Automatic Co And Odensity Calibration - Sanyo MCO-19MUVH Instruction Manual

Multi gas incubator
Table of Contents

Advertisement

AUTOMATIC CO
 
To calibrate the CO
By connecting a standard gas cylinder that has a density that is the same as the set CO
(usually 5% CO
and 5% O
2
the CO
and O
density display as necessary.
2
2
 
● Setting connected standard gas density
1. Press the MENU Key from the Top Display to display the menu, select Tools, and press the ENTER
Key.
 
 
T e m p
3 7 . 0
1
 
2
37.0
 
3
 
 
4
 
5
S t a t u s
 
6
O K
 
7
 
 
2. Press the Down Cursor Key from the Select Tools (1/2) Display to display the Select Tools (2/2) Display,
and then select Standard Gas Setting, press the MENU Key to display the menu, select OK, and press
the ENTER Key.
 
 
S e l e c t
 
 
T e m p / C O 2 / O 2
 
A l a r m
 
L C D / D A Q
 
U V
S e t t i n g
 
K e y
 
 
D a t e
 
 
 
 
S e l e c t
 
 
S t a n d a r d
 
S t a n d a r d
 
 
 
 
 
 
 
 
 
AND O
2
and O
density, a Semi-automatic one point Gas Calibration Kit (MCO-SG) is available.
2
2
), the CO
sensor and O
2
2
o
C
C O 2
T o o l s
( 1 / 2 )
C a l i b r a t i o n
S e t t i n g
S e t t i n g
L o c k
P W
S e t t i n g
T i m e
T o o l s
( 2 / 2 )
G a s
S e t t i n g
G a s
C a l i b r a t i o n
Get other manuals https://www.bkmanuals.com
DENSITY CALIBRATION
2
sensor can be calibrated to maintain the accuracy of
2
5 . 0 %
O 2
5.0
69
and O
density
2
2
5 . 0 %
M E N U
5 .
5.0
S e t
L o g
T o o l s
M E N U
O K
S v c
C a n c e l
M E N U
O K
S v c
C a n c e l