P R E P A R A T I O N F O R U S E
2
To Re-Configure the Auto Transformer
To Re-Configure the Auto Transformer
Re-Configure the
Transformer
2-20
Electrical Hazard
Only qualified service personnel familiar with the electrical shock hazards present inside the
equipment should perform any disassembly or corrective maintenance.
Step
1
Power down the System and disconnect the Main Power cord and Main Air
Supply.
2
Remove the right side panel.
To remove the panel:
1.
Unscrew the four quarter-turn locking screws.
2.
Gently pull the panel outward several inches, without damaging the ground
wire (which is attached to the panel).
3.
Detach the protective ground wire; be careful to reconnect the ground wire
upon re-assembly.
3
Locate the High Voltage rail and Voltage Configuration Card:
Voltage
Configuration
Card
High Voltage
Rail
ATS-545 & -645 Interface & Applications Manual
Action