Woodpecker UDS-E Instruction Manual

Woodpecker UDS-E Instruction Manual

Ultrasonic scaler
Hide thumbs Also See for UDS-E:

Advertisement

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the UDS-E and is the answer not in the manual?

Questions and answers

Summary of Contents for Woodpecker UDS-E

  • Page 2: Table Of Contents

    Contents 1. The installation and components of equipment ..........1 2. Product function and operation .................5 3. Sterilization and maintenance ................8 4. Precaution .......................11 5. After service ....................13 6. Symbol instruction ..................13 7. Environmental protection ................14 8. Manufacturer’s right ..................14 9.European authorized representative ..............14 10.
  • Page 3: The Installation And Components Of Equipment

    1.1.4 Automatic frequency tracking ensures that the machine always works on the best frequency and performs more steadily. 1.1.5 Controlled by PC, easy operation and more efficient for scaling. These features make UDS-E become a new generation product in the world dental market. 1.2 Components 1.2.1 The components of machine are listed in the packing list.
  • Page 4 1.3.11 Operating mode: Continuous operation 1.3.12 Type of protection against electric shock: Class II equipment 1.3.13 Degree of protection against electric shock: Type BF applied part 1.3.14 Degree of protection against harmful ingress of water: Ordinary equipment 1.3.15 Protection degree against water (used on the foot switch): IPX1 1.3.16 Applied part of the equipment: handpiece and tip 1.3.17 Degree of safety of application in the presence of a Flammable Anaesthetic Mixture with air or with Oxygen or Nitrous Oxide: Equipment not...
  • Page 5 Figure 2 1.4.2 Sketch map for connection of foot switch, power supply and main unit. Figure 3 1.4.3 Water way connection Sketch map for water bottle installation...
  • Page 6 Figure 4 1.4.4 Sketch map for connection of detachable handpiece Figure 5 1.4.5 Sketch map for how to install tip and endochuck with wrench...
  • Page 7: Product Function And Operation

    Figure 6 2. Product function and operation 2.1 Scaling function 2.1.1 Operation a) Open the packing box, make sure that all the parts and accessories are complete according to the packing list. Take the main unit out of the box and put it on a stable plane.
  • Page 8 The indicator of choosen mode is on. i) Select a suitable scaling tip as you need, screw it on the handpiece tightly by the torque wrench (See Figure 6). j) Step on the foot switch, the tip begins to vibrate. k) The normal frequency is extremely high.
  • Page 9 2.1.3 Instruction of torque wrench (See Figure 6) a) The torque wrench's structure is designed in special way which can control the strength of the scaling tip's installation properly and correctly. It also can guarantee the operator screw or unscrew the scaling tip effectively and keep their hands away from being scratched.
  • Page 10: Sterilization And Maintenance

    purified water, turn the water supply to max, make the machine work with auto- water supply for 30 seconds in order to keep the water way and spare parts clean and durable. f) When the water is lower than water level lower limit, please add water (See Figure 8).
  • Page 11 them with dental lube frequently, as sterilization and repeated pulling and inserting will reduce their working life. Change a new one once it is damaged or worn excessively. e) The following sterilizing methods are forbidden: ① Put handpiece into any liquid for boiling. ②...
  • Page 12 Fault Possible cause Solutions The tip is in loose contact. Contact our dealers or us. The connect plug between Contact our dealers or us. The scaling tip doesn't vibrate the handpiece and the circuit but there is water flowing out board is in loose contact.
  • Page 13: Precaution

    Fault Possible cause Solutions There is no water coming There is air in the water pipe. Turn the water control to the out from the handpiece Max, reinsert the bottle. (automatic water supply mode). If the problem still can't be solved, please contact with local dealer or manufacturer.
  • Page 14 Guilin Woodpecker Medical Instrument Co., Ltd. 4.1.13 The power supply is NOT waterproof. Please keep it dry and away from the water. 4.1.14 After operation, turn off power, and pull out the plug.
  • Page 15: After Service

    4.5.1 Environment temperature: +5°C to +40°C 4.5.2 Relative humidity: 30% ~75% 4.5.3 Atmosphere pressure: 70kPa to 106kPa 4.5.4 A temperature of the water at the inlet : not higher than +25°C 5. After service We offer one year free repair to the equipment according to the warranty card. The repair of the equipment should be carried out by our professional technician.
  • Page 16: Environmental Protection

    Atmospheric pressure Power switch for storage Authorised Representative in the EUROPEAN COMMUNITY 7. Environmental protection Please dispose according to the local laws. 8. Manufacturer’s right We reserve the rights to change the design of the equipment, the technique, fittings, the instruction manual and the content of the original packing list at any time without notice.
  • Page 19: Statement

    The pictures are only for reference. The final interpretation rights belong to GUILIN WOODPECKER MEDICAL INSTRUMENT CO., LTD. The industrial design, inner structure, etc, have claimed for several patents by WOODPECKER, any copy or fake product must take legal responsibilities.
  • Page 20: Table Of Operating Power Of The Tips

    TABLE OF OPERATING POWER OF THE TIPS Scaling Endodontics Tip Model Power Tip Model Power G 10 G 11 E10D Periodontics Tip Model Power E11D P2LD Cavity Preparation P2RD Tip Model Power...
  • Page 21 ZMN/WI-09-071 V3.3- 20190910...