Manual 41018v0800
Revision B
Keypad Lock
Locking the keypad prevents unauthorized keypad use. The keypad can be locked in two ways:
Manually by entering the keypad lock code.
■
Automatically by activating the Auto Keypad Lock feature in the Drug Library.
■
WARNING
Unauthorized View or Access.
Always guard the keypad lock code from unauthorized view or
access. Uncontrolled access by a patient or family member may
cause injury to the patient.
Automatic keypad lock:
»
When configured in the Drug Library, the keypad will lock automatically by activating the Auto
Keypad Lock feature by Care Area.
The keypad will lock 60 seconds after the RUN key is pressed.
To manually lock the keypad:
»
While the Pump is infusing, enter the keypad lock code 429. A popup message appears briefly
indicating that the keypad has been locked. See Figure 127.
The Keypad Lock icon appears in the upper left corner of the screen when the Pump is locked.
See Figure 128.
Pressing any key while the keypad is locked causes the same popup message to be
displayed. See Figure 127.
100
Figure 127. Keypad Lock Popup Message.
Figure 128. Keypad Lock Icon.