shape with the internal phase 0 °.
Dip-Swell synchronization sets the synchronization of the Dip/Swell
Synchronization can be selected from the list:
•
Sync Off
•
Sync On
If Sync Off is selected, Dip/Swell generation starts immediately after
phase of 0 °.
trigger. If Sync On is selected, Dip/Swell is synchronized with the internal
•
Dip-Swell ext. trigger enables external triggering in the Power
Dip/Swell mode.
External-trigger can be selected from the list:
• Input Off
• Input IN2
If Input Off is selected, Dip/Swell generation can be started only internally
(Trigger softkey or remote control). If Input IN2 is selected, Dip/Swell
the IN2 input. The input must remain low for 10 µs after the falling edge to
generation can be started internally or by the falling edge being applied to
be recognized properly.
Dip-swell features are only available in the 6003A/PQ Power Quality
Option.
11. The Voltage From Current submenu contains parameters that affect the
Voltage From Current function. This function is available only in Pac and
Pdc Extended modes.
•
State - The current output in Pac and Pdc Extended modes can be
configured to output a voltage by setting the "Enable" state.
•
The Equivalence Factor defines the ratio between simulated current
value and generated voltage. This equivalence factor is used for
converting the voltage output to an equivalent current. The factor is
expressed in V/A and has a range of 0.000002 V/A to 10 V/A.
Meter Menu
The Meter submenu contains parameters that can be measured by the Product.
Function sets the internal multimeter measure function.
To select measurements, choose an item from the list:
•
Voltage (dc voltage 0-12 V)
•
Current (dc current 0-25 mA)
•
Frequency (Frequency 1 Hz-15 kHz)
Current is measured across the mA and COM terminals. Voltage or frequency is
measured across the V and COM terminals.
Note
Front-Panel Operation
Setup Menu (Main Menu)
4
4-9