For the following models: 8104 8105 8106 SAFETYCHECKLIST KEEP unqualified/unauthorized personnel away from test area ARRANGE test stations in a safe and orderly manner NEVER touch products or connections during a test TURN OFF high voltage first in the event of a problem NEVER perform a test on energized circuitry or equipment BE SURE to always connect the return test lead first HANDLE test clips by insulation only, never touch clips directly...
FRONTPANELCONTROLS GRAPHIC LCD : 320 X 240 monographic LCD. SOFT KEYS: Multi-function parameter selection keys used to select screen and change parameters. NUMERIC DATA ENTRY : Keys used to enter numeric data. HIGH VOLTAGE ON INDICATOR: Flashing indicator to warn the operator high voltage is present.
BACKPANELCONTROLS 8104 BACK PANEL CONTROLS HIGH VOLTAGE OUTPUT TERMINAL : Connector used to attach the high voltage test lead, adapter box high voltage lead, or test fixture high voltage lead to the instrument. The connector is recessed for safety. CALIBRATION BUTTON : Used to put the instrument into calibration mode. SENSE (+) TERMINAL : Connector used to attach the current test lead or test fixture to the instrument.
Page 4
RETURN OUTPUT TERMINAL : Connector used to attach the return test lead or test fixture to the instrument. This connector provides the return current path for the high voltage, ground bond and continuity current. INPUT POWER RECEPTACLE : Standard IEC 320 connector for connection to a standard NEMA style line power (mains) cord.
INSTRUMENTSETUP WARNING: LOCATE A SUITABLE TESTING AREA WITH A THREE- PRONG, GROUNDED OUTLET. BE SURE THAT YOUR THREE-PRONG OUTLET HAS BEEN TESTED FOR PROPER WIRING. ALSO, MAKE SURE YOU READ THE SAFETY CHECKLIST OF THIS GUIDE BEFORE STARTING TO TEST. 1.
Page 6
CHANGINGTHETYPEOFTEST ADDING NEW TESTS From the Main Menu screen, press the “Setup Tests” soft key. On the Setup Tests screen, the test and step number will be highlighted. Press the “Add” soft key and specify the type of test with the soft key options given (ACW, DCW, IR, Continuity for 8104 and Run Test or Line Leakage Test for 8105 and 8106).
Page 7
Use the arrow keys (up, down) to highlight the test to be edited and hit the “Edit” soft key. Use the soft keys to edit parameters on the right side of the screen. Use the arrows (up, down, left, right) to scroll between test parameters on the left side of the screen. Use the number keypad to change a parameter value.
PERFORMINGATEST TEST CONNECTIONS (8104): The adapter box is an easy way to connect a Device Under Test (DUT) that terminates in either a two or three-prong line cord. The following diagram shows how to connect the adapter box to the instrument and the DUT.(See Figure 4) If you are testing a Class I product, you may elect to perform a ground bond test on the ground circuit of the DUT along with the hipot test.
Page 9
PERFORMINGATEST TEST CONNECTIONS (8105 & 8106): Under the DUT output ports on the back panel, plug the line lead from the adapter box into the L terminal and the neutral lead into the N terminal. Next attach the red ground lead from the adapter box to the GND terminal.
Page 10
PERFORMINGATEST TEST CONNECTIONS There is no ground circuit on the DUT of a Class II product. Therefore, test leads may be utilized to connect the OMNIA to a DUT. (See Figure 6) Plug the black return lead into the Return terminal and the black Kelvin lead into the Sense (–) lead.
TEST RESULTS VIEWING TEST RESULTS From the Perform Tests screen, press the “Results” soft key. (See Figure 7) Use the up and down arrows to select the test to be reviewed or use the left and right arrows to page up and page down. Results are displayed on the right hand side of the screen including the test type, reason for failure and test parameter results.
Page 12
For more support materials and testing information on OMNIA please visit us online at www.asresearch.com.You can also refer to the Operation and Service Manual. For additional information about these and other key features of the OMNIA, please consult the full Operation and Service Manual or call us toll-free 1-800-858-TEST (8378) or +1-847-367-4077...