HG-600 Geno/Grinder® 2010
9.0 ERROR MESSAGES
9.1 Lid Error
If the lid lock fails while the HG-600 Geno/Grinder® is running, the mill will stop running and
the screen will display a "LID OPEN" message. The timer will maintain the time remaining in
the run. To restart, close the lid and press the green RESUME button to finish the run. To
end the run and reset the timer, press the red STOP button. The "LID OPEN" message will
disappear once the lid is closed.
Figure 27 – Lid Open
9.2 Rate Error
If a "RATE ERROR" message appears on the screen, this indicates the HG-600
Geno/Grinder® is not operating within an acceptable range of the set rate. Dynamic loads
greater than 2 lb. can create a rate error. When stacking 4 or 6 titer plates, do not add
more than five 4 mm steel balls to a single titer plate well. Press the red STOP button to
discontinue operation and contact Cole-Parmer Service Technician at 1-732-623-0465 for
assistance.
1300-87055-1M, REV 6