Hp-Ib Check, Example Program 1; Program Comments - HP 83751A Programming Manual

Synthesized sweeper
Hide thumbs Also See for 83751A:
Table of Contents

Advertisement

HP-IB Check, Example Program 1

This lkst
functional. Connect a controller to the sweeper via an HP-IB cable. Clear and
reset the controller and type in the following program:
Run the program and verify that the REMOTE LED is lit on the sweeper. If
it is not, verify that the sweeper address is set to 19 and that the interface
cable is properly connected.
If the controller display indicates an error message, it is possible that the
program was entered in incorrectly. If the controller accepts the REMOTE
statement but the sweeper REMOTE LED does not tum on, perform the
operational checks as outlined in the respective Operating and Service
Manuals to kd the defective device.
1 0 :

Program Comments

20:
30:
40:
50:
60:
70:
program
is
to
verify that the HP-I5 connections
10
20
ABORT 7
30
LOCAL Source
CLEAR Source
40
50
REMOTE Source
60
70
PRINT "The source should now be in REMOTE. "
80
PRINT "Verify that the 'REMOTE' LED is on. 'I
90
END
Setup a variable to contain the HP-IB address of the source.
Abort any bus activity and return the HP-IB interfaces to their reset
states.
Place the source into LOCAL to cancel any Local Lockouts that may
have been setup.
Reset the source's parser and clear any pending output from the
source. Prepare the source to receive new commands.
Place the source into REMOTE.
Clear the display of the computer.
Print a message to the computer's display.
Getting Started Programming
Programming Typical Measurements
and interface are
1-51

Advertisement

Table of Contents
loading

This manual is also suitable for:

83751b83752a83752b

Table of Contents