Chapter 1 Safety 1.1 Safety Information For the continuing safety of the operators of this equipment, and the protection of the equipment itself, the operator should take note of the Warnings, Cautions and Notes throughout this handbook and, where visible, on the product itself. The following safety symbols may be used throughout the ha ndbook and on the equipment itself.
The MLS203 stages should be driven by the BBD series of brushless DC motor controllers. These controllers include the Thorlabs apt software suite, which provides an easy-to-use graphical interface for stand-alone operation as well as programming support for integrated applications.
Page 5
MLS203 Series XY Scanning Stage Fig. 2.1 MLS203 stage with MLS203P1 well plate adapter and 96 hole well plate, mounted on a microscope...
Chapter 3 Installation 3.1 Unpacking Note Retain the packing in which the unit was shipped, for use in future transportation. Caution Once removed from its packaging, the stage can be easily damaged by mishandling. The unit should only be handled by its base, not by any attachments to the moving platform.‘...
MLS203 Series XY Scanning Stage 3.2.2 Fitting the Finger Guards The stages are supplied with finger guards to avoid trapping of body parts or cables in the mechanism. These should be fitted before the stage is mounted on the microscope as follows: 1) Offer up the finger guards to the stage as shown below, and secure using the M3 x 16 bolts supplied.
Chapter 3 3.3 Electrical Connections The stage must be driven by a Thorlabs BBD series controller. Connect the motor leads to the MOTOR DRIVE connectors, and the encoder feedback leads to the FEEDBACK connectors. Ensure that the motor drive an d feedback leads for each motor are connected to the correct channel.
MLS203 Series XY Scanning Stage 3.4 Dimensions 3.4.1 MLS203-1 Dimensions Y AXIS MOTOR AND 260.0mm ENCODER CABLES X AXIS MOTOR AND All dimensions in mm [in.] 10.24in ENCODER CABLES 191.0mm 31.1mm 7.52in APPROX 1.22in 170.5mm 112mm REMOVABLE GUARD 6.71in 10.0mm 160.3mm...
Chapter 4 Operation 4.1 General Caution The MLS203 stages is designed to be driven by the Thorlabs BBD20x or RBD201 Brushless DC Motor Controllers. The stages are connected to the controller via 2 flying leads terminated in D-type connectors (FEEDBACK) a nd 2 leads terminated in round 8-Pin DIN Connectors (MOTOR DRIVE).
Page 11
MLS203 Series XY Scanning Stage Note The stage design incorporates several reference markers on the encoder scale, which are used to calibrate the position counter. The need for homing comes from the fact that on power up the motor (stage) is at a random position, so the value of the position counter is meaningless.
4.2 Maintenance The product is maintenance free. If any problems occur, the user should contact the local Thorlabs tech support for more information. 4.3 Transportation Caution When packing the unit for shipping, use the original packing. If this is not available, use a strong box and surround the unit with at least 100 mm of shock absorbent material.
Chapter 5 Specification Parameter Value Travel 110 mm x 75 mm (4.3" x 2.95") Max Speed 250 mm/s* Acceleration 2000 mm/s* Bidirectional Repeatability 0.25 µm Unidirectional Repeatability 0.25 µm Backlash N/A (No Leadscrew) Min Incremental Movement 0.1 µm Home Location Accuracy 0.25 µm Max Load 2.2 lbs (1 kg)
A range of attachment brackets is also available, which allow the stage to be fitted to specific microscopes from Nikon and Olympus - see ww w.thorlabs.com for more details.
MLS203 Series XY Scanning Stage Chapter 7 Regulatory 7.1 Declarations Of Conformity 7.1.1 For Customers in Europe See Section 7.3. 7.1.2 For Customers In The USA This equipment has be en tested an d found to comply with the limits fo r a Class A digital device, persua nt to part 15 of the FCC rules.
• left over parts of units disassembled by the user (PCB's, housings etc.). If you wish to return a unit for waste rec o very, please contact Thorlabs or your nearest dealer for further information. 7.2.2 Waste treatment on your own responsibility If you do not return an "end of life"...