8) Adjust the beam to reference line by pressing
9)
After adjustment is complete , press and hold slop button
to save the setting. All the indicators will light up & Buzzer sounds
indicating calibration is completed and setting are saved .
10) After saving the changes, restart the unit, check the X – axis for
accuracy; you should calibrate your instrument again if accuracy is
over spec.
11) Use the same way to calibrate Y-axis.
12) Place the instrument on the vertical application to calibrate Z-axis ,
repeat step 1 and step 2, the X&Y indicators will light up , adjust the
beam to reference line by pressing
to save the calibration.
calibration saved
or
- 9 -
or
, then repeat step 9