Time Difference Measurementsg - Kenwood CS-1022 Instruction Manual

Hide thumbs Also See for CS-1022:
Table of Contents

Advertisement

Adjust to the vertical scale
with <»
POSITION
Rise time
Fig. 18
[EXAMPLE]
For the example, the horizontal distance is 4.0 divisions.
The SWEEP TIME/DIV is 2 us. (See Fig. 18)
Substituting the given value:
Risetime = 4.0 (div) x 2 (us) = 8s
Risetime and falltime can be measured
by making use of
the alternate step 3 as described below as well.
4. Use the <»
POSITION control to set the 10% point to
coincide with the center
vertical graduation
line and
measure the horizontal distance to the point of the in-
tersection of the waveform
with the center horizontal
line. Let this distance be D,. Next adjust the waveform
position such that the 90% point coincides with the ver-
tical centerline and measure the distance from that line
to the intersection of the waveform with the horizontal
centerline. This distance is D, and the total horizontal
distance is then D, plus D, for use in the above relation-
ship in calculating the rise time or falltime.
Using the formula:
Risetime = (D, + D,) (div) x (SWEEP TIME/DIV
setting) x 'x 10 MAG"' value! (1/10)
Adjust 90% point to the center
and measure D,.
Adjust 10% point
to the center and
measure D,.
Rise time
Fig. 19
[EXAMPLE]
For the example, the measured D, is 1.8 divisions while D,
is 2.2 divisions.
If SWEEP
TIME/DIV
is 2 us we use the
following relationship. (See Fig. 19)
Substituting the given value:
Risetime = (1.8 + 2.2) (div) x 2 (us) =8 us
TIME
DIFFERENCE
MEASUREMENTS
(Unapplied to
CS-1020 and 1010)
This
procedure
is useful
in measurement
of time
dif-
ferences between two signals that are synchronized to one
another but skewed in time.
Procedure:
1. Apply the two signals to CH1 and CH2 INPUT jacks. Set
the vertical MODE to either ALT or CHOP mode.
Generally
for low frequency
signals CHOP
is chosen
with ALT used for high frequency signals.
2. Select the faster of the two signals as the SOURCE and
use the VOLTS/DIV and SWEEP TIME/DIV to obtain an
easily observed display.
Set the VARIABLE to CAL.
3. Using the 7 POSITION control set the waveforms to the
center of the CRT display and use the <»
POSITION
control to set the reference signal to be coincident with
a vertical graduation line.
4. Measure
the
horizontal
distance
between
the
two
signals and
multiply
this distance
in divisions
by the
SWEEP TIME/DIV setting.
If''x10 MAG" is being used multiply this again by 1/10.
Using the formula:
Time
= Horizontal distance (div) x (SWEEP TIME/DIV
setting) x '' x 10 MAG"' value (1/10)
[EXAMPLE]
For the example, the horizontal distance measured
is 4.4
divisions. The SWEEP TIME/DIV is 0.2 ms. (See Fig. 20)
Substituting the given value:
Time = 4.4 (div) x 0.2 (ms)=0.88 ms
tt
TT
T
Reference signal
Comparison signal
Time difference
Fig. 20
23