Result Message Examples

2062665-001 G
Sample Messages: Transactions from the monitor HL7 interface
OBR — Observation request segment
OBX — Observation result segment(s)
The structure of an ORU message looks like this:
MSH
PID
PV1
{
OBR
{OBX}
}
All message segments are terminated using a segment separator character,
which defaults to a carriage return.
The open and closed braces { } represent segments and segment groups.
Segments enclosed in square brackets [ ] are optional.
Trailing data fields for a segment are truncated if empty.
Following are several examples of the many possible HL7-formatted result
messages.
The monitor HL7 outbound interface can generate countless variations of these
examples.
Example 1
This example has one OBR segment followed by the corresponding OBX
segments collected at a user-configured time interval. The OBX segments
contain periodic vital sign data for heart rate, PVC, temperature, and invasive
blood pressures.
MSH|^~\&|VSP^080019FFFE134535^EUI-64|GE
Healthcare|||20130122134100||ORU^R01^ORU_R01|00409713
4535|P|2.6|||NE|AL||UNICODE UTF-8|||PCD_DEC_001^IHE
PCD^1.3.6.1.4.1.19376.1.6.1.1.1^ISO
PID|||HED12^^^PID^MR||LAZY^KITTY^^^^^L|||
PV1||R|SURGERY^^V2
OBR|1|080019FFFE13453520130122134100^VSP^080019FFFE13
4535^EUI-
64|080019FFFE13453520130122134100^VSP^080019FFFE13453
5^EUI-64|182777000^monitoring of
patient^SCT|||20130122134100
OBX|1||69965^MDC_DEV_MON_PHYSIO_MULTI_PARAM_MDS^MDC|1
.0.0.0|||||||X
19