2-6
Checks and adjustments
1
Adjust the collimator for desired position, see Figure 2 and Figure 3, by loosening
the four Allen screws and carefully displacing it. Check against free space X-axis and
Y-axis.
2
Tighten the Allen screws.
3
Release an exposure.
4
Compare the white borders on the film with those on the film exposed before work-
up. Adjust if necessary.
Siemens-Elema AB
Solna, Sweden
Modification of collimator holder
SRX-MK
RX B7-230.092.01.01.02
REV.01
10.95
MAMMOMAT 3000
Collimator holder