Do you have a question about the 9100c and is the answer not in the manual?
Questions and answers
Samwel
April 3, 2025
The oxygen is internally leaking from the pressure switch valve. What's the possible solution towards this?
1 comments:
Tom
April 3, 2025
@Samwel Try trimming the plastic O2 tubing to make sure it is flat, smooth and seats firmly into the quick connect ports. Other than that you may need to replace the manifold, which I believe is part # M1195025-S.
Page 1
GE Healthcare 9100c Anesthesia Machine Technical Reference Manual Software Revision 1.x Approved 2010−12−9 06:37 EET Junmin Ji (Rick) M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 2
M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Manglende overholdelse af denne advarsel kan medføre skade på grund af elektrisk stød, mekanisk eller anden fare for teknikeren, operatøren eller patienten. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 5
• Käesoleva hoiatuse eiramine võib põhjustada teenuseosutaja, operaatori või patsiendi vigastamist elektrilöögi, mehaanilise või muu ohu tagajärjel. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 6
• Brot á sinna þessari aðvörun getur leitt til meiðsla á þjónustuveitanda, stjórnanda eða sjúklings frá raflosti, vélrænu eða öðrum áhættum. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 7
Š br din juma neiev rošanas rezult t var rasties elektrisk s str vas trieciena, meh nisku vai citu faktoru izrais tu traumu risks apkopes sniedz jam, operatoram vai pacientam. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 8
• A não observância deste aviso pode ocasionar ferimentos no técnico, operador ou paciente decorrentes de choques elétricos, mecânicos ou outros. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 9
• Zanedbanie tohto upozornenia môže spôsobi zranenie poskytovate a služieb, obsluhujúcej osoby alebo pacienta elektrickým prúdom, mechanické alebo iné ohrozenie. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 10
Bu uyarõya uyulmamasõ, elektrik, mekanik veya di er tehlikelerden dolayõ teknisyen, operatör veya hastanõn yaralanmasõna yol açabilir. (ZH-CN) (ZH-HK) (ZH-TW) • viii M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Important The information contained in this Technical Reference Manual pertains only to those models of products which are marketed by GE Healthcare as of the effective date of this manual or the latest revision thereof. This Technical Reference manual was prepared for exclusive use by GE Healthcare service personnel in light of their training and experience as well as the availability to them of parts, proper tools and test equipment.
Page 12
9100c M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
1.5 System components ......1-6 1.6 Symbols used in the manual or on the equipment ..1-8 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 14
2.4 General Description ......2-14 2.5 9100c Ventilator features ..... . . 2-15 2.5.1 Safety features .
Page 15
4.5.3 O2 Cell ....... . 4-16 M1207026 xiii M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 17
9.7 Replace system switch assembly ....9-13 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 18
10.1.2 Test Tools ......10-4 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 19
10.20.2 Adjustable AGSS ..... 10-38 M1207026 xvii M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 20
10.20.3 Active AGSS ......10-40 10.21 CGO Valve, AGSS Installation, AGSS Flowtube . . 10-42 11 Schematics and Diagrams xviii M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
1 Introduction In this section This section provides a general overview of the 9100c line of anesthesia machines. 1.1 What this manual includes ..... . . 1-2 1.2 Standard service procedures .
9100c 1.1 What this manual includes Anesthesia Machine This manual covers the service information for the 9100c anesthesia machine and ventilator. It details the following components: • Ventilator components • Gas delivery components • Breathing system components • Frame components...
1.2 Standard service procedures 1.2.1 User’s Some sections of this manual refer you to the User’s Reference Manual for the 9100c anesthesia machine. To expedite repairs, you Reference Manual must have and be familiar with the User’s Reference Manual. Refer to the User’s Reference Manual for further information about the operation of the system.
Mode, and other optional features. Note Configurations available for this product depend on local market and standards requirements. Illustrations in this manual may not represent all configurations of the product. The 9100c product is not suitable for use in an MRI environment. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English...
• EZchange Canister (CO Bypass) • Gas scavenging (active, adjustable or passive) • Localized isolated electrical power outlets • Mounting arm M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
9100c 1.6 Symbols used in the manual or on the equipment Symbols replace words on the equipment, on the display, or in GE Healthcare manuals. No one device or manual uses all the symbols. Warnings and Cautions tell you about dangerous conditions that can occur if you do not follow all instructions in this manual: •...
Page 29
Read to top of float EZchange Canister (CO bypass) Caution: federal law prohibits dispensing without prescription Alarm silence touch key Menu touch key M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 30
This way up Positive pressure of flow sensor Pneumatic Output Negative pressure of flow sensor 1-10 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
The flow of gas through the breathing system. • Electrical signals between the anesthesia machine, including the breathing system, and the ventilator. • Ventilator mechanical subsystems. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
ON, to the common gas outlet, and into the breathing system. A pressure relief valve limits the maximum outlet pressure. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 35
53. Flow indicator 26. Free breathing check valve 54. To disposal system 27. Atmosphere 55. Active AGSS interface 28. Pop-off valve M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Decreasing the N O flow from the engagement point rotates the O sprocket away on the O knob, increasing the O concentration. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Scavenged gas paths: APL or pop-off valve. • Flow through the optional EZchange Canister: EZchange ON and EZchange OFF (CO bypass). M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
During inspiration, common gas flows from the machine into the inspiratory limb, upstream of the inspiratory check valve. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 40
As circuit pressure approaches the pressure placed on the APL disc, any additional pressure will cause the disc to lift and gas vents to the scavenging system. 2-10 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
(inspiratory check valve) to the patient. During inspiration, common gas flows into the inspiratory limb, upstream of the inspiratory check valve. M1207026 2-11 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 42
PEEP level. During exhalation, common gas flows backwards through the absorber into the expiratory limb, downstream of the expiratory check valve. 2-12 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 43
Excess gas vents to the scavenging system through the pop-off valve and the exhalation valve. M1207026 2-13 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
9100c 2.6 9100c Ventilator components Major components of the 9100c ventilator are found in different locations of the anesthesia machine. The ventilator package consists of: 1. Display Module – The Display Module includes: • Display Monitor Board • LCD Display •...
9100c 2.6.2 Control Sample The Control Sample Board (CSB) is located underneath the tabletop. Board 2-18 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
PEEP safety valve • Supply pressure sense switch • PEEP control valve • 200 mL reservoir • Calibrated bleed orifice M1207026 2-19 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
2 Theory of Operation 2.7.3 Sealed Lead A sealed lead acid battery supplies battery backup for the 9100c Acid Battery ventilator. The battery provides: • capacity to operate ventilator system for 30 minutes (fully charged); • the battery is internally fused (auto-resettable).
Flow valve control • PEEP valve drive and PEEP safety valve drive Figure 2-7 • Control Sample Board block diagram 2-22 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Pilot control of the exhalation valve is done with PEEP Control Valve (7), Supply Pressure Switch (9), and the PEEP Safety Valve (8). M1207026 2-25 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
If the pressure in the patient circuit exceeds 4 cm O, the pop-off valve opens to exhaust excess fresh gas flow at a rate up to 4 L/min. 2-26 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
The tube closest to the patient connection (13) must be connected to the port labeled “ P+”. Large Small Port Port M1207026 2-27 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 58
9100c 2-28 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
3.13 Electrical safety tests ......3-24 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 60
9100c WARNING After any repair or service of the 9100c system, complete all tests in this section. Before you do the tests in this section: • Complete all necessary calibrations and subassembly tests. Refer to the individual procedures for a list of necessary calibrations.
• Flow Sensor Negative • Flow Sensor Positive Then, complete the checkout procedure for the entire machine in the following sections. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
• The power cord is connected to a wall outlet. The mains indicator comes on when AC Power is connected. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
6. Connect the pipeline supplies one at a time and ensure that the corresponding gauge indicates pipeline pressure. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
If you overshoot a setting, turn the O flow control clockwise until the N O flow decreases to the previous setting before continuing the test. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
2. Connect the pipeline supplies or slowly open the cylinder valves. 3. Turn all flow controls full clockwise (zero flow). 4. If equipped, set the ACGO selector switch to Breathing System. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 66
Set the N O flow (using the The O flow must be greater flow control) to (L/min) than (L/min) M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Note: Hold the test adapter to the common gas outlet to avoid loosening. 8. Remove the test device and the adapter. 9. Replace the breathing system. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
5. Turn all of the flow controls fully clockwise (zero flow). 6. Reconnect the O pipeline supply. 3-10 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Large leak (if long filling time). Failure • The output of the pressure regulator is incorrect. • The pressure regulator is incorrectly connected. M1207026 3-11 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
8. Complete steps 5 through 7 for each vaporizer and vaporizer position. 9. Set the system switch to Standby. 10. Turn all of the flow controls fully clockwise (closed). 3-12 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
(1 L/min for one minute). Turn off all vaporizers at the end of the low-pressure leak test. 6. Flush the system with O M1207026 3-13 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 72
Set the vaporizer to Off. • Test the remaining vaporizers. 5. Disconnect the test device. 6. Replace the breathing system. 3-14 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
4. Keep the flowmeter of the test device vertical for accurate results. 5. Fully open the needle valve on the test device (counterclockwise). M1207026 3-15 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 74
3. Connect the test adapter port 1 to the common gas outlet. 4. Connect the test device to the test adapter port 2. 3-16 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 75
Agent mixtures from the low-pressure leak test stay in the WARNING system. Always flush the system with O after the low- pressure leak test (1 L/min for one minute). M1207026 3-17 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 76
1 L/min. • Continue the O flow for one minute. • Turn the O flow control fully clockwise (closed). 3-18 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
11. Test the apnea and low airway pressure alarms: • Turn all flow controls fully clockwise. • Set the Bag/Vent switch to Bag. M1207026 3-19 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 78
100%. • Turn the O flow control fully clockwise (zero flow). 14. Set the system switch to Standby. 3-20 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
O flush button to approximately 30 cm H • Release the O flush button. The pressure must not M1207026 3-21 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 80
10. Set the system switch to Standby. Make sure that there are no test plugs or other objects WARNING caught in the breathing system. 3-22 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
5. Make sure the following message is on the ventilator display: • “On Battery - Power OK?” 6. Connect the power cord again. 7. Make sure the alarm cancels. M1207026 3-23 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
2. Make sure that the resistance to ground is less than 0.2 between an exposed metal surface and the ground pin on the power cord. 3-24 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
4.5 Calibration ........4-13 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 84
4.6.4 Altitude Setting ......4-29 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
CSB and alarm MCU communication If one or more of these tests fail, the display provides a readout of the problem. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
The following sections in this manual are sequenced in the order that they appear on the screen. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
The Software Version includes: • CSB Version • DMB Version • Alarm Version The “System Time” displays the time set in the ventilator screen. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
4.4 Diagnostic Test/Tools The Diagnostic Tests/Tools menu includes a selection of items that look at individual subsystems of the 9100c Ventilator. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
— On or Off Electronic switches • Sensor Status — Connected or Disconnected • Breathing Circuit — Connected or Disconnected M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Battery Status “Battery Charging” — Battery is charging. “Battery is ON - Power is OK??” — System is running on battery. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
If any of the select buttons test fails, replace the front panel keyboard assembly. If the encoder knob test fails, replace the rotary encoder assembly. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
When the result is “Fail”, the icon background should be red. Remarks If one or more of these tests fail, please check the CSB, DMB, Vent Engine and the harnesses. 4-10 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
When setting the Inspiratory Valve or the PEEP Valve: • Turn the encoder clockwise to increase the values. • Turn the encoder counterclockwise to decrease the values. M1207026 4-11 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
The leak rate is the fresh gas flow needed to maintain 30 cm • The system should have a leak rate < 200 mL/min. 4-12 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
These procedures require you to disassemble and reassemble parts of the breathing system. Accordingly, the procedures are arranged to minimize the disassembly and reassembly process. M1207026 4-13 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Remarks Ensure that the breathing system is removed. “Fail” indicates a problem in the CSB (Airway pressure transducer) 4-14 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
9. Set the fresh gas flow to zero upon the completion of calibration Remarks “Fail” indicates a problem in the CSB (Airway pressure transducer). M1207026 4-15 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Place the sensor that passed the 21% test in the breathing system and expose it to 100% O “Fail” incidents a problem in the O Cell. Replace if necessary. 4-16 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
8. Connect the pediatric bag(1L) (F) to the calibration pipe using elbow fitting (G). Ensure that the test tubing is leak free. M1207026 4-17 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 100
The data is stored in EEPROM and is used during normal operation to compensate for the individual valve’s output characteristics. 4-18 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
“ P+” and “ P-”. If the calibration procedure fails, it indicates a problem in the flow measurement differential pressure transducers of the CSB. M1207026 4-19 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
6. Connect the tubing from calibration pipe to the calibration fitting. Ensure that the test tubing is leak free and the flow sensor tubes are connected. 4-20 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 103
The data is stored in EEPROM and is used during normal operation to compensate for the individual valve’s output characteristics. M1207026 4-21 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
5. Connect the calibration fitting (E) to the inspiratory flow port. 6. Connect the tubing from calibration pipe to the calibration fitting. Ensure that the test tubing is leak free. 4-22 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 105
The data is stored in EEPROM and is used during normal operation to compensate for the individual flow sensor output characteristics. M1207026 4-23 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
The data is stored in EEPROM and is used during normal operation to compensate for the individual flow sensor output characteristics. 4-24 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
The data is stored in EEPROM and is used during the calibration to compensate for the individual calibration orifice characteristics. M1207026 4-25 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
If the ventilator “Key Code” is changed, the new code should be recorded on the display module side panel label. 4-26 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Remarks Ensure the AC power is ON during the upgrade. M1207026 4-27 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Korean • French • Spanish • Portuguese The language setting is stored in EEPROM with the default setting as English. 4-28 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Altitude settings range from -400 to 3600 meters in increments of 100 meters. Remarks The setting is saved in EEPROM; the default value is 0 meter. M1207026 4-29 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 112
9100c 4-30 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
5.4.1 Zero the pressure gauge ....5-8 5.4.2 Checking the pressure gauge accuracy ..5-9 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
5. Remove the plug from the test port and connect a test device capable of measuring 689 kPa (100 psi). Remove Plug M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
7 kPa (1 psi). This adjusts the “low flow” regulated output to the high side of the specification so that the M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 116
7 kPa (1 psi). This adjusts the “low flow” regulated output to M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
11. Replace the large rear upper panel. 12. Perform the checkout procedure (Section 3). 5.1.3 Adjusting Important: Cylinder supplies in a 9100c machine must have all high- Cylinder Regulators pressure regulators set to the same pressure range. If a regulator is replaced, the replacement regulator must be set (as required) to the same specification as the one removed.
8. Disconnect the pressure meter and reconnect the tubings. 9. After all regulators have been tested, perform the Post-service checkout (Section 3). M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
The hypoxic guard is an internal system that is not serviceable. If the hypoxic guard is out of calibration or is malfunctioning, replace the entire flow control assembly. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
9. If required, repeat zero and span procedure (Section 5.4.1 and 5.4.2). 10. Replace the lens cover. Zero Adjust M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
0±1 cm H 40 cm H 40±2 cm H -5 cm H -5±2 cm H Pressure Gauge Fitting Test Device Low-pressure Source (Syringe) M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 122
9100c 5-10 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Pipeline and cylinder tests (Section 3.3). c. Flow control test (Section 3.4). d. Pressure relief tests (Section 3.5). e. O supply alarm test (Section 3.6). M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 125
Low-pressure leak test (Section 3.9). Alarm tests (Section 3.10). Breathing system tests (Section 3.11). k. Power failure test (Section 3.12). Electrical safety tests (Section 3.13). M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
1. Replace the free breathing flapper valve. 2. Replace the free breathing valve o-ring. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
6. Scissor the tail, ensure the tail is not longer than 3 mm. 7. Reassemble the system. 8. Perform the Preoperative Checkout Procedure. (refer to the 9100c User’s Reference manual). M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
You can verify this pressure by attaching a pressure test device to the regulator pressure port (shown below) and adjusting the regulator to 172 ±1.72 kPa (25 ±0.25 psi). M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
5. Remove the test adapter from the common gas outlet. 6. Reassemble the system. 7. Perform the Preoperative Checkout Procedure (refer to the 9100c User’s Reference Manual). M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Remove the patient Y-piece from the bag port handle. 10. Reassemble in reverse order. 11. Perform the Preoperative Checkout Procedure (refer to the 9100c User’s Reference Manual). M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 132
9100c 6-10 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
7.4 Alarm and Shutdown messages ....7-8 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Inspect the bellows for damage; replace. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 135
O flow Unable to begin mechanical No O supply Check O supply. ventilation Defective Manual/Mechanical Check Manual/Mechanical ventilation switch. ventilation switch M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Do not use a test plug that is small enough to fall into the breathing system. • Make sure that there are no test plugs or other objects caught in the breathing system. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
The pressure must not decrease. A pressure decrease large enough to see on the gauge indicates an unacceptable leak. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 138
10. Set the system switch to Standby. Make sure that there are no test plugs or other objects WARNING caught in the breathing system. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Power for the speaker comes from the Power Section of the DMB. Drive for the speaker comes from the Digital Section of the DMB board. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
0.5 seconds. to save power. Is the mains indicator light on? Make sure power is connected and circuit breakers are closed. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 141
Does the sensor see 21% O in room air? Calibrate O sensor. Replace O sensor. As sensors wear out, the measured% O decreases. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 142
This Check tubing connections. alarm is suspended for 9 Check alarm settings. breaths after you change the ventilator settings. 7-10 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
8.4 Software Installation ......8-6 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
When new Display Monitor Board is installed, the Function Upgrade options are reset to “not installed” (Section 4.6.1). To help ensure that, after repair or software replacement, the 9100c system will include all features that are currently installed and enabled, record the current Function Upgrade options installed.
The Control Board here only refer to the replacement Display Monitor Note Board for the 9100c system. The replacement Control Sample Board doesn’t need the appropriate optional features Key Code. 2. Access the Key Code Generator on the following web site.
Board kit (or on the Display Module) and the Key Code sticker to the left side panel of the Display Module. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Load the appropriate software revision (Section 8.4) using software USB disk that matches the software revision of the replaced Control Sample Board (CSB). M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
8. On the Altitude Setting menu of the Service Mode, reset the Altitude (Section 4.6.4). 9. Perform the preoperative checkout procedure. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
9.8 Replace O2 supply switch ..... . 9-15 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 150
9.17.1 To remove the Vent Engine ....9-39 9.17.2 Replacing Vent Engine components ..9-40 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 151
After repairs are completed, always perform the checkout procedure before returning the system to clinical use. Refer to Section 3 of this manual. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
(or close the O cylinder valve). • Push the O flush button to bleed O from the system. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
5. Remove the small upper rear panel (B) to access the access the electrical enclosure. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
3. If present, remove the inboard cylinders. 4. To remove the lower access panel, fully loosen the six screws that hold the panel in place. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
(access from below the rim of the tabletop). To remove the tabletop: 1. Loosen the two screws (A). 2. Slide the tabletop forward and lift to remove. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
To remove the upper shelf: 1. Loosen the ten screws (A). 2. Lift the upper shelf. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Note: Make sure to push the new check valve all the way back into the opening until it bottoms out on the shoulder. 6. Install the pipeline inlet fitting. M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
(5) 5. To reassemble, perform the previous steps in reverse order. 6. Perform the checkout procedure (Section 3). 9-10 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Note: A brass retaining ring keeps the filter inside the inlet adapter. 3. Thread a 6-mm screw (two turns only) into the brass retaining ring and pull it out. M1207026 9-11 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
The cylinder check valve is not a replaceable item. If the check valve cylinder check valve is defective, you must replace the complete cylinder supply module. 9-12 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Install the switch assembly through the gauge panel. • Push the knob collar in with the indicator up and turn it clockwise until it locks. M1207026 9-13 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 162
• Make sure that the display turns Off. 11. Reinstall the rear panel. 12. Perform the checkout procedure (Section 3). 9-14 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
10. Disconnect the gauge and plug the test port. 11. To reassemble, perform the previous steps in reverse order. 12. Perform the checkout procedure (Section 3). M1207026 9-15 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Note: Two screws from each side panel. Mounting Screw 8. To reinstall, perform the previous steps in reverse order. 9. Perform the checkout procedure (Section 3). 9-16 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
8. To reassemble, perform the previous steps in reverse order. 9. Complete the port valve checkout procedure described below (Section 9.10.2). M1207026 9-17 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
WARNING If the valve test tool is not removed before flowing gas through the system, pneumatic head damage could result. 9-18 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
The valve body, o-ring, and flapper do not come out with the block. They stay intact at the back of the vaporizer manifold. M1207026 9-19 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 168
16. Install the vaporizer front panel. 17. Perform the checkout procedure (Section 3). 9-20 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
7. Reassemble the machine in reverse order. 8. Perform the checkout procedure (Section 3). M1207026 9-21 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
9. Remove the knob assembly and protective shroud from the machine. 10. Remove the valve from the silicone output tubes. 9-22 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 171
13. Reconnect the wires to the flush pressure switch at the top of the valve(upper and lower terminals). 14. Replace the tabletop. M1207026 9-23 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
6. Lubricate the o-ring sparingly with Krytox (do not get Krytox on the flapper). 7. Insert the flapper assembly into the ACGO outlet with the flapper 8. Replace the cap. 9-24 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
5. Reassemble in reverse order. 6. Reinstall the breathing system. 7. Perform the checkout procedure (Section 3). M1207026 9-25 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
O-rings 5. Replace the Bag/Vent in reverse order. 6. Reinstall the breathing system. 7. Perform the checkout procedure (Section 3). 9-26 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
5. Replace the bag port in reverse order. 6. Reinstall the breathing system. 7. Perform the checkout procedure (Section 3). M1207026 9-27 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
• The inspiratory check valve rises during inspiration and falls at the start of expiration. Perform the post-service checkout procedure. 9-28 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
8. Install the module cover. 9. Install the EZchange module. 10. Verify that the ‘No CO2 absorption’ message appears on the screen when the absorber canister is released. 11. Perform the Preoperative Checkout Procedure (refer to the 9100c User’s Reference manual). M1207026 9-29 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English...
2. Remove the absorber, the vaporizers, gas cylinders, drawers and all auxiliary equipment. To prevent damage, do not tip the 9100c machine more CAUTION than 10 degrees from vertical. 3. Block the opposite wheels; then, block up the machine until there is enough room to remove the defective caster.
3. Remove the switch from the top shelf using the small needle- nose pliers. 4. To replace the switch, reassemble in reverse order. Connectors Switch M1207026 9-31 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
• the rear upper panel (Section 9.2), • the AC inlet access panel (Section 9.2), • the tabletop (Section 9.3). 9-32 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
2. Loosen the screws (C) that hold the two enclosures together (one in each corner). 3. Carefully separate the rear enclosure (A) from the front enclosure (B). M1207026 9-33 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
The flex cable (B) for the LCD display inserts into a ZIF (zero Note insertion force) connector on the DMB. 9-34 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 183
Carefully insert the cable until all the “fingers” are below the surface of the shell. Push the shell to the locked position. M1207026 9-35 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
2. Remove the nut and washer that hold the encoder to the keyboard. 3. Replace the encoder switch in reverse order (note the orientation of the attached ribbon cable). 9-36 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 185
LCD display to the mounting bracket. 2. If required, clean the enclosure window (if new, remove the protective film). 3. Reassembly in reverse order. M1207026 9-37 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Inlet Filter - located under the gas inlet valve (7) • Flow Control Valve (8) • Drive Gas Regulator (9) 9-38 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
6. Loosen the four screws to remove the Vent Engine. 7. To replace the Vent Engine, reassemble in reverse order. M1207026 9-39 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Note orientation of the flow valve. Also inspect the two o-rings that seal it to the manifold. Replace as necessary. 9-40 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
10.14 Breathing System ......10-19 10.14.1 Breathing system - front view ... . 10-19 M1207026 10-1 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 190
10.20.3 Active AGSS ......10-40 10.21 CGO Valve, AGSS Installation, AGSS Flowtube . . 10-42 10-2 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Refer to section 5.1 Low-pressure test device (digital manometer or test Refer to section 5.4.2 gauge) with an accuracy of ±2% of reading M1207026 10-3 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Not Shown Plug, stopper Refer to Section 6.6 M1210946 Tool to help disconnect tubing from Legris fittings 2900-0000-000 Test lung 0219-7210-300 10-4 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Refer to section 10.14 “Display Module” Refer to section 10.16 “Vaporizer manifold” Refer to section 10.10 “Tabletop pan components” Refer to section 10.12 10-6 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Refer to section 10.9 “Rear panel components” Refer to section 10.6 “Anesthetic Gas Scavenging System - AGSS” Refer to section 10.21 M1207026 10-7 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
M1198713-S Adapter, N O inlet and outlet M1169131-S regulator valve assembly M1198712-S Adapter, O inlet M1169127-S Adapter, O outlet M1169128-S M1207026 10-9 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
10.7 AC Power components 8 (9,10) CEE 7/7 AS 3112/GB2099 Euro China CEE 7/4 France Nema 5-15 BS 546 India and South Africa 10-10 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 199
Stud, Equal Potential, 6 mm 0208-0070-300 * Apply heatshrink tubing to terminals of black and white wires if not being used. M1207026 10-11 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
* If Air is selected as drive gas, the legris fitting will be connected to the Air adapter. Replace as necessary. 10-12 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Label Set, cylinder supply, O 1006-3854-000 1006-3855-000 Label Set, cylinder supply, N 1006-3856-000 Label Set, cylinder supply, Air * Lubricate sparingly with Krytox M1207026 10-13 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Screw, M4 x 30, cap head M1168479-S Valve, relief, 5.5 psi, 7/16-20 THD 1006-4128-000 Flexible tubing, 1/4 inch, mixed gas 1001-3064-000 10-14 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Legris Tee connector B Refer to section 10.18 Blank label (for machines without ACGO) M1203029-S * Lubricate sparingly with Krytox. 10-16 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Lock seat, Bellows M1208731-S O-ring, 2.9 ID 6.46 OD 1.78 W EP 70 DURO 1407-3409-000 * Lubricate sparingly with Krytox. M1207026 10-19 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
(or plug) and the cell cable are not included in the breathing system module. ** Lubricate sparingly with Krytox. 10-20 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
O-ring, 17 ID 20.6 OD M1198751-S Handspike bolt M1192913-S O-ring, 3.55 ID 7.15 OD M1168817-S * Lubricate sparingly with Krytox. M1207026 10-21 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
O-ring, 26 ID 31 OD M1198748-S O-ring, 31.5 ID 36.8 OD M1198747-S Pin, BTV M1187381-S * Lubricate sparingly with Krytox. 10-22 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
(does not include absorbent) Cover assembly, CO canister 1009-8240-000 Foam, CO canister (pack of 40) 1407-3201-000 O-ring 1407-3204-000 Canister, CO with handle 1407-3200-000 M1207026 10-23 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
M1168758-S Shaft hook, 61mm M1165452-S O-ring, 12.37 ID 17.61 OD M1169930-S Screw, M4x10 M1168626-S (10) * Lubricate sparingly with Krytox. 10-24 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
10 Illustrated Parts 10.14.9 Flow Sensor Item Description Stock number Flow sensor assembly M1174442-S1 Elbow connector M1171751-S M1207026 10-27 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 216
* Apply Loctite 242. ** Lubricate sparingly with Krytox. *** Rubber valve seats can not be removed from assembly (Item 12). 10-28 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
O-ring, 28.24 ID x 33.48 OD M1173359-S Filter, 2-micron (install course side DOWN) 1504-3708-000 Plug, 6 mm M1169143-S Screw, M4x20 M1168565-S Washer, M4 M1168751-S M1207026 10-29 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Speaker, with harness M1081516-S Screw, M3x6 0140-6219-128 * Refer to Section 8.2 for instructions regarding a replaced DMB or Display Module M1207026 10-31 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Vent pilot pressure OUT – BS Interface drive gas port 210 mm - 1/4 inch 1009-3164-000 CGO Valve – BS interface CGO 135 mm - 1/4 inch 1009-3164-000 10-34 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Harness, Bag/Vent switch 1009-5585-000 Harness, CO bypass 1407-3144-000 Harness, O cell to CSB M1194573-S Harness, CSB to Breathing system interface M1194577-S M1207026 10-35 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
9100c 10.20.2 Adjustable AGSS 4 (5) (9,10) 6 (5) 10-38 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 227
Screw, M6x20 M1189745-S Hook, AGSS M1192883-S Receiver cover, AGSS M1192887-S Bag with 30mm male connector 8004460 * Lubricate sparingly with krytox. M1207026 10-39 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
9100c 10.20.3 Active AGSS 4 (5) (9,10) 6 (5) 14 (23) 10-40 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 229
Active low flow with 30mm taper connector specific parts Connector, 30-mm ISO, male 1406-3555-000 Orifice, low flow 1407-3919-000 * Lubricate sparingly with krytox. M1207026 10-41 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Label, blank (for machines without flow indicator) 1009-3241-000 Flowtube, AGSS 1406-3560-000 Spring, AGSS flowtube M1093853 * For machines without ACGO, CGO assembly will be installed. 10-42 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 231
Schematic, AC inlet module; 220 - 240 V ... . . 11-9 M1207026 11-1 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 232
53. Flow indicator 26. Free breathing check valve 54. To disposal system 27. Atmosphere 55. Active AGSS interface 28. Pop-off valve 11-2 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 233
Note: Active AGSS systems with a 12.7 mm connector do not include the Flow Orifice and the Flow indicator. Figure 11-2 • Gas scavenging circuits M1207026 11-3 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 234
2175 2175 2175 2900 1450 2900 1450 2900 1450 N 2 0 Figure 11-3 • Tubing 11-4 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 235
11 Schematics and Diagrams Figure 11-4 • Wiring harnesses M1207026 11-5 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 236
9100c Figure 11-5 • Ventilator functional block diagram 11-6 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 237
11 Schematics and Diagrams Figure 11-6 • Control sample board block diagram M1207026 11-7 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 238
9100c Figure 11-7 • Display monitor board block diagram 11-8 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 239
Figure 11-8 • Schematic, AC inlet module; 100 - 120 V Figure 11-9 • Schematic, AC inlet module; 220 - 240 V M1207026 11-9 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 240
9100c 11-10 M1207026 M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Page 241
Wangzhuang Industrial Zone Phase II Wuxi, Jiangsu, P.R. China 214028 9100c Tel +86-510-85360178 Technical Reference Manual Fax +86-510-85360119 English M1207026 Printed in China All rights reserved M1207027 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English Reproduced from the electronic master in MATRIX...
Need help?
Do you have a question about the 9100c and is the answer not in the manual?
Questions and answers
The oxygen is internally leaking from the pressure switch valve. What's the possible solution towards this?
@Samwel Try trimming the plastic O2 tubing to make sure it is flat, smooth and seats firmly into the quick connect ports. Other than that you may need to replace the manifold, which I believe is part # M1195025-S.