1. Apply input power to 5300 ac receptacle.
2. Turn counter on with 5300 SAMPLE RATE control.
3. Set RANGE switch to CHK position. Display is a
Resolution
Selector
Display
10 kHz
0010.00
1 kHz
010.000
100 Hz
10.0000
10 Hz
0.00000
1 Hz
●000.000
.1 Hz
●00.0000
●Indicates overflow light is on.
Figure 9E-3-3. Self-Check Measurements
Adjust SAMPLE RATE for minimum display time
(full ccw).
function of RESOLUTION switch.
5300A Self Check
Annunciator
M, Hz
M, Hz
M, Hz, C
M, Hz, C
K, Hz, C
K, Hz, C
5300B Self Check
Display
Annunciator
000010.00±1
00010.000±1
0010.0000±1
010.00000±1
10000.000±1
0000.0000±1
Model 5305B
Operation
M, Hz
M, Hz
M, Hz, C
M, Hz, C
K, Hz, C
K, Hz, C
9E-3-3