PMF Control LED (brake disabled)
PCB 92-074
LED1
LED2
LED3
LED5
LED6
PMF Brake disabling for heavy use
If upon log file inspection high duty comes in evidence it' recommended to disable anode rotation
brake to reduce operative tube assembly temperature.
For more details refer to service instructions SPB7-115.840.01....
MAMMOMAT Balance
Maintenance Instructions
(yellow) bright during normal rotation phase (RUN)
(yellow) dark during breaking or alarm
(yellow) dark during breaking
(green) bright in stand by or rotation
(green) bright during rotation
SPB7-115.831.01
Rev.01
5 Rotating anode starter