Error codes
This section describes the error messages displayed by the instrument. Contact Technical Support if
the instrument needs to be sent for servicing.
0x8851
0x8852
0x8860
0x8861
0x8862
0x8863
0x8864
0x8867
0x8870
0x2001
0x2002
0x2004
0x2005
0x1100
0x1101
0x1102
0x1200
0x1301
0x1302
0x1303
0x1401
0x1402
0x1500
CTS
™
Xenon
™
Electroporation System User Guide
Error code
Appendix A Troubleshooting
Meaning
Subsystem firmware corrupted. Please update
firmware.
Subsystem firmware version mismatch. Please
update firmware.
Package is not suitable for this instrument.
Instrument class is not supported by this upgrade
package.
Unable to verify upgrade package signature.
Unable to extract upgrade files.
Unable to locate upgrade package.
Unable to recover from upgrade failure.
Unable to save more data, instrument file system is
full.
Unable to detect presence of consumable.
Consumable has not been commissioned.
Wrong consumable type has been loaded.
Consumable has been used.
Unable to start fluid extraction. Liquid sensor active.
Unable to detect fluid during extraction.
Unable to move purger during fluid extraction.
Unable to home cell mixer cup.
Unable to fill chamber from mixer.
Unable to fill chamber.
Unable to move filler pump to target position.
Unable to open stop cock.
Unable to close stop cock.
Tube is not present at extraction pump. (Extractor
tube is not inserted in properly).
A
Error codes
57
Need help?
Do you have a question about the CTS Xenon and is the answer not in the manual?
Questions and answers