MAMMOMAT 1000/3000/3000 Nova - Opdima?
Installation
Installation and Start-Up Instructions
Register 3
Print No.: SPB7-230.033.10.03.02
Replaces: SPB7-230.033.10.02.02
© Sie mens AG 2002
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model
_or_ design,_are_
reserved.
English
Doc. Gen. Date: 08.02
66 08 744

Advertisement

Table of Contents
loading

Summary of Contents for Siemens Opdima MAMMOMAT 1000

  • Page 1 MAMMOMAT 1000/3000/3000 Nova - Opdima? Installation Installation and Start-Up Instructions © Sie mens AG 2002 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model...
  • Page 2 0 - 2 Revision Chapter Page Revision Opdima? Register 3 SPB7-230.033.10 Page 2 of 4 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 3: Table Of Contents

    Replacing software ....... . . 5 - 7 Installing isolation board D707......5 - 7 Siemens-Elema AB Register 3 SPB7-230.033.10...
  • Page 4 Performing of DHHS maintenance ......6 - 2 7 ______ Changes to previous version _____________________________________ 7 - 1 Opdima? Register 3 SPB7-230.033.10 Page 4 of 4 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 5: General

    Upgrade of the MAMMOMAT software (generator and stand) to V2.2 (part No. 62 98 769, if the version to be upgraded is lower than V1.5, the AEC have to be upgraded, see Update Instruction, UI 001/96) Siemens-Elema AB Register 3 SPB7-230.033.10...
  • Page 6: Training Of Customer Support Engineers

    NOTE contains information provided with special emphasis to NOTE facilitate proper use of the equipment or proper execution of a procedure, i.e. hints, tips. Opdima? Register 3 SPB7-230.033.10 Page 2 of 4 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 7: Documents Required

    • Protective ground wire tester (44 15 899 RV090) • Service PC (e.g. Siemens-Nixdorf PCD 3NSX/20 or similar) with connecting cable, PC to generator (part No. 99 00 440 RE999) • Stereo calibration phantom (included in the delivery) (part No.64 30 701) •...
  • Page 8: Start-Up, Dhhs And X-Ray Decree Area Of Application

    Automatic exposure control (AEC) The result must be recorded in the document MAMMOMAT 300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy Attachment and Opdima? Measurement Certificates. Opdima? Register 3 SPB7-230.033.10 Page 4 of 4 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 9 Do not touch the pins in the contacts of the camera. The camera shall be used within 1030? C. The camera shall be transported or stored within 040? C. Siemens-Elema AB Register 3 SPB7-230.033.10 Page 1 of 2...
  • Page 10 2 - 2 Protective measures for CCD camera This page intentionally left blank. Opdima? Register 3 SPB7-230.033.10 Page 2 of 2 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 11 Risk of damaging components. Use ESD-equipment, ground prior to making contact and place the components on a conductive surface, see Technical Informa- tion TI 219. Siemens-Elema AB Register 3 SPB7-230.033.10 Page 1 of 2 Opdima? Solna, Sweden Rev.
  • Page 12 3 - 2 Protective measures This page intentionally left blank. Opdima? Register 3 SPB7-230.033.10 Page 2 of 2 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 13: Scope Of Delivery

    Printed circuit board D702 • MAMMOMAT software V2.2 or higher • Isolation p.c. board D707 kit • DICOM license key • Workstation table • Printer Siemens-Elema AB Register 3 SPB7-230.033.10 Page 1 of 2 Opdima? Solna, Sweden Rev. 03 08.02 SPS-UD Installation...
  • Page 14: Unpacking

    *. Only needed if a software reinstallation is necessary (see MAMMOMAT 1000/3000/3000 Nova - Opdima? Service Instructions). Store the documentation and software CDs in a proper place for future service use. Opdima? Register 3 SPB7-230.033.10 Page 2 of 2 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 15: General

    Fig. 1 MAMMOMAT 3000 Nova with digital biopsy and spot imaging system Opdima NOTE The workstation shall be placed at least 1.5 m from the object table of the stand. Siemens-Elema AB Register 3 SPB7-230.033.10 Page 1 of 26 Opdima? Solna, Sweden Rev.
  • Page 16: Removal Of Mammomat Stand Covers

    Remove the two screws (7/Fig. 2) at the top of the stand. Remove the cover (8/Fig. 2) by lifting it up. Fig. 2 Removal of covers Opdima? Register 3 SPB7-230.033.10 Page 2 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 17: Mammomat Stand With Single Rear Cover

    Pry loose the plastic strips (2/Fig. 3) on both sides using a screwdriver or similar tool. Remove the screws (3/Fig. 3) on both sides and remove the cover (4/Fig. 3). Fig. 3 Removal of covers Siemens-Elema AB Register 3 SPB7-230.033.10 Page 3 of 26 Opdima? Solna, Sweden Rev.
  • Page 18: Installation Of Biopsy Controller Holders

    Install the two standoffs (4/Fig. 4) by using the screws (5/Fig. 4) and the washers (6/Fig. 4). 985 – 5 mm Fig. 4 Rear cover, from the back side Opdima? Register 3 SPB7-230.033.10 Page 4 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 19 Install the rear cover (2/Fig. 5) on the stand and attach the biopsy controller (1/Fig. 5) to the holders. Switch on the power-on switch on the biopsy controller. Fig. 5 Biopsy controller Siemens-Elema AB Register 3 SPB7-230.033.10 Page 5 of 26...
  • Page 20: Modification Of Mammomat Stand

    Install the new board D702 Fig. 7 in the generator. If upgrading from a version lower than V1.5, the AEC have to be NOTE upgraded, see Update Instruction, UI 001/96. Opdima? Register 3 SPB7-230.033.10 Page 6 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 21: Replacing Software

    If not already done in conjunction with installation of printer (option), an isolation board D707 (see Fig. 10) shall be installed. Install the board according to Installation Instructions for Isolation P.c. Board D707, RX B7-120.031.03..Siemens-Elema AB Register 3 SPB7-230.033.10...
  • Page 22: Restoring Installation Parameters

    Select Main menu => Configuration => Mischelaneous=> Cassette loaded switch and set the switch to ON. Switch off the MAMMOMAT with the main circuit breaker. Opdima? Register 3 SPB7-230.033.10 Page 8 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 23: Connection Of Cables

    Z801 X880 Printer M3000 printer/ M3000 stand (option) Service PC X885 Biopsy unit M3000 generator (M3000 stand) Fig. 8 Connection of cables Siemens-Elema AB Register 3 SPB7-230.033.10 Page 9 of 26 Opdima? Solna, Sweden Rev. 03 08.02 SPS-UD Installation...
  • Page 24: Cables To Mammomat Generator (And Printer If Present)

    Use compartment according to the figure below. Serial number higher than 3000 Cable X206-X2 Serial number lower than 3000 Fig. 9 Cable-duct compartments Opdima? Register 3 SPB7-230.033.10 Page 10 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 25 This is to ensure electromagnetic compatibility of the equipment. Chassie ground clamp Fig. 10 Connection of cable to the generator Siemens-Elema AB Register 3 SPB7-230.033.10 Page 11 of 26 Opdima? Solna, Sweden Rev.
  • Page 26 The wall socket must be marked according to Fig. 24. Close the cable duct cover and install the front cover of the generator. Opdima? Register 3 SPB7-230.033.10 Page 12 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 27: Cables To Biopsy Unit

    1 (BN) 2 (BU) 3 (GNYE) Protective earth cable X885 Fig. 12 Connection of mains cable, protective earth cable and X885, serial number higher than 3000 Siemens-Elema AB Register 3 SPB7-230.033.10 Page 13 of 26 Opdima? Solna, Sweden Rev. 03 08.02...
  • Page 28: Cables To Ccd Camera

    Remove the screws (1/Fig. 14) and remove the cover (2/Fig. 14) of the object table arm. Fig. 14 Object table arm before installation Opdima? Register 3 SPB7-230.033.10 Page 14 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 29 Install a new thread (1/Fig. 15). Fig. 15 Installation of thread Install the holder (1/Fig. 16) with the pull relief device, using included Allen screws (2/Fig. 16). Fig. 16 Object table arm after installation Siemens-Elema AB Register 3 SPB7-230.033.10 Page 15 of 26 Opdima? Solna, Sweden Rev.
  • Page 30: Installation Of Cable Duct

    Attach the supplied plastic strip around the opening of the cable outlet cover (4/Fig. 17). Install the cable outlet cover (4/Fig. 17). Mount the previously removed stand covers. Opdima? Register 3 SPB7-230.033.10 Page 16 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 31: Installation Of Workstation

    Connect the cable from the control unit (2/Fig. 18) for the vertically adjustment to the connector marked “Handschalter” (3/Fig. 18). Connect the cable from the motor (4/Fig. 18) for the vertically adjustment to the other connector (5/Fig. 18). Siemens-Elema AB Register 3 SPB7-230.033.10 Page 17 of 26...
  • Page 32 Fig. 19 System site overview NOTE There must be enough space around the workstation table so that it safely can be vertically adjusted . Opdima? Register 3 SPB7-230.033.10 Page 18 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 33 Place the main unit to the right on its shelf (1/Fig. 20). Mount the MO unit shelf close to the main unit (2/Fig. 20). Place the MO unit to the right on the shelf (2/Fig. 20). Siemens-Elema AB Register 3 SPB7-230.033.10...
  • Page 34: Connecting The Workstation

    -Place the plastic cover (2/Fig. 22) on the other side of the connectors. -Fasten the plastic screws (3/Fig. 22). Biopsy controller Fig. 22 Mounting the touch protection Opdima? Register 3 SPB7-230.033.10 Page 20 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 35 Work- station station Power On-Off breaker Power On-Off breaker Power connection Power connection in MAMMOMAT 3000-stand in MAMMOMAT 3000-stand Fig. 23 Power distribution Siemens-Elema AB Register 3 SPB7-230.033.10 Page 21 of 26 Opdima? Solna, Sweden Rev. 03 08.02 SPS-UD Installation...
  • Page 36 Connect the wall socket for the workstation and printer power supply to the same branch circuit as the MAMMOMAT power sup- ply. (The workstation table can be connected to any branch circuit.) Opdima? Register 3 SPB7-230.033.10 Page 22 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 37 10 Camera interface via biopsy controller (X203) Keyboard interface 11 Monitor interface 6A Mammomat interface via biopsy controller (X204) 6B Biopsy controller interface (X204) Fig. 25 Workstation main unit connectors Siemens-Elema AB Register 3 SPB7-230.033.10 Page 23 of 26 Opdima? Solna, Sweden Rev.
  • Page 38: Connecting The Workstation To The Network

    Connect one end of the network cable (included in the delivery) to the workstation. Connect the other end of the network cable to a network socket. Log in with the user name “service”. The password is obtained from Siemens Service Centre.
  • Page 39 Add access to the network printer in the advanced service dialog, see Networked Opdima system in MAMMOMAT 1000/3000/3000 Nova - Opdima? Service Instructions. Select printer in the advanced service dialog, see Networked Opdima system in MAMMOMAT 1000/3000/3000 Nova - Opdima? Service Instructions. Siemens-Elema AB Register 3 SPB7-230.033.10 Page 25 of 26 Opdima? Solna, Sweden Rev.
  • Page 40: Strain Relieving The Cables

    Before the installed workstation table (option) is handed over to the customer, check that the workstation table can be vertically adjusted between the highest and lowest possible positions properly and safety. Opdima? Register 3 SPB7-230.033.10 Page 26 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 41: General

    Filling in of LINA CARD, see Page 6 - 2 • Performing of DHHS maintenance, see Page 6 - 2 The Opdima? external diaphragm must be used for tests and cali- NOTICE brations involving radiation. Siemens-Elema AB Register 3 SPB7-230.033.10 Page 1 of 4 Opdima? Solna, Sweden Rev.
  • Page 42: Starting Up The Mammomat

    The workstation is started according to instructions in Starting up the workstation and login in Supplement to the Instructions for Use MAMMOMAT 3000 - Opdima?. Use the user name “service” and the password obtained from Siemens service center when entering the system.
  • Page 43 MAMMOMAT. Therefore the DHHS Maintenance Instructions supplied with the MAMMOMAT can be removed. However, the Measurement Certificates supplied with the MAMMOMAT should be kept for future reference, if it contains previous measurement results. Siemens-Elema AB Register 3 SPB7-230.033.10 Page 3 of 4 Opdima? Solna, Sweden Rev.
  • Page 44 6 - 4 Final procedures This page intentionally left blank. Opdima? Register 3 SPB7-230.033.10 Page 4 of 4 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 45 Text change in Caution. Notice G Note, Notice G Caution (text change). Page 5 - 24 Notice G Note. Page 5 - 25 Notice G Note. Page 6 - 3 Siemens-Elema AB Register_3 SPB7-230.033.10 Page 1 of 2 Opdima? Solna, Sweden Rev. 03 08.02...
  • Page 46 7 - 2 Changes to previous version This page intentionally left blank. Opdima? Register_3 SPB7-230.033.10 Page 2 of 2 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
  • Page 47 150 mm Drilling template A Drill five holes with a 5 mm drill according to template A and B. Place template B underneath template A with no space between them. Check distances before drilling!
  • Page 48 This page intentionally left blank...
  • Page 49 140 mm 195 mm (to next hole) OPD00269...
  • Page 50 This page intentionally left blank...