Installation

Installing the new filter assembly

See figure 2.5 and 3.1.
1.
2.
3.
NOTICE
Fig. 3.1 Position sensor and drive motor of the filter assembly

Reinstalling the diaphragm-plate assembly

See figure 2.2 and 2.3.
NOTICE
1.
2.
3.
Siemens-Elema AB
Solna, Sweden
Position the new filter assembly so that the cog disc engages with the motor
pinion (12) and the rear guide. The angular position of the filter assembly in
relation to the motor pinion is arbitrary.
Install the two front guides (10).
Install the bearing (9) and secure with the retaining ring (8).
Do not touch the filters with bare fingers. Use lint-free cotton
gloves. Be careful not to damage the optical switch position
sensor (11) during the installation.
Be careful with the filter assembly.
Fit the diaphragm-plate assembly (5) to the diaphragm-plate holder, while
carefully putting the lamp holder into position, and fasten with the four Allen
screws (6).
Fit the lamp holder (4) to the diaphragm-plate holder and fasten with the two Allen
screws (3).
Replace the cable ties that were cut during the removal of the diaphragm-plate
assembly.
Register 5