About this document Liquiline Mobile CML18 About this document Warnings Structure of information Meaning This symbol alerts you to a dangerous situation. DANGER Failure to avoid the dangerous situation will result in a fatal or serious injury. Causes (/consequences) If necessary, Consequences of non- compliance (if applicable) ‣...
The battery may only be changed directly at the manufacturer' s premises or by the service organization. Intended use The Liquiline Mobile CML18 is a multiparameter mobile device for the connection of digital sensors with Memosens technology and optional operation by smartphone or other mobile devices via Bluetooth.
Liquiline Mobile CML18 Basic safety instructions Operational safety Before commissioning the entire measuring point: Verify that all connections are correct. Ensure that electrical cables and hose connections are undamaged. Do not operate damaged products, and protect them against unintentional operation.
Product description Liquiline Mobile CML18 Product description Product design A0040968 1 CML18 Protection cap Display screen with automatic screen rotation "Select" button "Next" button Memosens connection Area for wireless charging Status LED M12 connection Endress+Hauser...
Page 9
Liquiline Mobile CML18 Product description 3.1.1 Measuring parameters The mobile device is designed for digital Memosens sensors with an inductive plug-in head and fixed cable sensors with the Memosens protocol and no external power supply: • pH • ORP • pH/ORP combination sensors •...
Incoming acceptance and product identification Liquiline Mobile CML18 Incoming acceptance and product identification Incoming acceptance Verify that the packaging is undamaged. Notify the supplier of any damage to the packaging. Keep the damaged packaging until the issue has been resolved.
Dieselstraße 24 D-70839 Gerlingen Scope of delivery The scope of delivery comprises: • 1 Liquiline Mobile CML18 • 1 set of Operating Instructions in German • 1 set of Operating Instructions in English Inductive charger and power unit are available separately.
Electrical connection Liquiline Mobile CML18 Electrical connection Connecting the sensor 5.1.1 Connecting the Memosens sensor directly A0040973 2 Sensor connection Insert the sensor into the Memosens connection. Click the Memosens connection into place. 5.1.2 Connecting the Memosens sensor with M12 fixed cable connection...
Liquiline Mobile CML18 Electrical connection 5.1.3 Connecting the sensor via the Memosens M12 cable The M12 cable has two different connectors: • M12 connector for connecting to the device • Memosens connection for connecting a Memosens sensor A0040974 A0040972 A0041005 ...
Page 14
Electrical connection Liquiline Mobile CML18 The following applies to both options: • When device is switched on: • When charging starts, a flash symbol appears on the display and a confirmation tone sounds. • If charging stops before the battery is fully charged, another confirmation tone sounds.
Liquiline Mobile CML18 Electrical connection An acoustic signal indicates that charging is complete. During inductive charging, measurement via the integrated Memosens connection on the device is not possible. A message to this effect is shown on the display. Measurement via M12 cable is still possible.
Operation options Liquiline Mobile CML18 Operation options Overview of operation options There are two options for operating and configuring the device: • Internal operating menu with keys • SmartBlue app via Bluetooth® LE wireless technology → 20 Internal operating menu with keys 6.2.1...
Page 17
Liquiline Mobile CML18 Operation options 6.2.2 Structure and function of the operating menu Power-off Power-off Application Data logger Data logger Log interval Cond. unit Res. unit Erase data Erase grab values Abort Erase Erase continuous logs Abort Erase Data logger plot...
Page 18
Operation options Liquiline Mobile CML18 System/Language Power management Power save w. charger Power save w/o charger Power-off w. charger Power-off w/o charger Regulatory information Support links Support links Guidance 1 point calib. (ORP) 2 point calibration (pH and ISFET) Cell constant (inductive/...
Page 19
Liquiline Mobile CML18 Operation options Display structure A0044047 6 Schematic representation of the display structure Menu path/title of measuring screen Bluetooth status Battery level, charging information NAMUR indicator Measuring screen Date and time (displayed in main menu and if no sensor is connected)
Operation options Liquiline Mobile CML18 6.2.3 LED status indicator The status LED is used for the quick visualization of the sensor status. LED behavior Status Solid green Sensor working correctly Solid red No sensor connected Flashes red Sensor error Operation via SmartBlue app The SmartBlue App is available for download from the Google Play Store for Android devices and from the Apple App Store for iOS devices.
Page 21
Liquiline Mobile CML18 Operation options A0029747 8 SmartBlue App icon Bluetooth must be enabled on both devices. Enable Bluetooth → 26 A0044142 9 SmartBlue App Livelist The Livelist displays all of the devices that are within range.
Page 22
Operation options Liquiline Mobile CML18 User name >> admin Initial password >> device serial number Change the user name and password after logging in for the first time. The current measured values are displayed in the Home view. The device information (device tag, serial number, firmware version, order code) is also displayed.
Page 23
Liquiline Mobile CML18 Operation options A0048102 10 Home view of SmartBlue App with current measured values CML18 system and device information Shortcut to diagnostic list Overview of measured values of connected sensor General information and sampling option Operation is via 4 main menus:...
Page 24
Operation options Liquiline Mobile CML18 CML18_RA010905MHO A0048103 11 Main menus of the SmartBlue App Guidance Diagnostics Application System Menu Function Guidance Contains functions involving a self-contained sequence of activities, e.g. for calibration (= "Wizard", guided operation). Diagnostics Contains information on operation, diagnostics and troubleshooting, as well as configuration of the diagnostic behavior.
Liquiline Mobile CML18 Commissioning Commissioning Preliminaries Charge the device fully before initial commissioning.→ 12 Connect the sensor.→ 12 Function check WARNING Connection errors The safety of people and of the measuring point is at risk! ‣ Put the device into operation only if you can answer yes to all the following questions.
Commissioning Liquiline Mobile CML18 Setting the display language Navigate to: Display language Main menu >> System/Language >> Display language Press to scroll through the predefined values. Description of setting Configuration options Change the language of the operating menu. • Deutsch •...
Page 27
Liquiline Mobile CML18 Commissioning Press to scroll through the Device info. The following information about the device is shown on the display: • Manufacturer identification • Software version • Serial number • Designation • Extended order code 7.6.2 Adjusting the energy settings A maximum battery life of 48 h can be achieved via the energy settings.
Page 28
Commissioning Liquiline Mobile CML18 Power save w/o charger Description of setting Configuration options Set the time until the power save mode is activated if the device is running on • 1 min the battery. • 5 min • 15 min •...
Page 29
Liquiline Mobile CML18 Commissioning Description of setting Configuration options Switch signal sounds on/off • Enabled • Disabled Additional changes to the signal sounds can be made via the SmartBlue App. 7.6.4 Configuring M12 CSV Measured values can be output to other devices via the device' s M12 connection. The M12 USB data + charging cable→...
Page 30
Commissioning Liquiline Mobile CML18 7.6.7 Displaying regulatory information and approvals Navigate to: Regulatory information Main menu >> System/Language >> Regulatory information Press to display regulatory information and approvals. 7.6.8 Data logger Defining the log interval The log interval can only be changed if the data logger is deactivated.
Page 31
Liquiline Mobile CML18 Commissioning Description of setting Configuration options Enable/disable automatic data logger • On • Off Exit the menu. Once activated, the data logger automatically starts recording the measured values. If the data logger is activated, the display flashes alternately between the "Logging..."...
Operation Liquiline Mobile CML18 Operation Calibration To calibrate sensors, the calibration settings are first configured via the SmartBlue app. The calibration can then be started from the device. Configure the calibration settings in the SmartBlue app: Enable Bluetooth. → 26 Link the device to a mobile terminal via the SmartBlue app.
Page 33
Liquiline Mobile CML18 Operation 8.2.1 Saving the sample (Grab Sample) Samples can be assigned IDs and a user-definable text. By assigning an ID, samples can be assigned more easily to a measuring point, for example. IDs and the associated texts can be changed via the SmartBlue App.→ 33 In the measurement window, press ...
Page 34
Operation Liquiline Mobile CML18 Preparatory steps Install the SmartBlue app on a mobile terminal. → 20 Enable Bluetooth. → 26 Link the device to a mobile terminal via the SmartBlue app. → 20 Transferring the data Select the device in the SmartBlue app.
Page 35
Liquiline Mobile CML18 Operation A0042258 Select Data transfer. A0042261 Continue with Next. Endress+Hauser...
Page 36
Operation Liquiline Mobile CML18 A0042260 Select Data source. Select Grab sample logger for saved samples. Select Cont. data logger for data records of data logger. Press Ok to confirm. Press to discard changes and close the drop-down menu.
Page 37
Liquiline Mobile CML18 Operation A0042263 11. When the transfer is complete, press Next to continue. The result of the data transfer is displayed. A0042265 Endress+Hauser...
Page 38
Operation Liquiline Mobile CML18 12. Use Share ZIP file... to send the exported data records or save them locally. 13. Complete export by pressing Finish. Exporting to a computer Preparatory steps: Download the CML18 readout tool to the target computer and save it.
Page 39
Liquiline Mobile CML18 Operation Description of the individual elements of the export files Filename Name of the export file, based on the date/time of the first logged entry. If the sensor, sensor type or unit settings are changed, a new export file is created.
During a firmware update, do not switch off the device manually or disconnect it from the mobile terminal. A video tutorial on how to update the firmware is available on Endress+Hauser' s YouTube channel via the following link or QR code:...
Page 41
Liquiline Mobile CML18 Firmware update A0045926 13 Scan the QR code to access the instruction video Preparatory steps Download the firmware update package and save to the terminal. The current firmware update package can be found in the Downloads area on the product page at www.endress.com/CML18.
Diagnostics and troubleshooting Liquiline Mobile CML18 Diagnostics and troubleshooting 10.1 Diagnostics information via the local display 10.1.1 Accessing sensor information Navigate to: Sensor info Main menu >> Diagnostics >> Sensor info Press to access the sensor information. 10.1.2 Accessing calibration information Navigate to: Calibration info ...
Liquiline Mobile CML18 Maintenance Maintenance 11.1 Maintenance tasks 11.1.1 Cleaning ‣ Only clean with a damp cloth and commercially available cleaning agents. The device is resistant to: • Ethanol (for a short time) • Soap-based household cleaning agents • Dishwashing detergent...
(WEEE), the product is marked with the depicted symbol in order to minimize the disposal of WEEE as unsorted municipal waste. Do not dispose of products bearing this marking as unsorted municipal waste. Instead, return them to Endress+Hauser for disposal under the applicable conditions.
Liquiline Mobile CML18 Accessories 13.1 M12 USB data + charging cable Order code: 71496600 • Charging via cable • Data backup • Live data transfer A0047709 13.2 Protective cover Order code: 71530939 • Comprehensive protection • Extremely robust • Tabs and eyelets provide a range of securing options...
Technical data Liquiline Mobile CML18 Technical data 14.1 Input 14.1.1 Input power Wireless charging M12 connection 5 V; 0.6 A 14.1.2 Measured variables • pH • ORP • pH/ORP • Oxygen • Conductivity • Temperature 14.1.3 Measuring range → Documentation of the connected sensor 14.1.4...
Liquiline Mobile CML18 Technical data 14.3 Power supply 14.3.1 Supply voltage Inductive charging: use Qi-certified devices (min. 5 W output power) The power supply unit must supply an output current of at least 1500 mA. 14.3.2 Battery rated capacity 1 000 mAh (min. 950 mAh) 14.3.3...
Technical data Liquiline Mobile CML18 14.4.5 Electrical safety EN 61010-1 14.4.6 Pollution degree Complete device: Pollution level 4 Internal: Pollution level 2 14.5 Mechanical construction 14.5.1 Dimensions 54 (2.13) A0044044 14 Dimensions: mm (in) Endress+Hauser...
Page 49
Liquiline Mobile CML18 Technical data 14.5.2 Materials Components Material Housing Display window, light guide PMMA Buttons, cap M12 connection CuZn, nickel-plated 14.5.3 Materials not in contact with the medium Information according to REACH Regulation (EC) 1907/2006 Art. 33/1: The device battery contains the SVHC 1.3 propane sulton ; ethylene glycol dimethyl ether (CAS number 110-71-4) with more than 0.1% (w/w).
Need help?
Do you have a question about the Liquiline Mobile CML18 and is the answer not in the manual?
Questions and answers