Solving Common Monitoring Problems

Problem
Erratic trace.
Series 50 Fetal Monitor
Users Only
Suspicious Fetal Movement
Profile.
External Monitoring Uterine Activity (Toco) - External
Artifact on Toco trace.
displayed when using
Err 3
the Series 50 T with an old
8040A Monitor.
Chapter 4 - Monitoring
Possible Causes
Monitoring FHR using Ultrasound
Patient walking heavily.
Maternal movement due
to patient ambulating.
Walk frequency is being
recorded.
Monitoring Intrauterine Pressure (IUP) - Internal
IUP sensitivity setting is
incorrect.
Solving Common Monitoring Problems
Solutions
Ask patient to walk quietly.
Turn off FMP.
None.
Contact your Philips Medical
Response Center to change
setting. Once the setting is
changed, you will only be able
to monitor IUP with the Series
50T.