3.3.1
Attaching the Magnetic Sensor (EZU-RV2S)
The procedure of attaching the Magnetic sensor is as follows.
1) Attach the Spacer for EZU-RV2S to the Magnetic sensor.
2) Attach the Magnetic sensor into the attachment with the correct
direction as shown in Fig. 15.
3) Insert the hold of the attachment into the projected part of the
probe.
Magnetic sensor
attachment
Magnetic sensor
(EZU-RV2S: Black)
Fig. 15
1) Never attach the Magnetic sensor attachment to the probe in the
incorrect direction, otherwise it may result in false diagnosis.
2) Never forget to attach the Spacer for EZU-RV2S to the Magnetic sensor,
otherwise it may result in false diagnosis.
Spacer for EZU-RV2S
Projected part of the
EZU-RV2S (Place the projected
part upward as shown here)
How to attach the Magnetic Sensor Attachment
CAUTION