LITHOSTAR MODULARIS Error List
NOTE
Info, E00 and E99
E00
Error Text:
Possible
cause/action:
E99
Error Text:
Possible
cause/action:
Water system, E10
LITHOSTAR MODULARIS circuit diagram J1048, sheets 13 and 14.
E10
Error Text:
Possible
cause/action:
E11
Error Text:
Possible
cause/action:
E12
Error Text:
Possible
cause/action:
Siemens AG
Medical Solutions
Error messages from the SIREMOBIL Iso-C have not been listed.
(Refer to the SIREMOBIL Iso-C service documents)
System was switched on.
The system was switched off without deleting the therapy data.
Press the Reset button.
–
E12
Cooling system losing water.
Is water running out of the hoses?
P
-switch o.k.?
min
Board D3.X3 connection o.k.?
Fill the cooling circuit.
Maximum pressure in the cooling circuit exceeded.
Pressure switch P
Are the hoses blocked, kinked?
P
-switch o.k.?
max
Board D3.X3 connection o.k?
Is the cooling unit operating?
Check the fill level in the cooling circuit.
Minimum pressure in the cooling circuit exceeded.
Pressure switch P
Is water running out of the hoses and hose connections?
Air in the cooling system?
Check the fill level in the cooling circuit.
SPL1-130.840.02
Rev. 03
04.05
responded.
max
responded.
min
Page 1 of 14