Hl7 Communication; Observation Results Interfaces - GE B40 Reference Manual

Patient monitor
Hide thumbs Also See for B40:
Table of Contents

Advertisement

HL7 Communication

Observation results interfaces

2062665-001 G
HL7 Interface: HL7 Standard Background
HL7 messages are passed between computer systems whenever a valid
transaction occurs. For example, receiving a completed test result on the
monitor system would cause an HL7 message to be generated and sent. After a
message is sent, the receiving system processes the message. When processing
is complete, the receiving system is designed to generate an application level
acknowledgment that is returned to the sending system.
Since the HL7 standard provides flexibility in message content and format, and
in communication protocol options, its implementation requires mutual
agreement between the sending and receiving computer systems on the
following items:
message formats
acknowledgment protocol
communication protocol
data handling
Communication between HIS personnel and GE interface personnel is essential
to determine the customer-specific use of the HL7 standard.
NOTE: This document is not intended to provide instructions for implementing
and using the HL7 standard. Details of the HL7 standard can be found in
Health Level Seven Version 2.6.
The HL7 observation results reporting interface for this device is for unsolicited
trended data. ORU messages are sent unsolicited to the host at a user-
configured time interval. The monitor HL7 outbound interface pauses to wait for
an application level acknowledgment form the host. The HIS typically uses the
trended result message content in either of the following ways:
To make test results available for access on the hospital computer system.
To generate charges for completed tests.
7