8.3
Application Measurement
The system support following measurement types:
Abdomen measurements - Used for measurements of abdominal organs (liver, gall
bladder, pancreas and kidney, etc.) and large abdominal vessels.
OB measurements- Used for measurements of fetal growth indexed (including EFW) as
well as GA and EDD calculations. The fetus can be evaluated through growth graph
analysis and fetal biophysical profile.
Cardiac measurements- Used for left ventricle function measurements and
measurements of main artery and vein parameters, etc.
Gynecology measurements - Used for the uterus, ovary and follicles, etc.
Small Part measurements – Used for small parts such as thyroid.
Urology measurements - Used for prostate, seminal vesicle, renal, adrenal, micturated
and testicle volume.
Vascular measurements – Used for carotid, cerebral, upper and lower extremities vessels,
etc.
Pediatric measurements - Used for hip joint measurement.
Nerve measurements – Used for nerve structure measurements.
Emergency measurements - Used for emergency abdomen, obstetric, superficial etc.
8.4
Measurement Accuracy
Table 1 Error of 2D Images
Parameter
Distance
Full screen
Area (Trace)
Full screen
Area (ellipse,
Full screen
circle)
Circumference
/
Angle
Full screen
Distance (iScape)
Full screen
Table 2 Volume Measurements
Value
Parameter
Range
Volume
Full screen
8-4 Measurement
Value Range
Within ±3%; or when the measured value is less
than 40 mm, the error is less than 1.5 mm.
Within ±7%; or when the measured value is less
than 16 cm
Within ±7%; or when the measured value is less
than 16 cm
Within ±10%
Within ±3%.
Linear probe: within ±5%
Other probe: within ±10%
Within ±10 %; or when the measured value is less than 64 cm
the error is less than 6.4 cm
Error
2
, the error is less than 1.2 cm
2
, the error is less than 1.2 cm
Error
3
.
2
.
2
.
3
,