It is important to remember fever is a symptom, not a disease, and is just one of the many possible signs of illness. If your child exhibits signs of a serious illness, contact your physician.
3 Simple Steps:
Note: Users have to shut down the thermometer and start up again to take another measurement and/or change age setting. The Braun Age Precision™ Thermometer turns off automatically 20 seconds after the measurement or after 90 seconds without measurement.
Braun Age Precision™ colour-coded display that offers a colour indication along with the temperature readout and helps you better understand the temperature reading based on patient's age. The associated colour will display 5 sec after the temperature reading was completed.
The Age Precision™ colour-coded display for interpreting of fever has been designed for use in the underarm mode only according to the table below.
If you use the thermometer in the oral or rectal mode by placing it at those sites, please do not use the display lights programmed into the thermometer for determining fever. Instead, refer to the oral and rectal tables respectively as shown below to determine if you have a fever.
Note: Due to human physiology, temperature measurements taken from different sites on the same person show variability. For example, rectal readings tend to run higher than oral as well as underarm readings taken from the same person.
This method is used for babies and young children.
This method is suitable for children who are old enough to safely keep the thermometer inside the mouth. We recommend using a universal disposable (single-use) probe cover specifically intended for stick thermometers when taking an oral measurement. Such probe covers are generally available in stores from various manufacturers.
Commonly used for babies, young children, or when it is diffi cult to take an oral or underarm temperature.
When the thermometer is turned on, the display will show a test, after which the last measured temperature will be shown for approximately 1 second. Letter "M" will be displayed next to the temperature reading.
When the battery symbol on LCD display twinkles continually , please replace with a new battery.
When measured temperature is higher than 42.9°C, LCD will display "Err °C" and relevant age icon. Red backlight stays on for 5 seconds.
When ambient temperature is out of range (see Product Specifi cations below) or other system errors occur such as short or broken circuit, LCD will display "Err" to indicate system error. Red backlight on for 5 seconds.
This device was designed and manufactured for a long service life, however it is generally recommended to have the device inspected once a year by the Authorized Service Center located in your country to ensure correct function and accuracy.
Note: The accuracy checking is not a free service and therefore we recommend that you contact the Authorized Service Center to get a quotation before you send out the product.
Displayed temperature range: | 32.0°C – 42.9°C (89.6°F – 109.2°F) |
Operating ambient temperature range: | 10 – 40°C (50 – 104°F) and relative humidity range of 15%~95% |
Display resolution: | 0.1°C or °F |
Accuracy for displayed temperature range 35.5°C – 42°C (95.9°F – 107.6°F): | Maximum Laboratory Error ± 0.1°C (± 0.2°F) at an ambient of 23±5°C |
outside this range: | ±0.2°C/0.4°F (within 32.0°C – 42.9°C (89.6°F – 109.2°F)) |
Battery type: | 3V DC, 1X CR1632 |
Battery life: | 500 times measurements within 2 years |
Service life: | 2000 measurements |
Storage/Transport temperature and relative humidity: | -25 – 55°C ± 2°C and 15-95 RH% ± 5% |
If the device is not used within specifi ed temperature and humidity ranges the technical accuracy of the measurement cannot be guaranteed.
Type BF applied parts
See instruction for use
Operating temperature
Storage temperature
Keep dry
Internally powered equipment.
Continuous operation.
Not intended to be sterilised.
Not for use in an oxygen rich environment.
IP22: Protected against solid foreign objects of 12.5 mm diameter and greater. Protected against vertically falling water drops when the device is tilted up to 15 degrees.
This thermometer is specifi ed to operate at 1 atmospheric pressure or at altitudes with an atmospheric pressure up to 1 atmospheric pressure (760-1060 hPa).
Equipment with type BF applied parts.
Subject to change without notice.
This appliance conforms to the following standards:
Standard Reference Edition Title: EN 12470-3: 2003 Clinical thermometers – Part 3: Performance of compact electrical thermometers (non-predictive and predictive) with maximum device.
EN 60601-1: 2006 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance.
EN ISO 14971: 2012 Medical devices – Application of risk management to medical devices.
EN ISO 10993-1: 2009 Biological evaluation of medical devices – Part 1: Evaluation and Testing.
EN 60601-1-2: 2007 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: electromagnetic compatibility – Requirements and tests.
EN 980: 2008 Symbols for use in labeling of medical devices.
EN 1041: 2008 Information supplied by the manufacturer of medical devices.
EN 60601-1-11: 2010 Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance – Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. This product conforms to the provisions of the EC directive 93/42/EEC.
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC. For detailed description of EMC requirements please contact your authorized local Service Centre (See guarantee card). Portable and mobile RF communications equipment can aff ect MEDICAL ELECTRICAL EQUIPMENT.
This product contains batteries and recyclable electronic waste. To protect the environment, do not dispose of it in the household waste, but take it to appropriate local collection points.
1Herzog L, Phillips SG. Addressing concerns about fever. Clin Pediatr (Phila) 2011 May; 50(5):383-90.
2Sund-Levander M, Forsberg C, Wahren LK. Normal oral, rectal, tympanic and axillary body temperature in adult men and women: a systematic literature review. Scand J Caring Sci 2002 June; 16(2):122-8.
Guidance and manufacturer's declaration – electromagnetic emissions | |||||
The PRT2000 equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the PRT2000 should assure that it is used in such an environment. | |||||
Emissions Test | Compliance | Electromagnetic environment – guidance | |||
RF Emissions CISPR 11 | Group 1 | The ME equipment uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. | |||
RF Emissions CISPR 11 | Class B | Complies | |||
Harmonic emissions IEC 61000-3-2 | Not Applicable | The ME equipment is solely battery powered. | |||
Voltage fl uctuations / flicker emissions | Not Applicable | ||||
Non-Life Support Equipment Separation Distance Calculation (3Vrms / 3V/m compliance) | |||||
Rated maximum output power of transmitter (W) | Separation distance according to frequency of transmitter (m) | ||||
150 kHz to 80 MHz in ISM bands![]() | 80 MHz to 800 MHz ![]() | 800 MHz to 2.5 GHz ![]() | |||
0.01 | / | 0.12 | 0.23 | ||
0.1 | / | 0.38 | 0.73 | ||
1 | / | 1.2 | 2.3 | ||
10 | / | 3.8 | 7.3 | ||
100 | / | 12 | 23 |
⠀
Guidance and manufacturer's declaration – electromagnetic immunity | |||
The PRT2000 is intended for use in the electromagnetic environment specified below. The customer or the user of the PRT2000 should assure that it is used in such an environment. | |||
Immunity test | IEC 60601 test level | Compliance level | Electromagnetic environment – guidance |
Electrostatic discharge (ESD) IEC 61000-4-2 | ±6kV Contact ±8kV Air | Complies | Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. |
Radiated RF IEC 61000-4-3 | 3 V/m 80MHz to 2.5GHz | Complies | Field strengths outside the shielded location from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than 3 V/m. Separation distance calculation provided above. If a known transmitter is present the specifi c distance can be calculated using the equations. |
Conducted RF IEC 61000-4-6 | 3Vrms 150kHz to 80MHz | Not Applicable (no electrical cabling) | |
Electrical fast transient IEC 61000-4-4 | ±2kV power line ±1kV I/O lines | Not Applicable | The ME equipment is solely battery powered. |
Surge IEC 61000-4-5 | ±1kV differential ±2kV common | Not Applicable | |
Power frequency magnetic field IEC 61000-4-8 | 3 A/m | Complies | Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. |
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 | >95% dip 0.5 cycle 60% dip 5 cycles 70% dip 25 cycles 95% dip 5 sec. | Not Applicable | The ME equipment is solely battery powered. |