3M Clean-Trace LM1 User Manual page 31

Luminometer
Hide thumbs Also See for Clean-Trace LM1:
Table of Contents

Advertisement

Synchronization
Instrument not registered
Failure
Instrument
becomes
Temporary software malfunction
unresponsive
Unexpectedly
Instrument was dropped or
high-test results
subjected to a physical shock
Error: 7
Firmware upgrade problem
Error: 9
Cap was open while reading a test
Error: 11
Instrument temperature is > 40°C
Error: 15
Error: 12
Instrument temperature is < 5°C
Error: 16
Error: 1
Error: 2
Error: 4
Error: 6
Error: 8
Error: 10
Error: 13
Error: 14
Error: 28
Error: 17
Error: 31
Error: 18 - 30
Error: 33 - 123
For USB or Bluetooth® synchronization, make sure you are
signed-in to the Sync Manager on the workstation, then initiate
a synchronization again.
For Wi-Fi synchronization, make sure the Luminometer has
3M™ Clean-Trace™
been registered in the
1. Re-start the instrument: Press and hold the On/Off button for
about 3 to 5 seconds. Then tap [Restart].
2. Contact 3M or your local distributor if the failure persists.
1. Switch off the power: Press and hold the On/Off button for
about 3 to 5 seconds. Then tap [Power off].
2. Wait at least 5 minutes, then switch the power On.
3. Contact 3M or your local distributor if the failure persists.
1. Re-start the instrument: Press and hold the On/Off button for
about 3 to 5 seconds. Then tap [Restart].
2. Re-run the firmware upgrade.
3. If the problem persists, contact 3M or your local distributor.
1. Close the cap.
2. Repeat reading the test.
3. If the problem persists, contact 3M or your local distributor.
1. Use the 3M Clean-Trace Luminometer at ambient
temperature less than or equal to 40°C.
2. If the problem persists, contact 3M or your local distributor.
1. Use the 3M Clean-Trace Luminometer at ambient
temperatures greater than or equal to 5°C.
2. If the problem persists, contact 3M or your local distributor.
1. Re-start the instrument: Press and hold the On/Off button for
about 3 to 5 seconds. Then tap [Restart].
2. Repeat the task using the 3M Clean-Trace Luminometer.
3. If the problem persists, contact 3M or your local distributor.