Embedded Computer
Programming
(C-Size Systems)
IBASIC Programming
Absolute Addressing
and Select Code 8
The Register Offset and
Register Number
External Computer
Programming
80 16-Channel Relay Multiplexer Registers
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
If the relay multiplexer is part of a C-Size VXI system, the fastest throughput
is achieved using an embedded computer. The embedded computer allows
you to access the registers from the VXIbus backplane, and thus, there is no
parsing of SCPI command headers.
When the relay multiplexer is programmed using the E1300/E1301
Mainframe's Instrument BASIC (IBASIC), two methods of accessing the
registers are through absolute addressing or using select code 8.
Absolute addressing is faster than select code 8 since the complete register
address (including the A16 starting location 1FC000
select code 8 is used, the IBASIC processor must calculate the complete
register address based on the logical address specified
Depending on whether absolute addressing or select code 8 is used, either a
register offset or register number is specified as part of the register address.
Absolute addressing specifies a register offset, which is the register's
location in the block of 64 address bytes. For example, the multiplexer's
Channel Register has an offset of 08
register, the offset is added to the base address to form the register address
(using a logical address of 80):
register address = base address + register offset
= 1FC000
+ (80 * 64)
16
= 1FC000
+ 1400
16
or
= 2,080,768 + (80 * 64) + 8
= 2,080,768 + 5120 + 8 = 2,085,896
Using select code 8 requires that you specify a register number. The
register number is the register offset/2. Referring to
Register with an offset of 08, is register number 4.
When the multiplexer is programmed by an external computer through the
E1300/E1301 Mainframe or E1405/E1406 Command Module, the registers
are accessed using
DIAG:PEEK?
.
VXI:WRITE
DIAG:PEEK?/DIAG:POKE
Throughput speed using
DIAG:PEEK?
VXI:READ?
and
VXI:WRITE
(including the A16 starting location 1FC000
and
specify the device logical address and register offset only.
VXI:WRITE
Thus, the E1300/E1406 processor must calculate the complete register
address which decreases throughput speed.
IBASIC programming using absolute addressing or select code 8 is faster
than either
DIAG:PEEK?
and
because the registers are accessed from the VXIbus backplane rather than
from the HP-IB. Also,
READIO
) is specified. When
16
(Figure
. When you write a command to this
16
+ 08
16
16
+08
= 1FD408
16
16
16
Figure
and
, or
DIAG:POKE
VXI:READ?
and
VXI:READ?/VXI:WRITE
and
DIAG:POKE
because the complete register address
) is specified.
16
DIAG:POKE
or
VXI:READ?
and
WRITEIO
are not parsed.
B-1).
B-1, the Channel
and
is faster than
VXI:READ?
and
VXI:WRITE
Appendix B