Remote alarm terminal
The remote alarm terminal is located at the rear right side of the incubator.
The alarm is outputted from this terminal. Contact capacity is DC 30 V, 2 A.
Contact output:
between COM. and N.O.
At normal
At abnormal
Note:
The alarm is actuated when the power cord is disconnected from the outlet or the power switch is OFF.
• The remote alarm cannot be stopped by pressing the alarm buzzer stop key (BUZZER) since the remote
alarm is not conjunct with the alarm buzzer stop key (BUZZER).
between COM. and N.C.
Open
Close
Remote alarm terminal