6) After the use of the probe, clean and disinfect and sterilize
the probe immediately and using RVS, also clean and disinfect
and sterilize the Magnetic sensor attachment. In case of using
RVS with an EZU-RV2S, also clean and disinfect and sterilize
the Spacer for EZU-RV2S.
7) Store the probe, the Magnetic sensor attachment and the Spacer
for EZU-RV2S in the environment indicated in "6. Maintenance
and Safety Inspection".
 Do not use the probe if the image and the frequency do not match
the probe. An incorrect acoustic output can result in burns or
other injuries to the patient.
 Scan for the minimum length of time necessary for the diagnosis
and at the lowest possible output. There is the possibility that
the patient's tissues could be affected. For details about the
acoustic output, please refer to the operation manual of the
ultrasound diagnostic instrument.
 The acoustic lens of the probe is very thin and delicate. So
if the surface of the acoustic lens is wiped for removing
remaining ultrasound jelly or cleaned, use the soft cloth or
tissue paper and handle with care.