Intraoperative probe
Table of Contents

Advertisement

Keep the acoustic power low and minimize the ultrasound
exposure time for the examination of an early pregnancy.
Do not expose the connector to water or other liquids. The
connector is not waterproof.
Do not hit or drop the probe. The probe is easily damaged
by mechanical shock.
Do not use detergents and disinfectants other than listed in "8.2
Supplier's list of the Probe" and "8.3 Supplier's List of the
Magnetic sensor attachment and the Spacer for EZU-RV2S".
Use a sterile probe cover to avoid staining or damaging the
acoustic lens.
Clean, disinfect and sterilize the probe, Magnetic sensor
attachment and Spacer for EZU-RV2S before the first use as
they are not sterilized in the factory.
Use only the soft cloth or tissue to clean the acoustic lens.
Do not use this probe with other equipment except for those
specifically approved in the manual. Use with unapproved
equipment can result in an electric shock, burn, or other
injury to the patient or operator and damage to the probe
and the other equipment.
Do not wash, disinfect or sterilize using procedures other
than those specified in this manual. Infection could result
due to incomplete washing disinfection or sterilization. It
can also result in damage to the probe or reduced performance.
CAUTION
-23-
Q1E-EP1234