ABB AquaProbe FEA200 User Manual

Electromagnetic flowmeter insertion-type flow sensors
Hide thumbs Also See for AquaProbe FEA200:

Advertisement

Quick Links

User Guide OI/FEA100/200–EN Rev. C
AquaProbe FEA100 / FEA200
Electromagnetic flowmeter
Insertion-type flow sensors
Maximum performance,
minimum hassle
Introduction
The AquaProbe FEA100 / FEA200 flow sensor is
designed for measurement of the velocity of water. The
flow sensor is available in four standard lengths and can
be installed in any pipeline of internal diameter from 200
mm (8 in.) to 8000 mm (360 in.), through a small
tapping.
The flow sensor is designed for use in survey
applications such as leakage monitoring and network
analysis and in permanent locations where cost or space
limitations preclude the use of conventional closed pipe
meters.
This User Guide provides installation, connection,
security, start-up and basic setup details for the flow
sensor only. The AquaProbe sensor is available for
operation with either a WaterMaster transmitter (FET100)
or an AquaMaster3 transmitter (FET200).
This User Guide should be used in conjunction with the
following publications:
WaterMaster flowmeter (FEA100):
User Guide – OI/FET100-EN
Programming Guide – IM/WMP
User Guide Supplement, PROFIBUS RS485 Physical
Layer (FEX100-DP) – IM/WMPBS–EN
User Guide Supplement, PROFIBUS FEX100-DP
Parameter Tables – IM/WMPBST–EN
AquaMaster flowmeter (FEA200):
User Guide – OI/FET200-EN
Programming Guide COI/FET2XX-EN
MODBUS Tables Supplement –
COI/FET2XX/MOD/TBL-EN
ScrewDriver profiling and Configuration software:
User Guide – OI/SDR

Advertisement

Table of Contents
loading

Summary of Contents for ABB AquaProbe FEA200

  • Page 1 User Guide OI/FEA100/200–EN Rev. C AquaProbe FEA100 / FEA200 Electromagnetic flowmeter Insertion-type flow sensors Maximum performance, minimum hassle Introduction The AquaProbe FEA100 / FEA200 flow sensor is This User Guide should be used in conjunction with the designed for measurement of the velocity of water. The following publications: flow sensor is available in four standard lengths and can WaterMaster flowmeter (FEA100):...
  • Page 2 We are an established world force in the design and manufacture of instrumentation for industrial process control, flow measurement, gas and liquid analysis and environmental applications. As a part of ABB, a world leader in process automation technology, we offer customers application expertise, service and support worldwide.
  • Page 3: Table Of Contents

    AquaProbe FEA100 / FEA200 Electromagnetic flowmeter – insertion-type flow sensors Safety ............................... 3 Health & Safety ........................3 Electrical Safety – CEI/IEC 61010-1:2001-2 ................3 Symbols – CEI/IEC 61010-1:2001-2 ..................3 Product Recycling Information ....................4 Product Disposal ........................5 Restriction of Hazardous Substances (RoHS) ................
  • Page 4 AAquaProbe FEA100 / FEA200 Electromagnetic flowmeter – insertion-type flow sensors Appendix A ............................27 Velocity Profiles Background ....................27 Testing the Flow Profile for Symmetry ..................29 A.2.1 Partial Velocity Traverse ....................29 A.2.2 Single Entry Point Method .................... 29 A.2.3 Dual Entry Point Method ....................
  • Page 5: Aquaprobe Fea100 / Fea200 Electromagnetic Flowmeter - Insertion-Type Flow Sensors

    AquaProbe FEA100 / FEA200 Electromagnetic flowmeter – insertion-type flow sensors 1 Safety 1 Safety Information in this manual is intended only to assist our customers in the efficient operation of our equipment. Use of this manual for any other purpose is specifically prohibited and its contents are not to be reproduced in full or part without prior approval of the Technical Publications Department.
  • Page 6: Product Recycling Information

    AquaProbe FEA100 / FEA200 Electromagnetic flowmeter – insertion-type flow sensors 1 Safety The equipment is protected through double insulation. This symbol, when noted on a product, indicates a potential hazard which could cause serious personal injury and/or death. The user should reference this instruction manual for operation and/or safety information. This symbol, when noted on a product enclosure or barrier, indicates that a risk of electrical shock and/or electrocution exists and indicates that only individuals qualified to work with hazardous voltages should open the enclosure or remove the barrier.
  • Page 7: Product Disposal

    Note. The following only applies to European customers. ABB is committed to ensuring that the risk of any environmental damage or pollution caused by any of its products is minimized as far as possible. The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/EC) that came into force on August 13 2005 aims to reduce the waste arising from electrical and electronic equipment;...
  • Page 8: Safety Conventions

    1.11 Service and Repairs None of the instrument's components can be serviced by the user. Only personnel from ABB or its approved representative(s) is (are) authorized to attempt repairs to the system and only components formally approved by the manufacturer should be used.
  • Page 9: System Schematic

    AquaProbe FEA100 / FEA200 Electromagnetic flowmeter – insertion-type flow sensors 2 System Schematic 2 System Schematic AquaMaster3 FET200 WaterMaster FET100 Transmitter Transmitter AquaMaster 3 Flow Sensor Fig. 2.1 System Schematic Caution. Care of the Equipment  The tip of the flow sensor is a precision-built part of the equipment and must be handled with care.
  • Page 10: Mechanical Installation

    AquaProbe FEA100 / FEA200 Electromagnetic flowmeter – insertion-type flow sensors 3 Mechanical Installation 3 Mechanical Installation 3.1 Location – Environmental Conditions 60 °C (140 °F) Maximum –20 °C (–4 °F) Minimum A – Within Temperature Limits 10 m (30 ft) IP68 (NEMA 6P) B –...

This manual is also suitable for:

Aquaprobe fea100

Table of Contents