10 - 6
Oscilloscope diagrams
•
MAMMOMAT in mAs mode, see Fig. 3 - Fig. 10.
NOTE
Tube P40 MoW,
MAMMOMAT 3000 Nova
* tungsten anode 0.3
30 kV 20 mAs (P = 4.7 kW)
Fig. 3
30 kV
1 V/T (1 T <=> 5 kV)
158 mA
1 V/T (1 T <=> 40 mA)
127 ms
20 ms/T
NOTE
M1000/3000 Nova
Installation and Start-Up
Checks indicated by
filter disc Mo 0.03 / Rh 0.025 (molybdenum rotating anode).
9
7
1
0
0
M
A
M
Record the measured kV and mAs value in the Test protocol, page
22 - 10.
Register 3
SPB7-230.033.11
Rev. 06
Checks with high voltage
*
shall be disregarded in installations with
Tube P40 MoW,
MAMMOMAT 1000/3000 Nova
molybdenum anode 0.3
30 kV 20 mAs (P = 3.75 kW)
Fig. 4
30 kV
1 V/T (1 T <=> 5 kV)
125 mA
1 V/T (1 T <=> 40 mA)
160 ms
20 ms/T
Page 6 of 10