Download Print this page
ABB MagMaster MF/F Upgrade

ABB MagMaster MF/F Upgrade

Electromagnetic flowmeter

Advertisement

Quick Links

SP033 Adam Street Wynnum SPS - WaterMaster - Electromagnetic Flowmeter -
User Guide IM/MM2WM–EN Rev. A
MagMaster
Electromagnetic flowmeter
MagMaster to WaterMaster
electronics upgrade
Introduction
These instructions describe how to upgrade an existing
MagMaster MF/F flowmeter (DN80 to DN1600)
installation with remote transmitter for use with the new
remote WaterMaster transmitter.
This guide should be read in conjunction with the
following publications:
– WaterMaster FET100 User Guide, OI/FET100–EN
– WaterMaster Programming Guide, IM/WMP
Q-Pulse Id: VM486
OIML R49 Class 1 450mm - User Guide
Active: 08/04/2016
Page 1 of 12

Advertisement

loading

Summary of Contents for ABB MagMaster MF/F

  • Page 1 MagMaster to WaterMaster electronics upgrade Introduction These instructions describe how to upgrade an existing MagMaster MF/F flowmeter (DN80 to DN1600) installation with remote transmitter for use with the new remote WaterMaster transmitter. This guide should be read in conjunction with the following publications: –...
  • Page 2 We are an established world force in the design and manufacture of instrumentation for industrial process control, flow measurement, gas and liquid analysis and environmental applications. As a part of ABB, a world leader in process automation technology, we offer customers application expertise, service and support worldwide.
  • Page 3: Table Of Contents

    SP033 Adam Street Wynnum SPS - WaterMaster - Electromagnetic Flowmeter - OIML R49 Class 1 450mm - User Guide MagMaster Electromagnetic flowmeter Contents Contents Electrical Installation ........................2 Configuration DIP Switches ..................... 2 MagMaster Sensor Cable ........................ 3 MagMaster Sensor Cable Wiring to WaterMaster Transmitter ..........3 2.1.1 Sensor Coil Resistance ....................
  • Page 4: Electrical Installation

    SP033 Adam Street Wynnum SPS - WaterMaster - Electromagnetic Flowmeter - OIML R49 Class 1 450mm - User Guide MagMaster Electromagnetic flowmeter 1 Electrical Installation 1 Electrical Installation Note. These upgrade instructions and the use of WaterMaster with existing MagMaster sensors is supported only on the MF/F product code style of sensors, not the MF/E type.
  • Page 5: Magmaster Sensor Cable

    M1 and M2 with the meter turned off. Note. An alarm on the WaterMaster display indicates if the coil resistance is too low for the WaterMaster transmitter to operate. For the MF/F flowmeter sensor this alarm should not occur. Contact ABB Service if it does.
  • Page 6: Programming & Configuration

    Exit The 'Sensor Setup' programming menu enables the WaterMaster electronics to be 'Retro-Fitted' to MagMaster MF/F sensors. For retrofit applications an extra menu (Sensor Setup) appears at the end of the menu tables; it is not present on factory-calibrated WaterMaster flowmeters. Note that after completing configuration of the retrofit installation, this menu disappears.
  • Page 7 SP033 Adam Street Wynnum SPS - WaterMaster - Electromagnetic Flowmeter - OIML R49 Class 1 450mm - User Guide MagMaster Electromagnetic flowmeter 3 Programming & Configuration [Sensor Setup] continued [Sensor Size (Bore)] Nominal bore size of flow sensor in mm: [DN1]* [DN65]* [DN760]...
  • Page 8 SP033 Adam Street Wynnum SPS - WaterMaster - Electromagnetic Flowmeter - OIML R49 Class 1 450mm - User Guide MagMaster Electromagnetic flowmeter 3 Programming & Configuration MF/F DN Cal Fact 1 to Ss Cal Fact 2 to Sz Multiplier Multiplier 102.3 102.0 101.9...
  • Page 9 SP033 Adam Street Wynnum SPS - WaterMaster - Electromagnetic Flowmeter - OIML R49 Class 1 450mm - User Guide MagMaster Electromagnetic flowmeter 3 Programming & Configuration [Sensor Setup] continued [Trim St] Calibration 'trim' adjustment in % applied to the main span [Span Ss]. Is set to 0.000%.
  • Page 10 Caution. Changing this setting to [Calibrated] causes the 'Sensor Setup' menu to disappear. This sensor calibration data (type, size, span, zero, etc.) is then LOCKED. Modification is then possible only by ABB Service engineers. Only when programming of calibration settings is complete, fully checked and the flowmeter is working correctly adjust this setting to 'Calibrated'.
  • Page 11 Repair Centre. — Metals and Minerals — Oil, Gas & Petrochemical — Pulp and Paper ABB Limited Tel: +44 (0)1453 826661 Drives and Motors Fax: +44 (0)1453 829671 — AC and 6 Drives, AC and DC Machines, AC Motors to —...
  • Page 12 With regard Stonehouse to purchase orders, the agreed particulars Gloucestershire GL10 3TA shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this Tel: +44 1453 826 661 document.