6 Reference information
14
Switch time
Operating life time
Environment
Physical data
Internal volume, in/out to port
IV-908
PV-908
Flow channel diameter
IV-908
PV-908
Valve principle
Functions
Degree of protection
Wetted materials
Channel and
Distribution plates
Chemical resistance
Power requirement
Power consumption
UniNet 2 address
Inlet and outlet tubing
IV-908
PV-908
Dimensions,
H x W x D
Weight
EMC Standards
<260 ms between two adjacent positions
>50 000 cycles, two adjacent positions
+4 to +40 °C
20-95% relative humidity 84–106 kPa (840–1060 mbar)
atmospheric pressure
26 μl
7 μl
1.2 mm
0.8 mm
Motor controlled valve
Switching 8-way, 8 positions controlled from UNICORN
IP 43
PEEK (polyetheretherketone)
The wetted parts are resistant to organic solvents and
salt buffers commonly used in chromatography of
biomolecules, except 100% ethylacetate, 100% hexane
and 100% tetrahydrofuran (THF)
32 V DC ±10% from the system pump
Up to 9 W
0–9
5/16"-24 UNF 2B for tubing with 3/16" outer diameter
UNF 10–32 2B "Fingertights" for capillary tubing 1/16"
outer diameter
135 × 80 × 120 mm
1.2 kg
This product meets the requirement of the EMC Directive
89/336/EEC through the harmonized standards
EN 50081-1 (emission) and EN 50082-1 (immunity)
Note: The declaration of conformity is valid for the
instrument when it is
• used in laboratory locations
• used in the same state as it was delivered from
GE Healthcare except for alterations described
in the user manual
• connected to other CE labelled GE Healthcare
instruments or other products as recommended.