Instructions for use COMPACTtorque 636 - 0.553.0810 COMPACTtorque 636C - 0.553.0830 1 User instructions 1 User instructions Dear User Congratulations on purchasing this KaVo quality product. By following the instruc‐ tions below you will be able to work smoothly, economically and safely.
Instructions for use COMPACTtorque 636 - 0.553.0810 COMPACTtorque 636C - 0.553.0830 2 Safety | 2.1 Description of safety instructions 2 Safety 2.1 Description of safety instructions Warning symbol Structure DANGER The introduction describes the type and source of the hazard.
Page 6
This service interval depends on the frequency of use and should be adjusted accordingly. Service may only be carried out by KaVo-trained repair shops using original KaVo replacement parts. 4/22...
350,000 to 400,000 rpm Recommended operating pressure 2 to 3 N (0.45 to 0.68 lbs) COMPACTtorque 636 turbines can be attached to all hoses with Borden 2-hole connections. COMPACTtorque 636 turbines can be attached to all hoses with 4-hole connec‐...
Instructions for use COMPACTtorque 636 - 0.553.0810 COMPACTtorque 636C - 0.553.0830 4 Start up and shut down | 4.1 Mounting the permanent connection 4 Start up and shut down WARNING Hazard from non-sterile products. Infection hazard for care provider and patient.
Instructions for use COMPACTtorque 636 - 0.553.0810 COMPACTtorque 636C - 0.553.0830 4 Start up and shut down | 4.2 Check the amount of water 4.2 Check the amount of water CAUTION Overheating of the tooth due to lack of cooling water.
Instructions for use COMPACTtorque 636 - 0.553.0810 COMPACTtorque 636C - 0.553.0830 4 Start up and shut down | 4.4 Check the seal 4.4 Check the seal CAUTION Missing or damaged seal When the seal is damaged, it can cause malfunctions or premature failure.
Instructions for use COMPACTtorque 636 - 0.553.0810 COMPACTtorque 636C - 0.553.0830 5 Operation | 5.1 Attach the medical device 5 Operation Note At the beginning of each workday, the water-conducting systems should be rinsed for at least 2 min. without the instrument being attached; if there is a risk of contamination from reflux or back suction, the system must be rinsed for 20-30 seconds.
Instructions for use COMPACTtorque 636 - 0.553.0810 COMPACTtorque 636C - 0.553.0830 5 Operation | 5.4 Removing the milling tool or diamond grinder CAUTION Hazard from defective chucking system. The tool can fall out and cause injury. ▶ Pull on the tool to check if the chucking system is functioning properly and that the tool is firmly clamped.
Instructions for use COMPACTtorque 636 - 0.553.0810 COMPACTtorque 636C - 0.553.0830 6 Troubleshooting | 6.1 Cleaning the spray nozzle. 6 Troubleshooting 6.1 Cleaning the spray nozzle. CAUTION Hazard from insufficient amount of spray water. Insufficient spray water can cause the medical device to overheat and damage the tooth.
Instructions for use COMPACTtorque 636 - 0.553.0810 COMPACTtorque 636C - 0.553.0830 7 Reprocessing methods according to ISO 17664 | 7.1 Preparation at the site of use 7 Reprocessing methods according to ISO 17664 7.1 Preparation at the site of use WARNING Hazard from nonsterile products.
7.2.3 Manually cleaning the interior Can only be done with KaVo CLEANspray or KaVo DRYspray. ▶ Cover the medical device with the KaVo CLEANpac bag, and place it on the corresponding care adapter. Press the spray button three times for 2 seconds each time.
The efficacy of manual internal disinfection must be demonstrated by the manu‐ facturer of the disinfection agent. With KaVo products, use only disinfection agents that have been released by KaVo with respect to the compatibility of mate‐ rials (e.g. WL-cid / made by ALPRO).
Reduced product life. ▶ Perform proper care regularly! Note KaVo only guarantees that its products will function properly when the care prod‐ ucts used are those listed as accessories, as they were tested for proper use on our products. 7.5.1 Servicing with KAVOspray COMPACTtorque 636 ▶...
▶ Carry out the servicing according to the instructions in the section, "Care with KaVo Spray". 7.5.3 Servicing with KaVo SPRAYrotor COMPACTtorque 636 KaVo recommends servicing the product after each time it is used, i.e. after each automatic cleaning and before each sterilisation. ▶ Unscrew the COMPACTtorque turbine from the hose.
Can be used with the COMPACTtorque 636C only. The COMPACTtorque 636 cannot be fitted in the QUATTROcare! KaVo recommends servicing the product after each time it is used, i.e. after each automatic cleaning and before each sterilisation. ▶ Remove the cutter or grinder.
Instructions for use COMPACTtorque 636 - 0.553.0810 COMPACTtorque 636C - 0.553.0830 7 Reprocessing methods according to ISO 17664 | 7.6 Packaging Servicing the clamping chuck KaVo recommends cleaning and servicing the chuck system once a week. See also: Instructions for Use KaVo QUATTROcare PLUS 2124 A...
▶ Immediately remove the product from the steam steriliser after the sterilisa‐ tion cycle! The KaVo medical device has a maximum temperature resistance up to 138 ℃ (280.4 °F). Select a suitable procedure (depending on the available autoclave) from the fol‐...
Instructions for use COMPACTtorque 636 - 0.553.0810 COMPACTtorque 636C - 0.553.0830 8 Tools and consumables 8 Tools and consumables Available from dental suppliers. Material summary Mat. No. Spare turbine with wrench 0.553.8941 Spare turbine without wrench 0.553.8921 Cleanpac 10 units 0.411.9691...
12 months from the date of the invoice, subject to the following conditions: In case of justified complaints, KaVo will honour its warranty with a free replace‐ ment or repair. Other claims of any nature whatsoever, in particular with respect to compensation, are excluded.