Page 2
Usability BEFORE USING THIS SYSTEM This system should be used only by employees adequately trained in the use of this equipment. Before using this system, the operator should be thoroughly acquainted with the instructions for use and safety recommendations provided in this manual.
Page 3
Content Overview System overview.......8 Startup ...........16 Imaging system ......26 Examination ........32 Image postprocessing .....44 Image documentation.....54 Shutdown........62 XPR8-270.622.01.01.02...
Page 4
Quick Guide Layout This Quick Guide contains pairs of opposite pag- es. As a rule, the right page shows the steps to be performed, the left page overviews, more de- tailed information, and figures. Text layout Prerequisites Instructions ...
Table of Contents System overview...........8 Mobilett Mira Max system ........9 Startup ............16 Charging the system batteries ......17 Switching on the system ........17 Transport............19 Charging the detector battery ......21 Changing the detector battery ......23 Workflow............24 Switching on the imaging system and the generator ........25 Imaging system...........26...
Page 6
Storing the detector..........63 Switching off ............65 Disconnecting the power cable ......65 XPR8-270.622.01.01.02...
Page 7
More detailed and complete descriptions can be found in the Mobilett Mira Max Operator Manual. Note The Quick Guide is only valid in conjunction with the Operator Manual and the safety information it contains.
System overview Mobilett Mira Max system Basic system (1) Arm system (2) X-ray tube assembly with collimator (3) Imaging system with touch screen (4) Transport handle (5) Chassis (6) Flat detector MAX wi-D or MAX mini Chassis (1) Touch screen with display panel...
Page 10
(8) (9) (10) (11) (11a) (11b) (11c) Note If the battery charge indicators are inactive, your system is not ready. Connect the system to the mains supply and charge the batteries. XPR8-270.622.01.01.02...
Page 11
System overview Key switch option (1) Key position OFF (2) Key position ON (3) Key position LOCK Control and display panel (1) “Generator ready” LED display (GREEN) (2) “Exposure” LED display (YELLOW) (3) LED display for operating mode A = Analog D = Digital (4) Button for toggling between analog and dig- ital mode...
Page 13
System overview Arm system with X-ray tube assembly The Mobilett Mira Max has a double articulated arm with a counterbalance. The X-ray tube as- sembly and articulated arm remain in the set po- sition without additional locks. (1) X-ray tube assembly...
Page 15
System overview Mobile MAX detectors (1) Detector MAX wi-D, for free exposures (2) Detector MAX mini, for free exposures (3) Detector indicators Detector indicators (1) Detector status LED (2) Wi-Fi status LED (3) Battery status LED (4) Infrared sensor - needed for attachment of a detector (MAXswap) (5) Power button...
Startup Note Always set the key switch to position O (key switch variant) or completely switch off with the OFF button on the numeric keypad (numeric lock variant) while the system is not being used. Note The unit can be operated by mains power or batteries.
Startup Charging the system batteries Connect the power cable to the socket. The system batteries are automatically charged when the system is connected to the mains, irre- spective of whether the system is currently switched on or not. Battery charge indicators Mains operation LED: system is connected to mains power 6 Battery charge LED bars showing the current...
Page 18
Note When transporting the unit across longer dis- tances or uneven floors, the imaging system should be switched off. (1) Handle for direction and speed control (2) Safety switch (back of handle) XPR8-270.622.01.01.02...
Startup Transport You can move the unit by means of the motor drive. For this, the Mobilett Mira Max must be switched on. Prior to transport The power cable and the LAN cable must be disconnected from the wall outlets.
Page 20
(1) Charger (2) Batteries (3) 4 charging LEDs for each battery (4) Power input with green power indicator - Green LED is illuminated when power is on. 0-25% 25-50% 50-75% 75-100% 100% Charging levels Fault signal XPR8-270.622.01.01.02...
Startup Charging the detector battery The battery of the MAX wi-D is automatically charged when the detector is properly inserted in the holder at the back of the unit, indicated by a blinking battery LED of the detector. External battery charger Batteries of MAX wi-D and MAX mini are inter- changeable and can be charged both in the ex- ternal battery charger.
Startup Changing the detector battery Let the battery run down completely. Open the locking system of the battery as shown in the figure. Remove the discharged battery. Orientate the charged battery in the correct position: battery connector (1) facing the de- vice connector (2).
2. Switch on the generator and the imaging system. 3. Register the patient: Select patient from worklist or enter patient data. 4. Align the Mobilett Mira Max. 5. Release the articulated arm and lift the X-ray tube assembly. 6. Put the detector into a sterile cover.
Startup Warm-up time from a cold start After a cold start (battery completely empty), it is possible to operate the detector, but with re- duced image quality. Therefore, avoid the bat- tery getting empty. The detector reaches good image quality again after a maximum of 120 minutes (typically 60 minutes).
Imaging system The Mobilett Mira Max is equipped with the FLUOROSPOT Compact (syngo FLC) digital imaging system. User interface layout (1) Display area for images, patient lists and dia- logs (2) Control area (3) Tabs of the task cards for mode selection...
Page 28
(1) Tabs of study lists (2) Headings line (3) Study entries (4) Scroll bar (5) Buttons (6) Image preview (Only if the study contains images or series.) (7) Status line with message (8) Number of studies being displayed XPR8-270.622.01.01.02...
Imaging system Screen saver With the screen saver, you can protect the screen from unauthorized viewing of patient data. Tap this icon on the Tools subtask card. See also Locking the system Study lists You can access the stored study data via study lists in Patient mode: ...
Page 30
Onscreen keyboard When entering of text is required, the onscreen keyboard is displayed. You can close the onscreen keyboard at any time by tapping Close. You can reopen the onscreen keyboard by tapping this icon. XPR8-270.622.01.01.02...
Imaging system Patient registration If your system is not connected to an HIS/RIS or in an emergency, you must register the patient to be examined. Emergency registration If you are not logged-in, tap Emergency in the login window. (Only necessary, if HIPAA Security is activated.) ...
Examination Preparing the wi-D detector MAX- It is possible to use the MAX wi-D and MAX mini swap detectors with the FLUOROSPOT Compact imag- ing system shared with more than one acquisi- tion system. Automatic connection In Patient mode, the available MAX wi-D and MAX mini detectors are checked.
Page 35
Examination Attaching the clip-on grid The clip-on grid is used for free exposures on the MAX wi-D. (1) Clip-on grid for MAX wi-D (2) Detector MAX wi-D Detector front must face the backside of the clip-on grid. Put the detector with the bottom in the clip- on grid as show in above figure.
Page 36
Note In analog mode, the digital flat detector is not active. In this case either film or CR cassettes must be used. XPR8-270.622.01.01.02...
Examination Starting the examination Register the patient, if necessary. Select the correct entry from the study list. Select the Examination tab. Selecting the operating mode You can work either in digital mode (D) or in analog mode (A). ...
Page 39
Examination Changing the patient view/laterality Tap this icon. Tap the correct View Position, if required Tap the correct Image Laterality, if required. Confirm with OK. Selecting the MAX detector for free exposure In case of more than one detector is connected: ...
Examination Preparing exposure Aligning the Mobilett Mira Max Release the articulated arm and lift the X-ray tube assembly. Move the arm to the required position by lat- eral rotation. Move the detector to the required position. ...
Page 42
Exposures via a two-stage manual button (1) Starting position (2) Preparation Duration: 2.5 s. Ready after four short acous- tic signals. (3) Exposure Duration: As long as the radiation indicator is lit and the signal sounds. Note The exposure button can also be pressed directly to the second pressure point (3).
Examination Releasing an exposure The imaging system indicates "Ready" with a green checkmark. Exposure phases Preparation: The green ready indicator goes out and only lights again when the rotational anode has reached its full speed (after ap- prox. 2.5 s). It is indicated via four short acoustic signals.
Image postprocessing Image display Tap the study entry in the Examined Patients list, if necessary. Tap the Examination or Postprocessing tab card. - The last acquired or processed image of the selected study is displayed. Scrolling images Tap one of these icons on the Tools or Image subtask card.
Image postprocessing Image processing Tap the Process. subtask card, if necessary. Zooming an image continuously Tap this icon. - A slider to control the zoom is displayed. Drag the slider to zoom the image. You can now pan the image. Zooming an image to 1:1 pixel display ...
Page 48
Rotate the image by 90 degrees counter-clockwise Rotate the image by 90 degrees clockwise Rotate the image by 1 degree counter- clockwise Rotate the image by 1 degree clockwise XPR8-270.622.01.01.02...
Page 49
Image postprocessing Rotating an image Tap this icon. - The available rotation functions are dis- played. Rotating the image by a fixed angle Tap the required icon once or more than once. - The image is rotated by the selected angle for each tap.
Page 51
Image postprocessing Windowing images Tap this icon. - The sliders for window width / contrast and for window center / brightness are dis- played. Tap and move your finger within the image. - Upward: Increase window center/bright- ness values. - Downward: Decrease window center/ brightness values.
Image postprocessing Image annotation Tap the Graphics subtask card, if necessary. Adding a laterality label Tap this icon on the Graphics subtask card. - The Label L/R window with the available labels and label positions is displayed. Tap the required label L or R.
Image documentation Image overview Tap the study entry in the Examined Patients list, if necessary. Tap the Documentation tab. - The image catalog (overview page) of the study is displayed. Displaying an image in full size Tap the required image. ...
Page 56
Note It is possible to configure hosts where images are automatically sent to. The image transfer or- der is also configurable. Note The archive node is indicated by an asterisk * . Note During transfer, an icon displayed bottom right, indicates the running transfer process.
Image documentation Transferring images (DICOM Send) Tap the Send subtask card, if necessary. Mark the images to be transferred. Sending to default destination Tap this icon. The images are sent to the configured default target(s) in the background. Sending to other destinations ...
Page 58
Note It is possible to configure a hardcopy destination where images are automatically sent to. The im- age transfer order is also configurable. Note During transfer, an icon displayed bottom right, indicates the running transfer process. XPR8-270.622.01.01.02...
Image documentation Filming/printing images (DICOM Print) Tap the Film subtask card, if necessary. Mark the images in the order you want to ex- pose them. Sending to default hardcopy camera/printer Select the number of Copies to be exposed, if necessary.
Image documentation Deleting images Tap the Select subtask card, if necessary. Tap this icon. Tap the images you want to delete one after the other. - The images are marked with a white cross. It you want to delete the images marked with a cross: ...
Shutdown Locking the system The system can be intermediately locked to se- cure it from unauthorized access to data, driving and X-ray, without having to shut it down, so that the user can continue to work when return- ing to the unit. Key lock variant: ...
Page 64
Note To avoid delays due to flat batteries, we recom- mend that you switch off the Mobilett Mira Max completely when it is not in use and connect it to the mains supply to charge the batteries. XPR8-270.622.01.01.02...
Page 65
Shutdown Switching off Switching off will shut down the complete sys- tem, i.e. the imaging system, the generator, and driving. Power off with key switch variant Turn the key switch to position OFF. Power of with keypad variant Press the OFF button on the numeric keypad on the control and display panel.
Page 66
Index 1:1 pixel display Emergency registration Exam Manager Examination starting Exporting data Annotation Exposure Arm system preparing Automatic windowing releasing External battery charger Basic modes Battery charge indicators Filming images Flip indicator Flipping Changing an image detector battery FLUOROSPOT Compact Charging detector battery system batteries...
Page 67
15, 33 MAX wi-D detector storing Storage of modifications MAXswap Store Monitor Mirroring Storing an image the detector Mobilett Mira Max Study lists system Switching off Switching on imaging system and generator Onscreen keyboard the system Operating mode syngo FLC...
Page 68
System batteries charging System overview Task cards Text in the image Transferring images Transport Unlocking the system 45, 55 Unmarking images User interface layout Warm-up time Windowing Wireless detector status Wireless network WLAN Workflow X-ray tube assembly Zooming XPR8-270.622.01.01.02...
Page 71
All names shown in figures are purely fictional. Similarities to existing people are entirely coincidental. Siemens Healthcare GmbH reserves the right to change the im- plementation and specifications of this document without pri- or notice. For up-to-date information, please contact your local Siemens representative.