1.2.
1.2.1.
1.2.2.
1.2.3.
E 1-2
CHARACTERISTICS
Safety Characteristics
Performance Characteristics, Specifications
Properties
expressed
in numerical
values with stated tolerance are guaranteed
by the manufacturer.
Specified non-tolerance values indicate those that could be nominally expected from the mean of a
range of identical instruments.
If not stated otherwise, relative tolerances relate to the set value.
Frequency
frequency range
setting range
—
sine wave
—
square wave
—
triangular wave
—
pos. sawtooth
—
neg. sawtooth
setting
measuring unit
indication
max. resolution
setting error
temperature coefficient
long term drift
aging
phase noise
phase jitter rms
signal -noise ratio (SNR}
0.1 mHz — 2 MHz
0.1 mHz — 2.147 MHz
0.1 mHz — 2.147 MHz
0.1 mHz — 200 kHz
0.1 mHz—
20 kHz
0.1 mHz—
20kHz
Hz, kHz
8-digits
0.1 mHz
£1 x 10-6
< 0.2 ppm/K
< 0.3 ppm
within 7 hours
< 1 ppm/year
< — 80 dBe/Hz
< 3 mrad
= 55 dBc
depending on function and wave form
numerical keys
decimal point key
dimension key Hz/kHz
step function
selectable with key Hz/kHz.
When controlling via IEC/IEEE bus
frequency values can only be entered
in Hz.
7-segment LED-display;
decimal point free selectable
1 kHz from carrier
for a 30 kHz band centred on the carrier
excluding + 1 Hz about the carrier.