5.5 Drying
1) Wipe all parts of the needle guide bracket with a single-use,
fluff-free wipe or towel to remove moisture from the surface of them.
2) Dry all parts of the needle guide bracket naturally in an ambient
temperature between 15-30°C for a minimum of 4 hours. Alternatively
they can be dried using a drying heater at a temperature of less
than 55°C.

5.6 Packaging

Packing the device into the sterile bag is recommended until using it
for keeping cleaned and disinfected condition.

5.7 Assembly and Inspection

Assemble the needle guide bracket and inspect it for any damage such
as crack, scratch or deformation. Do not use it if any damage is found.

5.8 Storage

Store the equipment in a cool, dustproof and dark space to avoid
high temperature, humidity and direct sunlight.
Do not perform any sterilization procedure to the needle guide bracket.
The needle guide bracket is not applied to any sterilization procedure.
If sterilization procedure is performed to it, it may suffer serious
damage and become unusable.