Cerna Mini Microscopes
4.3.2
Fluorescence Filter Installation
The SFMGFP2 ships with an MDF-GFP filter.
Note: The SFM module does not come with a filter cube nor a filter set. Use filter
cube item # MDFM-MF2 to install your filter set in the Filter Cube Housing.
To install the MDF-GFP filter:
M3 x 10 mm
Screw
1.
Use a 1.5 mm hex key to remove the screws (2X) from either side of the
Filter Cube Housing cover.
2.
Use a 2.5 mm hex key to loosen the screw at the center of the cover to
slide out the filter cube.
3.
Install your filter set inside the filter cube. See the MDFM-MF2 manual for
detailed instructions- https://www.thorlabs.com/manuals.cfm
4.
Slide the filter cube on the Filter Cube Housing cover and tighten the
screw at the center of the cover
Page 12
2X M2 x 5 mm
Screws
Figure 7
Replacing the Filter Set
.
Chapter 4: Getting Started