Page 2
Indications for Use/Intended Use Symbols on the product The LCD Monitor is intended to provide 4K3D and 2D color video displays of surgical and other Consult the instructions for use medical images. Follow the directions in the instructions for use for parts of the unit on which this Notes symbol appears.
Page 3
IEC 60601-1 standards. If expense. in doubt, consult qualified Sony service You are cautioned that any changes or personnel. modifications not expressly approved in this manual could void your authority to operate this 3.
Page 4
Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of the LMD-XH550MT. Otherwise, degradation of the performance of this equipment could result. If the LMD-XH550MT will be used adjacent to or stacked with other equipment, normal operation of the LMD-XH550MT under such configurations should be verified via observation.
Page 5
Guidance and manufacturers declaration electromagnetic immunity The LMD-XH550MT is intended for use in the electromagnetic environment specified below. The customer or the user of the LMD-XH550MT should assure that it is used in such an environment. IEC 60601 Compliance Immunity test Electromagnetic environment ...
Page 6
Guidance and manufacturers declaration electromagnetic immunity The LMD-XH550MT is intended for use in the electromagnetic environment specified below. The customer or the user of the LMD-XH550MT should assure that it is used in such an environment. IEC 60601 Compliance Immunity test Electromagnetic environment ...
Page 7
If the measured field strength in the location in which the LMD-XH550MT is used exceeds the applicable RF compliance level above, the LMD-XH550MT should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the LMD- XH550MT.
Page 8
The LMD-XH550MT is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of the LMD-XH550MT. Otherwise, degradation of the performance of this equipment could result.
Page 9
Top side: 30 cm (11 in.) or more Warning on power connections for medical use Consult with Sony qualified personnel for the Customers in the U.S.A. and Canada should use following types of installation location. the following type of power cord.
Page 10
Your co-operation shall facilitate of Sonys limited warranty applicable to this proper disposal & help prevent potential negative product. consequences/hazards to the environment and...
Initial Setup menu ........33 Table of Contents Preset menu .........34 Troubleshooting ..........34 Error Messages ..........35 Precaution ............12 Specifications ..........35 On Safety ...........12 Dimensions ............41 On Installation ...........12 3D Viewing Angle (vertical) ......42 Safety precautions for using this unit ..12 Licenses ............
When connecting the unit to medical Precaution equipment, refer to “Precautions when connecting this unit to medical equipment”. On Safety Precautions when connecting this unit to medical equipment Operate the unit only with a power source as specified in the “Specifications” section. Before using this unit for medical purposes, be ...
update of equipment connected to the IT- addition, over a long period of use, because of NETWORK; and the physical characteristics of the liquid crystal upgrade of equipment connected to the IT- display, such “stuck” pixels may appear NETWORK.
The fan for cooling the unit is built in. When the still images are displayed in the same position on fan error indication appears on the screen, turn the screen continuously, or repeatedly over off the power and contact an authorized Sony extended periods. dealer. Images that may cause burn-in...
Do not keep the protection plate surface/ monitor surface in contact with a rubber or 1) A.I.M.E. is a registered trademark of Sony vinyl resin product for a long period of time. Corporation. The finish of the surface may deteriorate or the coating may come off.
Page 16
This unit is a high definition and high performance multi-format LCD monitor compatible with the micro-polarizer 3D. Exclusive 3D glasses optimized for this unit are supplied considering long-time work. You can use it for both 2D and 3D with the 2D/3D selection function.
Location and Function of Parts and Controls Front Panel Power indicator When you press PORT A or PORT B when it is lit in white, an image from the input Indicator Operating State connector assigned to the PORT A/B is displayed.
Page 18
CUSTOM buttons Turns on or off the assigned function. You can adjust the assigned function by pressing the / buttons. (Refer to the Custom Button of the System Configuration menu on page 33 and of the Initial Setup menu on page 34.) The following functions are assigned in the default setting.
Rear Panel To use the connectors on the rear panel, remove the cable cover. For details on the cable cover, see page 23. Cable cover (L) cover to the right. The cable of the 12G-SDI 1 input/output connector can be connected or When connecting or disconnecting the cable of the connector on the rear panel, remove this disconnected without removing the cable cover...
Page 21
Active through output connector for serial digital component signals (4K/HD). SDI 1 and Note SDI 2 outputs are available. Consult with Sony qualified personnel at using 3G-SDI input/output connector (BNC type) this connector. input connector Caution Input connector for serial digital component For safety, do not connect the connector to signals (HD/SD).
Page 22
Be sure to connect equipment or cables 12V 2.5A (DC output) connector specified by Instructions for Use of this monitor Outputs 12 volts for connected external to the 12G-SDI/3G-SDI output connectors. If equipment. you connect unspecified equipment or cables, ...
Preparation Connecting AC input connector Make sure that the (On)/ (Standby) switch is set to (Standby). AC power cord Remove the cable cover (L). Slide down the cable cover (L) while sliding Plug holder for the the slide locks (2 pcs), which are located at supplied AC power cord the bottom of the monitor, in the direction of the arrow.
Attach the cable cover (L). Turning on the Monitor/Switching Input Slide the cable cover (L), while sliding the Settings slide locks (2 pcs), which are located at the bottom of the monitor, in the reverse Connect the plug of the AC power cord to an direction of step2.
Initial Setting If you have selected the wrong area, change the Color Temperature setting (page 28). When you turn on the unit for the first time after purchasing it, select the area and language To set the using language where you intend to use this unit. You can select one of seven languages (English, Chinese, Japanese, Italian, Spanish, German, and To set the using area...
Using the Menu The unit is equipped with an on-screen menu for making various adjustments and settings such as picture control, input setting, setting change, etc. Press the / button to select the tab. The selected tab is shown in blue, and setting items of the selected tab appear.
About the control navigation Adjustment Using the Menus Depending on the state, the operation buttons of the unit light as shown below: White light: Operable state. Green light: Operating state. Items Off: Unable to operate. The screen menu of this monitor consists of the following items.
Pattern Skip Menu Setting Gamma Select the appropriate gamma mode Input/Output Configuration (page 30) from “1.8,” “2.0,” “2.2,” “2.4,” “2.6,” “DICOM,” “Highlight,” “HLG,” or “Auto.” Clone Out “DICOM” is for reference, not for Port A Input Select diagnostic purposes. Select “HLG” when the input signal is Port B Input Select HDR-HLG.
To display the corresponding signal with “Enhanced Format,” use a Screen Control menu Premium High-Speed HDMI cable The Screen Control menu is used to set the image within a length of 3 meters (Sony product recommended). display setting for each input. Menu Setting...
Menu Setting Menu Setting 3D Signal Format Select the 3D signal format. Port C Input Sets the input connector for the third Single Cable DS: screen. Select to display signals of the 3G Select from among “12G-SDI 1,” “12G- Level-B Dual Stream system in 3D.
Menu Setting Menu Setting Color Space Select the color gamut from “Auto,” Input Name Set the name of the 12G-SDI 1, 12G-SDI “BT.709,” or “BT.2020.” 2, 3G-SDI, Display Port, HDMI, DVI-D When “Auto” is selected, the color input connector. gamut changes according to the ...
Page 32
Menu Setting Menu Setting A.I.M.E. Display Input Connector Name Always displayed while the A.I.M.E. function is activated. Select Port Off: Control Lock Not displayed. Language You can select the menu or message language from the following Signal Format languages.
Menu Setting Menu Setting Remote Backlight Control Off: Turns off backlight control. Remote Mode Sets the remote function when the Standard: REMOTE connector is connected to the Sets the backlight to standard optional Foot Switch FS-24. brightness mode. ...
Menu Setting Troubleshooting Port B Input Select Sets the input connector that is skipped when changing the input signal by pressing the PORT B button. Set to “Not Skip” or “Skip” for the 12G- This section may help you isolate the cause of a SDI 1, 12G-SDI 2, 3G-SDI, Display Port, problem and as a result, eliminate the need to HDMI, DVI-D connector.
Error Messages Specifications Picture performance When the following messages appear on the screen, turn off the power and contact an LCD panel TFT Active Matrix authorized Sony dealer. Pixel efficiency 99.99% Messages Description Viewing angle (panel specification) Fan Error A malfunction occurs in the fan.
Page 36
Optional accessories WARRANTY, OR FOR ANY OTHER REASON 3D Eye Shield Kit WHATSOEVER. CFV-E30SK SONY WILL NOT BE LIABLE FOR CLAIMS OF 2D Eye Shield Kit ANY KIND MADE BY USERS OF THIS UNIT OR CFV-E20SK MADE BY THIRD PARTIES.
Page 37
Pin assignment 12V 2.5A (DC output) connector SERIAL REMOTE (RS-232C) connector D-sub 9-pin, female Pin number Signal Pin number Signal 12 V REMOTE (stereo mini jack) connector Pin number Signal On (short to sleeve) Off (open) 5V 2.0A (DC output) connector Pin number Signal...
Page 38
Available signal formats The unit is compatible with the signal systems shown below: Signal format 12G-SDI 1 12G-SDI 2 3G-SDI SDI Single Link SD-SDI 720×487/59.94i 4 : 2 : 2 YCbCr 10bit × × 720×576/50i 4 : 2 : 2 YCbCr 10bit ×...
Warning Dimensions Be sure to set the tightening torque value to the following value. Torque value: 1.7±0.1 N·m Front Make sure the tightening torque value is at this value. If the torque value is not appropriate, the mounting part may become damaged or the screws may become loosened, and in the worst case, it may cause injury or damage to the unit due to the unit falling.
3D Viewing Angle (vertical) Licenses For details on the license, refer to “Software License Information” on the CD-ROM. Side View 3D Viewing Angle (vertical) Crosstalk ratio ≤ 7% A (Typical) B (Typical) C (Typical) 1 000 mm 37° 2 000 mm...
Need help?
Do you have a question about the LMD-XH550MT and is the answer not in the manual?
Questions and answers