DC voltage to use for each gain setting.
Page 17
Initialize the AI calibration using the DAQmx Initialize External Calibration VI, as shown below. Note Refer to the NI-DAQmx function parameters for the LabVIEW input values. LabVIEW Block Diagram NI-DAQmx Function Call Call with the DAQmxInitExtCal following parameters: deviceName: Dev_name password: calHandle: &myCalHandle...
Page 19
Configure the calibrator to output a tone with the amplitude and frequency listed in the 0 dB gain row of Table 19. Table 19. Calibrator Output for AI AC Coupled Calibration Adjustment on NI 4492/4497/4499 Calibrator Output Device Gain Amplitude (V Frequency (Hz) 0 dB 6.36...
10. Finish the AI calibration with the DAQmx Close External Calibration VI, as shown below. Use the action if there has been any error during the AI calibration or if you do not cancel want to save the new AI calibration constants in the device EEPROM. Use the action if you want to save the new AI calibration constants in the device EEPROM.
Need help?
Do you have a question about the NI 449 Series and is the answer not in the manual?
Questions and answers